Getting Started

Private Community String

Public Community String

This menu entry allows you to set the Private Community string.

This string is effectively a password. Without this string, SNMP commands cannot change values in the instrument.

NOTE.

The Private String is necessary for SNMP access to write changes into the instrument.

This menu entry allows you to set the Public Community string.

This string is effectively a password. Without this string, SNMP commands cannot read values from the instrument.

NOTE.

The Public String is necessary for SNMP access to read values from the instrument.

Incoming Inspection

The incoming inspection procedures are optional procedures to check the functionality of your instrument. These procedures require no equipment other than an external monitor or display. For a more robust inspection, see the performance verification procedures in the WVR7200 Specifications and

Performance Verification Technical Reference that is included on the Product

Documentation CD that shipped with your instrument.

Basic Turn On and Self

Test

The following test can performed with or without the external monitor. Using the external monitor allows you to check that the DVI connector on the rear panel is working properly.

1. Connect a monitor to the EXT DISPLAY connector on the rear panel of the instrument. This connector is a DVI-I connector with socket contacts. The output supports DVI monitors directly and analog PC (RGB) monitors using a

DVI-I to VGA adapter.

2. Connect the AC line cord to the rear of the instrument and to a 100 to

240 VAC source.

10 WVR7200 Waveform Rasterizers User Manual

Getting Started

Figure 2: Rear panel

3. Press the Power button on the front panel to turn the instrument on. It is located on the bottom left corner of the front panel.

WVR7200 Waveform Rasterizers User Manual 11

Getting Started

Figure 3: Front panel

4. After about 30 seconds, the power-on diagnostic page should appear on the screen.

5. Verify that all self tests pass. Any failures are shown in red. The results of the power-on diagnostics are erased from the screen, but you can view the results by selecting CONFIG > Utilities > View Diagnostics Log > SEL or by viewing them on the instrument Web page.

6. After the diagnostics are finished, the instrument state is restored. When the progress indicator in the status bar is finished, the instrument has finished initializing.

7. Record Pass or Fail for the POST (Power On Self Test) in the test record.

8. If it is still open, exit the Diagnostics Log.

Front Panel Test

1. Press and hold the PRESET button.

2. Use the GENERAL knob to navigate to Recall Preset.

3. Press the right and down arrow keys to navigate to Recall Factory Preset and press the SEL button.

12 WVR7200 Waveform Rasterizers User Manual

Getting Started

4. Press the FULL button to change the display screen from 4-tile view to single-tile view.

5. Press the HELP button to display the help screen.

6. Press all of the other buttons, one at a time.

Each button should flash as you press it. Most buttons show related help text.

Some buttons, such as the presets, show the same information. The HELP button, the four arrow keys, and the SEL button do not show help text, but are used to navigate the help panes and content.

Figure 4: Instrument online help

7. Press the right arrow key until the Help Contents pane in the upper-left corner of the help display is highlighted.

8. Turn the GENERAL knob and check that the selector box moves up and down the list.

9. Press the HELP button to exit help.

Fan Test

You should be able to hear the fans and feel air coming out of the back of the instrument. At low temperatures, the fans will turn slowly and be very quiet.

WVR7200 Waveform Rasterizers User Manual 13