OPERATION
When the low temperature option is fitted an extra 100 cm space on the left-hand side of the system is required.
Unpacked, the largest subassembly will fit through a door aperture of 85 cm. Check the door aperture to ensure the system can be assembled in its designated area.
5.2.1.3 Electrical Services
A 3-phase 63 A supply with one 32A outlet and two 16A outlets. Additionally four single-phase outlets are required as described in drawing OD-01-00-015-C.
Use only the power cables supplied.
Do not connect the electrical power supply circuit to any other devices. Limit the electrical noise in the system by attaching the earth cable exclusively to an external earth terminal with a resistance of less than 0.5 ohms.
Fit a line voltage regulator if the power supply voltage fluctuates more than ±10 %.
Locate the mains outlet on the wall behind the system. The mains outlet should be of the circuit breaker type. (Outlet and connecting plugs are not supplied). The mains plug should be readily accessible by the operator when the equipment has been installed.
In areas where the mains power supply is unreliable an ‘uninterruptible power supply’ (UPS) is recommended. The UPS should have specifications of 10 kVA with 3-phase input and single phase output.
V Hz
Maximum mains current
A
Main fuse
A
Maximum power
W
X-ray generator
CCD detector &
KMW200CCD cooler
1/N AC 230 ±10% 50/60
1/N AC 230 ±10% 50/60
24
1
32
1.6
4200
300
2.5 6.3 400 Kappa geometry, X-ray goniometer, video monitor, halogen lamp, fibre optics
1/N AC 230 ±10% 50/60
Computer and peripherals
100 - 240 50/60
Cryojet controller
ILM201
100 - 240
100 - 230
50/60
50/60
KMW30000C Cooler 1/N AC 230 ±10% 50/60
10
2
0.6
1.2
10
2.5 – 5
0.8 –
1.6
6.3 300
Version 1.4 Xcalibur_Manual_v1.4
Page 26
Xcalibur
USER MANUAL
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 10 1. Health and Safety Information
- 10 1.1 General
- 11 1.2 Electrical Safety
- 11 1.2.1 Potential Electrical Hazards
- 11 1.2.2 Recommended Precautions
- 11 1.2.3 First Aid
- 12 1.3 Mechanical Handling Safety
- 12 1.4 Safe Mechanical Practice
- 12 1.5 Moving Parts
- 12 1.6 X-ray Radiation
- 13 1.7 Extreme Temperatures
- 14 1.8 Vacuum
- 14 1.9 High Pressures
- 14 1.10 Hazardous or Toxic Materials
- 14 1.11 Modifications and Service
- 15 2. Introduction
- 15 2.1 Scope
- 15 2.2 How To Use This Manual
- 15 2.3 System Overview
- 16 3. Specifications
- 16 3.1 Environmental Requirements
- 16 3.2 Services
- 16 3.2.1 Electrical Supply
- 16 3.2.2 Water Cooling
- 17 3.2.3 Helium Gas Supply (where applicable)
- 17 3.3 Performance Data
- 17 3.3.1 X-ray Tube (Typical Operating Conditions)
- 18 3.3.2 Sapphire 2 CCD Detector
- 19 3.3.3 Sapphire 3 CCD Detector
- 20 3.3.4 Onyx CCD Detector
- 20 Onyx CCD Detector Theta and Resolution Ranges
- 21 3.3.5 PC CCD Interface
- 21 3.3.6 Four-circle Kappa Geometry X-ray Goniometer
- 22 4.4 Electrical Data
- 23 4. Technical Description
- 23 4.1 Overview
- 24 4.2 CCD Detector Technology
- 25 4.2.1 Beryllium Window
- 26 4.2.2 Phosphor
- 26 4.2.3 Taper
- 26 4.2.4 CCD
- 26 4.2.5 Readout Speed
- 26 4.2.6 Binning
- 27 4.2.7 Dark Current and MPP Mode
- 27 4.2.8 Radiation Damage
- 27 4.2.9 Full Well Depth and 18-bit Digitisation
- 27 4.2.10 Anti-blooming
- 27 4.2.11 Vacuum
- 27 4.2.12 Fast Shutter
- 28 4.2.13 Zingers and Cosmic Ray Events
- 28 4.3 Four-Circle Kappa Geometry Goniometer
- 29 4.4 X-ray Generator
- 30 4.5 Software
- 30 4.5.1 Directory Structure
- 31 4.5.2 Basic Menu Philosophy
- 31 4.6 KMW200CCD Chiller
- 31 4.7 KMW3000C Chiller
- 31 4.8 Low Temperature Option
- 31 4.9 Safety Features
- 32 5. Handling, Installation, Storage and Transit Information
- 32 5.1 Reception and Handling
- 32 5.1.1 Delivery
- 32 5.1.2 Unpacking
- 33 5.1.3 Mechanical Handling
- 33 5.1.3.1 Weights, Dimensions and Lifting Points
- 34 Delivery
- 34 5.2 Installation and Setting to Work
- 34 5.2.1 Preparation of Site and Services
- 34 5.2.1.1 Environmental Requirements
- 34 5.2.1.2 System Layout
- 35 5.2.1.3 Electrical Services
- 36 5.2.1.4 Water Supply
- 36 5.2.1.5 Low Temperature Option
- 36 5.2.1.6 CCD Camera Pumping
- 36 5.2.1.7 Helijet Option
- 37 5.2.2 Setting to Work
- 37 5.2.2.1 Equipment Required
- 37 5.2.2.2 Personnel Required for Installation
- 37 5.2.2.3 Setting up Procedures
- 39 Storage
- 40 6. Operation
- 40 6.1 Controls and Indicators
- 41 6.2 Initial Switch on Procedure
- 42 6.3 X-ray Tube Warm-up Procedure
- 43 6.4 Software
- 43 6.4.1 Software Updates
- 44 6.4.2 Software Installation
- 44 6.4.2.1 MGC interface software
- 45 6.4.2.2 CrysAlis Software
- 45 6.4.3 Changing Machine Correction and Set-up Files
- 46 6.5 Normal Operation
- 46 6.5.1 General Commands
- 48 6.5.2 Changing Xcalibur Settings
- 52 6.5.3 Standard Diffraction Experiment
- 52 6.5.3.1 Crystal Mounting and Alignment
- 54 6.5.3.2 Diffraction Photographs to Determine Crystal Quality
- 55 6.5.3.3 Unit Cell Determination
- 57 6.5.3.4 Data Collection
- 58 6.5.3.5 Data Processing and Reduction
- 59 6.5.3.5.1 Orientation Matrix
- 60 6.5.3.5.2 Run List
- 61 6.5.3.5.3 Scan Width
- 61 6.5.3.5.4 Background Evaluation
- 62 6.5.3.5.5 Special Corrections
- 63 6.5.3.5.6 Outlier Rejection
- 64 6.5.3.5.7 Output Format
- 65 6.5.3.6 Changing the Output Format from Data Reduction
- 66 6.5.3.7 Absorption Correction
- 74 6.5.3.8 GRAL - Space Group Determination
- 80 6.5.3.9 Structure Solution and Refinement
- 80 6.5.4 Ewald explorer
- 86 6.5.5 Dc Movie - Replay of Data Collection Movie
- 87 6.5.6 Reconstruction of Precession Photographs
- 91 6.5.7 Dc opti - Optimisation of Data Collection Strategy
- 96 6.5.8 Indexing and Data Reduction of Incommensurate Samples
- 97 6.5.9 Indexing and Data Reduction of Twinned Samples
- 99 6.5.10 Extracting Data from Powder Samples
- 100 6.5.11 Refining of Machine Parameter File
- 103 6.5.12 Glossary of CrysAlis Commands
- 110 6.6 Normal Shutdown
- 110 6.7 Emergency Shutdown
- 111 6.7.1 Emergency Shutdown Procedure
- 112 7. Mechanical Changeover of Detectors and X-ray Sources
- 112 7.1 Interchange of CCD Detectors
- 112 7.1.1 Installation of a Sapphire 2 and Sapphire 3 CCD detectors
- 113 7.1.2 Removal of a Sapphire 2 and Sapphire 3 CCD detectors
- 113 7.1.3 Installation of the Onyx CCD camera
- 116 7.1.4 Removal of the Onyx CCD camera
- 117 7.2 Procedure for Interchange of the Molybednum and Copper Enhance X-ray Source
- 121 8. Maintenance Schedules
- 121 8.1 Introduction
- 121 8.2 Weekly Maintenance Schedule
- 121 8.3 Monthly Maintenance Schedule
- 122 8.4 Six Monthly Maintenance Schedule
- 122 8.5 Yearly Maintenance Schedule
- 123 8.6 10,000 Hours Maintenance Schedule
- 124 9. Maintenance Instructions
- 124 9.1 Special Tools
- 124 9.2 Refining the Machine Parameter File
- 125 9.3 Changing the X-ray Tube of Enhance
- 126 9.4 Changing the X-ray Tube of Enhance ULTRA
- 128 9.5 X-ray Beam Stop Alignment
- 129 9.6 Changing the Collimator of Enhance
- 130 9.7 Changing the Collimator of Enhance Ultra
- 130 9.8 Aligning the X-ray Collimator of Enhance
- 130 9.9 Aligning the Enhance X-ray Source
- 133 9.10 Aligning the Enhance Ultra X-ray Source
- 133 9.10.1 X-ray Beam Alignment of Enhance Ultra
- 135 9.10.2 Optic Alignment of Enhance Ultra
- 136 9.10.3 Collimator Alignment of Enhance Ultra
- 136 9.10.4 Aligning the beam to the centre of the goniometer – Enhance Ultra
- 137 9.11 Checking the Door Safety Interlocks
- 137 9.12 Checking the Emergency stop
- 138 9.13 Checking the X-ray Radiation Levels
- 138 9.14 CCD Detector – Pumping Out Vacuum
- 140 9.15 Dismantling Xcalibur
- 143 10. Trouble Shooting
- 146 11. Spares
- 148 12. Disposal Instructions
- 148 12.1 X-ray Tube and CCD Detector
- 148 12.2 Third Party Equipment
- 149 13. Additional Information
- 149 13.1 Third Party Information
- 150 13.2 Drawings
- 150 13.2.1 Mechanical Drawings
- 150 Xcalibur Suggested Layout
- 150 System and Component Dimensions
- 152 13.2.2 Electrical Drawings
- 154 14. CE Conformity notice
- 155 Appendices
- 155 Appendix 1 X-ray Tubes Wave Lengths
- 155 Appendix 2 Standard Crystal Parameters
- 155 Appendix 3 Temperature Scales Conversion
- 155 Appendix 4 Maintenance Records
- 160 Appendix 5 Example of Local Rules for the Xcalibur System Set Up at Oxford Diffraction