STARTING UP. EUTECH INSTRUMENTS CYBERSCAN PHION 510 PHION METER, CyberScan pH 510, CyberScan Ion 510
Add to my manuals
36 Pages
advertisement
Instruction Manual pH 510/ Ion 510
2 STARTING UP
Connect the accessory connectors at the rear of the instrument panel. During operation, it is important that water does not get onto the BNC connector. Also avoid touching the connector with soiled or wet hands.
2.1 Connecting the Sensor Electrode
Slide the electrode connector of the electrode over the socket of the meter’s BNC connector marked ‘INPUT’.
The meter can accept any standard pH, ORP, or Ion Selective Electrode with a BNC connector. Align the slot of the electrode connector with the posts of the socket. Rotate the connector clockwise until it locks. For separate reference electrodes, push the electrode pin into jack marked ‘REF’.
2.2 Connecting the Temperature Probe
Insert the temperature probe (provided) into the connector marked ‘ATC’.
2.3 Connecting the A/C Adapter
Slide in the adapter jack of the A/C adapter into the back panel marked ‘DC’ until it is firmly seated. Ensure that the power is switched off before installation and the correct voltage adapter is used.
2.4 Connecting the Chart Recorder
You can connect chart recorders or external output devices for continuous printout. Plug in the input connector and ground pin of the chart recorder into the ports marked ‘REC’ and ‘GND’ respectively.
Sensor electrode Chart recorder
Temp. probe Separate ref. electrode Ground
AC/DC Adapter
4
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 5 INTRODUCTION
- 5 Introducing the 510 Bench Series
- 5 Custom LCD
- 6 Keypad
- 7 Rear Instrument Panel
- 8 STARTING UP
- 8 Connecting the Sensor Electrode
- 8 Connecting the Temperature Probe
- 8 Connecting the A/C Adapter
- 8 Connecting the Chart Recorder
- 9 Connecting the Electrode Holder (Optional)
- 10 Steps Prior to Calibration
- 10 CALIBRATION
- 10 pH Calibration
- 13 Ion Concentration Calibration (For Ion 510 only)
- 15 mV Calibration
- 16 Temperature Calibration
- 17 MEASUREMENT
- 17 Automatic Temperature Compensation (ATC)
- 17 Manual Temperature Compensation (MTC)
- 18 Taking Measurements
- 19 HOLD Function
- 19 Storing and Recalling Data from Memory
- 21 SETUP FUNCTIONS
- 21 P1.0 Clear Memory
- 22 P2.0 Viewing Electrode Diagnosis
- 23 P3.0 Selecting pH Buffer Standard
- 23 P4.0 Meter Reset
- 24 pH Electrode Maintenance
- 24 Storage
- 24 After Use
- 25 Electrode Cleaning
- 26 pH Electrode Rejuvenation
- 27 TROUBLESHOOTING GUIDE
- 27 Error Messages
- 28 Troubleshooting
- 29 ADDITIONAL INFORMATION
- 29 pH and Temperature
- 29 Use of Standard pH Buffers
- 29 Standard pH Buffers
- 30 List of Accessories
- 32 Specifications
- 33 WARRANTY
- 33 RETURN OF ITEMS