- Industrial & lab equipment
- Measuring, testing & control
- Multimeters
- Fluke
- 1654B06-TPOLEKIT/DVDE-tester
- User manual
advertisement
Maintaining the Tester
Maintaining the Tester
Cleaning
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
1. Turn the meter off and remove all test leads.
2. Shake out any dirt that may be in the terminals.
3. Soak a new swab with alcohol. Work the swab around each terminal.
Testing and Replacing the Batteries
Battery voltage is continuously monitored by the tester. If the voltage falls below
6.0 V (1.0 V/cell), the low battery icon appears on the display, indicating that there is minimal battery life left. The low battery icon continues to appear on the display until you replace the batteries.
Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the
battery icon () appears.
Be sure that the battery polarity is correct. A reversed battery can cause leakage.
Replace the batteries with six AA batteries. Alkaline batteries are supplied with the tester but you can also use 1.2 V NiCd or NiMH batteries. You can also check the battery charge so that you can replace them before they discharge.
Warning
To avoid electrical shock or personal injury, remove the test leads and any input signals before replacing the battery. To prevent damage or injury, install ONLY specified replacement fuses with the amperage, voltage, and speed ratings shown in the
General Specifications section of this manual.
To test the batteries:
1. Turn the rotary switch to the V position.
to initiate the battery test. The Voltage function display clears and is replaced by the measured battery voltage in the secondary display for 2 seconds, the Voltage function display then returns.
41
1652C/1653B/1654B
Users Manual
To replace the batteries (refer to Figure 24):
2. Remove the test leads from the terminals.
3. Remove the battery door by using a standard-blade screwdriver to turn the battery door screws (3) one-quarter turn counterclockwise.
4. Press the release latch and slide the battery holder out of the tester.
5. Replace the batteries and the battery door.
Note
All stored data will be lost if the batteries are not replaced within approximately one minute (Models 1653B and 1654B only).
6. Secure the door by turning the screws one-quarter turn clockwise.
42
Figure 24. Replacing the Batteries
apx028f.eps
Maintaining the Tester
Testing the Fuse
A fuse test is performed each time you turn on the tester. If leads are plugged into the L and PE terminals, the fuse test is skipped. If a blown fuse is detected, testing is disabled, FUSE appears on the primary display, and the tester issues a warning beep.
You can also perform a manual check of the fuse.
To manually check the fuse:
1. Turn the rotary switch to either or switch setting.
2. Short the leads and press and hold .
3. If the fuse is bad, FUSE will appear on the display to indicate the tester is damaged and needs repair. Contact Fluke Service for repair (see
Contacting Fluke).
43
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 1 1652C/1653B/1654B Users Manual
- 2 Warranty
- 3 Table of Contents
- 5 List of Tables
- 7 List of Figures
- 9 Introduction
- 9 How to Contact Fluke
- 10 Safety
- 12 Unpacking the Tester
- 14 Operating the Tester
- 14 Using the Rotary Switch
- 15 Understanding the Pushbuttons
- 17 Understanding the Display
- 23 Input Terminals
- 23 Using the IR Port
- 24 Error Codes
- 25 Power-On Options
- 27 Making Measurements
- 27 Measuring Volts and Frequency
- 28 Measuring Insulation Resistance
- 29 Measuring Continuity
- 30 Measuring Loop/Line Impedance
- 30 Loop Impedance (Line to Protective Earth L-PE)
- 33 Earth Resistance Testing by Loop Method
- 33 Line Impedance
- 36 Measuring RCD Tripping Time
- 40 Measuring RCD Tripping Current
- 42 RCD Testing in IT Systems
- 42 Measuring Earth Resistance
- 44 Testing Phase Sequence
- 45 Memory Mode
- 46 Storing a Measurement
- 46 Recalling a Measurement
- 47 Clearing Memory
- 48 Uploading Test Results
- 49 Maintaining the Tester
- 49 Cleaning
- 49 Testing and Replacing the Batteries
- 51 Testing the Fuse
- 52 Specifications
- 52 Features by Model
- 53 General Specifications
- 54 Category Ratings and Usage
- 54 Electrical Measurement Specifications
- 54 Insulation Resistance (RISO)
- 55 Continuity (RLO)
- 56 Loop Tests (ZI)
- 57 RCD/FI Tests ((T, I(N)
- 58 Earth Tests (RE)
- 58 AC Voltage Measurement (V)
- 58 Continuity Testing (RLO)
- 59 Insulation Resistance Measurement (RISO)
- 60 No Trip and Hi Current Modes RCD/FI
- 60 Prospective Earth Fault Current Test (PSC/IK)
- 61 RCD Testing
- 61 RCD Types Tested
- 61 Test Signals
- 62 Tripping Speed Test ((T)
- 62 Maximum Trip Time
- 63 RCD/FI-Tripping Current Measurement/Ramp Test (I(N )
- 63 Earth Resistance Test (RE )
- 64 Phase Sequence Indication
- 64 Mains Wiring Test
- 65 Operating Ranges and Uncertainties per EN 61557
- 66 Operating Uncertainties per EN 61557