MAINTENANCE INSTRUCTIONS
3. Record details of the test in the local radiation log, stipulating the date, person testing, outcome and signature.
9.13 Checking the X-ray Radiation Levels
Task Time:
20 minutes
When:
Once a month
Tools:
Radiation meter
Procedure
1. With the X-ray generator on (at normal settings) and the shutter closed, use the radiation meter to sweep the area around the X-ray tube housing, fast shutter and collimator for any radiation leak (inside the enclosure).
2. Close the protective enclosure doors and open the X-ray shutter (command sh o in CrysAlis
CCD)
3. Sweep the outside of the enclosure using the radiation meter, paying particular attention to the plane of the X-ray tube, the area of the secondary beam stop (mounted on the enclosure window) and the door seals.
4. Record details of the test in the local radiation log, stipulating the date, person testing, outcome and signature.
9.14 CCD Detector – Pumping Out Vacuum
Task Time:
16 hours
When:
Once a year
Tools:
Special ‘T’ tool
1 set of Allen keys
Rotary vacuum pump capable of obtaining a pressure of 1x10
-1
mbar
Vacuum tubing and adaptor
Procedure
1. Switch off the cooling water / KMW200CCD chiller unit.
2. Detach water pipes from top of CCD detector (these are self-sealing, although there may be a few drops of water, have something absorbent to hand)
3. Unscrew top access plate of CCD detector (See figure 9.14.1 for details of specific CCD detector).
Version 1.4 Xcalibur_Manual_v1.4
Page 129
Xcalibur
USER MANUAL
Sapphire 2
MAINTENANCE INSTRUCTIONS
Access plate of Sapphire 2 CCD detector removed. Secured by one screw at centre.
Pump out port (with ‘T’ tool attached)
Sapphire 3
Access plate of Sapphire 3
(false front to head – spring loaded). Secured by 4 grub screws at corners (2 shown).
Pump out port (with ‘T’ tool attached)
Onyx
Pump out port (with cover unscrewed)
Figure 9.14.1 View of the access plates and pump out ports of the Sapphire and Onyx CCD detectors
4. Screw into place the ‘T’ tool.
Version 1.4 Xcalibur_Manual_v1.4
Page 130
Xcalibur
USER MANUAL
advertisement
Related manuals
advertisement
Table of contents
- 10 1. Health and Safety Information
- 10 1.1 General
- 11 1.2 Electrical Safety
- 11 1.2.1 Potential Electrical Hazards
- 11 1.2.2 Recommended Precautions
- 11 1.2.3 First Aid
- 12 1.3 Mechanical Handling Safety
- 12 1.4 Safe Mechanical Practice
- 12 1.5 Moving Parts
- 12 1.6 X-ray Radiation
- 13 1.7 Extreme Temperatures
- 14 1.8 Vacuum
- 14 1.9 High Pressures
- 14 1.10 Hazardous or Toxic Materials
- 14 1.11 Modifications and Service
- 15 2. Introduction
- 15 2.1 Scope
- 15 2.2 How To Use This Manual
- 15 2.3 System Overview
- 16 3. Specifications
- 16 3.1 Environmental Requirements
- 16 3.2 Services
- 16 3.2.1 Electrical Supply
- 16 3.2.2 Water Cooling
- 17 3.2.3 Helium Gas Supply (where applicable)
- 17 3.3 Performance Data
- 17 3.3.1 X-ray Tube (Typical Operating Conditions)
- 18 3.3.2 Sapphire 2 CCD Detector
- 19 3.3.3 Sapphire 3 CCD Detector
- 20 3.3.4 Onyx CCD Detector
- 20 Onyx CCD Detector Theta and Resolution Ranges
- 21 3.3.5 PC CCD Interface
- 21 3.3.6 Four-circle Kappa Geometry X-ray Goniometer
- 22 4.4 Electrical Data
- 23 4. Technical Description
- 23 4.1 Overview
- 24 4.2 CCD Detector Technology
- 25 4.2.1 Beryllium Window
- 26 4.2.2 Phosphor
- 26 4.2.3 Taper
- 26 4.2.4 CCD
- 26 4.2.5 Readout Speed
- 26 4.2.6 Binning
- 27 4.2.7 Dark Current and MPP Mode
- 27 4.2.8 Radiation Damage
- 27 4.2.9 Full Well Depth and 18-bit Digitisation
- 27 4.2.10 Anti-blooming
- 27 4.2.11 Vacuum
- 27 4.2.12 Fast Shutter
- 28 4.2.13 Zingers and Cosmic Ray Events
- 28 4.3 Four-Circle Kappa Geometry Goniometer
- 29 4.4 X-ray Generator
- 30 4.5 Software
- 30 4.5.1 Directory Structure
- 31 4.5.2 Basic Menu Philosophy
- 31 4.6 KMW200CCD Chiller
- 31 4.7 KMW3000C Chiller
- 31 4.8 Low Temperature Option
- 31 4.9 Safety Features
- 32 5. Handling, Installation, Storage and Transit Information
- 32 5.1 Reception and Handling
- 32 5.1.1 Delivery
- 32 5.1.2 Unpacking
- 33 5.1.3 Mechanical Handling
- 33 5.1.3.1 Weights, Dimensions and Lifting Points
- 34 Delivery
- 34 5.2 Installation and Setting to Work
- 34 5.2.1 Preparation of Site and Services
- 34 5.2.1.1 Environmental Requirements
- 34 5.2.1.2 System Layout
- 35 5.2.1.3 Electrical Services
- 36 5.2.1.4 Water Supply
- 36 5.2.1.5 Low Temperature Option
- 36 5.2.1.6 CCD Camera Pumping
- 36 5.2.1.7 Helijet Option
- 37 5.2.2 Setting to Work
- 37 5.2.2.1 Equipment Required
- 37 5.2.2.2 Personnel Required for Installation
- 37 5.2.2.3 Setting up Procedures
- 39 Storage
- 40 6. Operation
- 40 6.1 Controls and Indicators
- 41 6.2 Initial Switch on Procedure
- 42 6.3 X-ray Tube Warm-up Procedure
- 43 6.4 Software
- 43 6.4.1 Software Updates
- 44 6.4.2 Software Installation
- 44 6.4.2.1 MGC interface software
- 45 6.4.2.2 CrysAlis Software
- 45 6.4.3 Changing Machine Correction and Set-up Files
- 46 6.5 Normal Operation
- 46 6.5.1 General Commands
- 48 6.5.2 Changing Xcalibur Settings
- 52 6.5.3 Standard Diffraction Experiment
- 52 6.5.3.1 Crystal Mounting and Alignment
- 54 6.5.3.2 Diffraction Photographs to Determine Crystal Quality
- 55 6.5.3.3 Unit Cell Determination
- 57 6.5.3.4 Data Collection
- 58 6.5.3.5 Data Processing and Reduction
- 59 6.5.3.5.1 Orientation Matrix
- 60 6.5.3.5.2 Run List
- 61 6.5.3.5.3 Scan Width
- 61 6.5.3.5.4 Background Evaluation
- 62 6.5.3.5.5 Special Corrections
- 63 6.5.3.5.6 Outlier Rejection
- 64 6.5.3.5.7 Output Format
- 65 6.5.3.6 Changing the Output Format from Data Reduction
- 66 6.5.3.7 Absorption Correction
- 74 6.5.3.8 GRAL - Space Group Determination
- 80 6.5.3.9 Structure Solution and Refinement
- 80 6.5.4 Ewald explorer
- 86 6.5.5 Dc Movie - Replay of Data Collection Movie
- 87 6.5.6 Reconstruction of Precession Photographs
- 91 6.5.7 Dc opti - Optimisation of Data Collection Strategy
- 96 6.5.8 Indexing and Data Reduction of Incommensurate Samples
- 97 6.5.9 Indexing and Data Reduction of Twinned Samples
- 99 6.5.10 Extracting Data from Powder Samples
- 100 6.5.11 Refining of Machine Parameter File
- 103 6.5.12 Glossary of CrysAlis Commands
- 110 6.6 Normal Shutdown
- 110 6.7 Emergency Shutdown
- 111 6.7.1 Emergency Shutdown Procedure
- 112 7. Mechanical Changeover of Detectors and X-ray Sources
- 112 7.1 Interchange of CCD Detectors
- 112 7.1.1 Installation of a Sapphire 2 and Sapphire 3 CCD detectors
- 113 7.1.2 Removal of a Sapphire 2 and Sapphire 3 CCD detectors
- 113 7.1.3 Installation of the Onyx CCD camera
- 116 7.1.4 Removal of the Onyx CCD camera
- 117 7.2 Procedure for Interchange of the Molybednum and Copper Enhance X-ray Source
- 121 8. Maintenance Schedules
- 121 8.1 Introduction
- 121 8.2 Weekly Maintenance Schedule
- 121 8.3 Monthly Maintenance Schedule
- 122 8.4 Six Monthly Maintenance Schedule
- 122 8.5 Yearly Maintenance Schedule
- 123 8.6 10,000 Hours Maintenance Schedule
- 124 9. Maintenance Instructions
- 124 9.1 Special Tools
- 124 9.2 Refining the Machine Parameter File
- 125 9.3 Changing the X-ray Tube of Enhance
- 126 9.4 Changing the X-ray Tube of Enhance ULTRA
- 128 9.5 X-ray Beam Stop Alignment
- 129 9.6 Changing the Collimator of Enhance
- 130 9.7 Changing the Collimator of Enhance Ultra
- 130 9.8 Aligning the X-ray Collimator of Enhance
- 130 9.9 Aligning the Enhance X-ray Source
- 133 9.10 Aligning the Enhance Ultra X-ray Source
- 133 9.10.1 X-ray Beam Alignment of Enhance Ultra
- 135 9.10.2 Optic Alignment of Enhance Ultra
- 136 9.10.3 Collimator Alignment of Enhance Ultra
- 136 9.10.4 Aligning the beam to the centre of the goniometer – Enhance Ultra
- 137 9.11 Checking the Door Safety Interlocks
- 137 9.12 Checking the Emergency stop
- 138 9.13 Checking the X-ray Radiation Levels
- 138 9.14 CCD Detector – Pumping Out Vacuum
- 140 9.15 Dismantling Xcalibur
- 143 10. Trouble Shooting
- 146 11. Spares
- 148 12. Disposal Instructions
- 148 12.1 X-ray Tube and CCD Detector
- 148 12.2 Third Party Equipment
- 149 13. Additional Information
- 149 13.1 Third Party Information
- 150 13.2 Drawings
- 150 13.2.1 Mechanical Drawings
- 150 Xcalibur Suggested Layout
- 150 System and Component Dimensions
- 152 13.2.2 Electrical Drawings
- 154 14. CE Conformity notice
- 155 Appendices
- 155 Appendix 1 X-ray Tubes Wave Lengths
- 155 Appendix 2 Standard Crystal Parameters
- 155 Appendix 3 Temperature Scales Conversion
- 155 Appendix 4 Maintenance Records
- 160 Appendix 5 Example of Local Rules for the Xcalibur System Set Up at Oxford Diffraction