6. Demonstration

Fig.6-5 Waveform of the Noise-Removed Signal

Example 5: Application of X-Y Function

Examine the Phase Difference between Signals of two Channels

Example: Test the phase change of the signal after it passes through a circuit network.

Connect the oscilloscope with the circuit and monitor the input and output signals of the circuit.

For the examination of the input and output of the circuit in the form of X-Y coordinate graph, please operate according to the following steps:

(1) Set the probe menu attenuation coefficient for 10X and that of the switch in the

probe for 10X (see “How to Set the Probe Attenuation Coefficient” on P14).

(2) Connect the probe of channel 1 to the input of the network and that of Channel 2 to the output of the network.

(3) Push down the Autoset button, with the oscilloscope turning on the signals of the two channels and displaying them in the screen.

(4) Turn the VOLTS/DIV knob, making the amplitudes of two signals equal in the rough.

(5) Press the Display button and recall the Display menu.

(6) Press the H3 button and choose XY Mode as ON.

The oscilloscope will display the input and terminal characteristics of the network in the Lissajous graph form.

(7) Turn the VOLTS/DIV and VERTICAL POSITION knobs, optimizing the

77