DIGITAL CONDUCTIVITY METER SERVICE MANUAL

6.0 TROUBLE SHOOTING CHART

MALFUNCTION PROBABLE CAUSE CORRECTIVE ACTION

Meter does not zero Contaminated Probe Clean thoroughly

with isopropyl

alcohol and air dry

Needs zero adjustment Adjust zero as per 7.2.

Meter reads above or Needs CAL adjustment Check Pre-amp output below calibration as per 7.0

number during CAL check

CU reading erroneous Contaminated or Clean probe with when CAL check is okay damaged probe. isopropyl alcohol

and air dry

Inspect probe for

physical damage.

Check electrical

characteristics as

per 7.0

No Display Indication Lack of supply Check connection and/or

voltage replace batteries

MEASURE/CAL switch Replace Display PC Bd.

defective P/N 152-03-0000

A-D Converter Replace A-D PC Bd.

defective P/N 152-05-0000

+/- 5 Supply In-op Replace Display PC Bd.

P/N 152-03-0000

7.0 CALIBRATION

7.1 With the back panel removed set R-18 (case hole marked CAL) to approximately mid scale. The exact setting is not important at this time, however, the voltage at A-15 should be approximately 1.0 volts above the zero reference (A-14).

7.2 Remove the probe if attached and depress the measure switch (M). Adjust

R-19 (case hole marked ZERO) for a display of 000 +/- 1.

The information contained in the accompanying document is proprietary and confidential, and may not be copied in any manner whatsoever without prior written consent of Emcee Electronics,

Inc. The document and the material therein may not be used for any purpose other than that intended by Emcee Electronics, Inc.

COPYRIGHT 2002 EMCEE ELECTRONICS, INC.

REVISION DATE: FEBRUARY 23, 1996

PAGE 6 OF 14

advertisement

Key Features

  • Handheld, battery-operated
  • Measures fuel conductivity
  • High-range ohmmeter
  • Probe included
  • Digital display
  • Calibration feature

Related manuals

Frequently Answers and Questions

What is the purpose of the Emcee Digital Conductivity Meter Model 1152?
The Emcee Digital Conductivity Meter Model 1152 is used to measure the conductivity of fuel samples.
How do I calibrate the Emcee Digital Conductivity Meter Model 1152?
The calibration procedure is described in detail in section 7.0 of the manual. It involves using a CU substitution box and adjusting specific potentiometers on the meter.
What are the battery requirements for the Emcee Digital Conductivity Meter Model 1152?
The meter uses three 6 volt Alkaline Nema 1414A batteries (Emcee P/N 152-90-5341). Replacing them with any other battery type will void the CSA and UL Intrinsically Safe Rating.