Manuals and User Guides for Velleman DVM171THD. We found 4 manuals for free downloads User manual, User's manual

Below you will find brief information for DVM171THD. This device logs temperature and humidity with 32,000 readings, dew point indication, USB interface, and user-selectable alarms. It supports multi-mode logging, long battery life, and customizable measuring cycles for versatile environmental monitoring.

Key Features

  • Memory for 32,000 readings (16,000 temp, 16,000 humidity)

  • Dew point indication

  • USB interface

  • User-selectable alarms

  • Analysis software

  • Multi-mode start

  • Long battery life

  • Selectable measuring cycle (2s–24h)

Pages: 27 Velleman DVM171THD User manual

Brand: Velleman Category: Measuring & layout tools Size: 412 KB

Languages: German, English, Spanish, French, Dutch

Open manual

Other manuals and documents

40 Pages

User manual

Languages: German, English, Spanish, ...

27 Pages

User's manual

Languages: German, English, Spanish, ...

User manual

Languages: Italian

Frequently Answers and Questions

How do I start logging manually?

Press the start button until the green and yellow LED flash once together. The green LED will then flash every 10 seconds to confirm logging.

What do the LED signals mean?

Green LED: Logging active. Red LED: Humidity alarm. Yellow LED: Temperature alarm. Both LEDs off: Low battery or logging inactive.

How do I replace the battery?

Open the battery compartment with a pointed object, remove the old battery, insert a 3.6V lithium battery, and close the case.

How do I recondition the sensors?

Heat the datalogger to +80°C (humidity <5%) for 36 hours, then store it at +20–30°C (humidity >74%) for 48 hours.

What to do when memory is full?

The yellow LED flashes every 60 seconds. Download data and clear memory to resume logging.