Omega FMA3100, 3100ST, 3300, 3300ST Series Owner Manual


Add to my manuals
29 Pages

advertisement

Omega FMA3100, 3100ST, 3300, 3300ST Series Owner Manual | Manualzz

FMA 3100/3100ST/3300/3300ST Series

Thermal Mass Flow Sensors and Meters

M-4270/0707, pg. 2 of 29

READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR

OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY RESULT IN INJURY

TO YOU OR DAMAGE TO THE FLOW SENSOR.

T A B L E O F C O N T E N T S

A. Introduction ............................................................................................. 4

1. Unpacking........................................................................................... 4

2. Product Overview And Principle Of Operation .......................................... 4

B. Installation............................................................................................... 5

1. General Considerations......................................................................... 5

2. Mounting The Flow Sensor Or Flow Meter ............................................... 6

3. Tubing Connections .............................................................................. 7 a) General ....................................................................................... 7 b) High Flow Units (0-100 L/min and above) ............................................ 7

5. Electrical Connections ........................................................................... 7 a) Overview .................................................................................... 7 b) Connecting the 6 Pin Mini Din Connector............................................. 8 c) Connecting the 6 Pin Mini Din Connector And FMA 3000C Cable........... 9 d) Connections for the 9 Pin D Sub Connector........................................ 10 e) Connections for the 15 Pin D Sub Connector ...................................... 11 f) Using the 0-5VDC Output Power Adapter Package .............................. 12

C. Operation ............................................................................................. 13

1. Warm-Up .......................................................................................... 13

2. Verification Of Zero ............................................................................ 13

3. Flow Readings.................................................................................... 13

4. Zero Adjustments ................................................................................ 14

5. Recalibration...................................................................................... 15

D. Maintenance And Product Care................................................................ 15

1. General............................................................................................. 15

2. Returning Units For Repair Or Recalibration............................................ 15

E. Specifications......................................................................................... 16

F. Dimensions............................................................................................ 17

G. Gas K Factors ........................................................................................ 25

H. Trouble Shooting Guide .......................................................................... 26

M-4270/0707, pg. 3 of 29

A.

Introduction

1. Unpacking

All sensors are suitably packaged to prevent damage during shipping. If external damage is noted upon receipt of the package, please contact

Omega Engineering immediately.

Open the package from the top, taking care not to cut too deeply into the package. Remove all the documentation and contents. Take care to remove all the items and check them against the packing slip. The products should also be checked for any concealed shipping damage. If any shortages or damage is noted, please contact Omega Engineering to resolve the problem.

Contents of Box- Sensor, Calibration Certificate & Manual

FMA 3100 Series shown; FMA 3300 series has an integrated display.

Caution: Take care not to

drop

your sensor. Read the installation section of this manual before providing power or tubing connections to the unit. Any damage caused by improper installation or careless handling will not be repaired under warranty (see limited warranty on page 28 for more details).

2. Product Overview and Principle of Operation

The FMA 3100/3300 Series flow sensors and flow meters from Omega

Engineering are capable of measuring virtually any clean, dry gas as low as 0-20 sccm or as high as 0-500 l/min. Repeatable results are achieved using a patented thermal mass flow sensor design. This proven design

M-4270/0707, pg. 4 of 29

minimizes zero drift while maintaining fast response and linear outputs with virtually no maintenance.

The FMA 3100/3300 Series utilizes thermal sensing technology. A portion of the gas flowing through the unit is redirected into a small sensor tube.

This tube has two coils on the outside. The first coil introduces a small amount of heat into the gas stream. As the gas passes through the tube, heat is transferred from one coil to the other. The flow rate is proportional to the amount of heat transfer. Smart electronics analyze the amount of temperature change in the second coil and provide a linearized analog output. A patented system insures that the zero remains stable and the sensor is extremely repeatable.

The output of the thermal mass flow sensor is directly related to the specific heat characteristic of the gas being measured. A sensor is calibrated for one gas but may be used with other gases by applying a correction factor to the output. The calibration gas for each specific flow sensor or flow meter is detailed on the product label.

B.

Installation

Caution: Do not exceed the pressure, temperature or power operating ranges detailed in the SPECIFICATIONS section of this manual. Omega Engineering shall not be liable for any damage or injury caused by incorrect operation of their products.

1. General Considerations

It is recommended that a safety shut-off valve be installed upstream

(before) of the sensor.

All wetted parts should be checked for compatibility with the gas to be used. If there are any incompatibilities e.g. highly corrosive gas, then the unit may be damaged or fail prematurely. Such damage will not be repaired under warranty.

Units should be installed in a clean, dry environment with an ambient temperature that is as stable as possible. Avoid areas with strong magnetic fields, strong air flows or excessive vibration.

Pressure drop across the FMA 3100/3300 Series is approximately 0.08 psi

(6 millibar) at 100% of the rated flow.

M-4270/0707, pg. 5 of 29

0.08

0.07

0.06

0.05

0.04

Pressure Drop vs. Percentage of the Full Scale

Rated Flow for the FMA 3100/3300 Series

6

5

4

3

0.03

0.02

0.01

2

1

0

0 10 20 30 40 50 60 70 80 90 100

0

% of Full Sca le Ra ted Flow

Typica l Pre ssure Drop

2. Mounting the Flow Sensor or Flow Meter.

The FMA 3100/3300 Series sensors have no particular orientation requirements so may be mounted in any convenient position.

It is recommended that units be fixed to a suitable substrate using the two

4-40 mounting holes provided.

Mounting View from Bottom

(mounting hardware not included with sensor)

M-4270/0707, pg. 6 of 29

3. Tubing Connections a) General

All tubing must be clean, dry and purged with clean dry air before installation of the flow sensor or flow meter.

If the gas to be used may contain particles then a filter (20 microns or less) should be installed upstream of (before) the unit.

Caution: Use only the fittings factory installed on the unit. If the fittings are removed the calibration of the unit may be effected and leaking may occur. If different fittings are required please contact Omega Engineering for assistance.

When connecting the sensor to the tubing, take care not to over-tighten the fittings or leaking may occur. b) High Flow Units (0-100 L/min and above)

It is recommended that tubing with a 0.5” (12.7 mm) outside diameter,

0.40” – 0.42” (10.16 mm – 10.67 mm) inside diameter is used. Using any other tube size may effect the calibration.

High flow units should be installed with at least a 5” (127 mm) long straight length of tube immediately before the sensor. For flow rates above 200 L/min a straight length of at least 22” (559 mm) is recommended. Using lengths shorter than these recommendations will effect the calibration of the unit.

The manufacturer is able to calibrate high flow units for specific tubing configurations. Please contact Omega Engineering for further details.

4. Electrical Connections

Caution: Incorrect wiring may cause severe damage to the unit.

Applying an AC voltage (115VAC or 230VAC) directly to the unit will cause damage. Read the following instructions carefully before making any connections. a) Overview

The FMA 3100/3300 Series provides a 0-5VDC analog output proportional to the flow rate. This output may be connected to a display, data acquisition system or voltmeter with an impedance of greater than 2.5 kΩ

(kilo ohms).

A stable D.C. power supply is required to operate the unit. The voltage and current requirements depend on the configuration of the unit. Full details may be found in the Specification section of this manual.

M-4270/0707, pg. 7 of 29

Connecting wires should be as short as possible to avoid voltage drops.

Twisted conductor cable should be used if the length of the wiring is to be longer than 2 meters.

Units are supplied with either a 6 pin mini DIN type connector (requires mating cable assembly), a 9 Pin D Sub connector or 15 Pin D Sub connector.

Caution: Cutting off the integrated connectors on the unit IS

NOT RECOMMENDED and will void the product warranty. Mating cables should be ordered for each unit.

b) Connecting To The 6 Pin Mini Din Connector

Using a suitable mating connector the pins of the integrated connector should be wired as follows:

Connecting To The Integrated 6 Pin Connector

Pin Numbers For

Integrated Connector

Pin 2 should be connected to the Positive of the power source.

Pin 6 should be connected to the Negative (Ground) of the power source.

Pin 3 provides the signal output and should be connected to the positive terminal of the display, data acquisition system or voltmeter.

PIN 1 is the signal negative (ground) and should be connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.

On the integrated 6-pin connector, PINS 4 & 5 are unused.

Caution: Do not short the output signal pins or allow them to contact the power wires at any time. DAMAGE WILL RESULT!

M-4270/0707, pg. 8 of 29

c) Connecting The 6 Pin Mini Din Connector and FMA 3000C

Cable

The two mating connectors should be pushed together and the pigtail leads wired as follows:

Connecting To The Integrated 6 Pin Connector Using A FMA 3000C Cable

The pigtail lead of the FMA 3000C cable assembly should be connected as follows:

The RED wire (Pin 2 on connector) should be connected to the Positive of the power source.

The BLACK wire (Pin 6 on the connector) should be connected to the

Negative (Ground) of the power source.

The ORANGE wire (Pin 3 on the connector) provides the signal output and should be connected to the positive terminal of the display, data acquisition system or voltmeter.

The BROWN wire (PIN 1 on the connector) is the signal negative

(ground) and should be connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.

On the FMA 3000C cable, the GREEN and YELLOW wires, if present, are unused. On the integrated 6-pin connector, PINS 4 & 5 are unused.

The wire colors above describe the pigtail leads of the 50-C-X cable assembly and may not correspond with the internal wiring of your flow sensor.

Caution: Do not short the output signal wires or allow them to contact the power wires at any time. DAMAGE WILL RESULT!

M-4270/0707, pg. 9 of 29

d) Connections For The 9 Pin D Sub Connector

Using a suitable mating connector the pins should be wired as follows:

Connecting To The Integrated 9 Pin Connector

Pin Numbers For

Integrated Connector

PIN 3 should be connected to the Positive of the power source.

PIN 4 should be connected to the Negative (Ground) of the power source.

PIN 2 provides the signal output and should be connected to the positive terminal of the display, data acquisition system or voltmeter.

PIN 8 is the signal negative (ground) and should be connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.

Pins 1, 5, 6, 7 and 9 are not used.

Caution: Do not short the output signal pins or allow them to contact the power connections at any time. DAMAGE WILL

RESULT!

M-4270/0707, pg. 10 of 29

e) Connections For The 15 Pin D Sub Connector

Using a suitable mating connector the pins should be wired as follows:

Connecting To The Integrated 15 Pin Connector

Pin Numbers For

Integrated Connector

PIN 7 should be connected to the Positive of the power source.

PIN 5 should be connected to the Negative (ground) of the power source.

PIN 2 provides the signal output and should be connected to the positive terminal of the display, data acquisition system or voltmeter.

PIN 10 is the signal negative (ground) and should be connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.

Pins 1, 3, 4, 6, 8, 9, 11, 12, 13, 14, and 15 are not used.

Caution: Do not short the output signal pins or allow them to contact the power connections at any time. DAMAGE WILL

RESULT!

M-4270/0707, pg. 11 of 29

f) Using a 0-5VDC Output Power Adapter Package.

An optional 0-5VDC Output Power Adapter Package is available for use with the FMA 3100/3300 series. This consists of a power source (115VAC or 230VAC), a connection hub and cable assembly with pig-tail (soldered wire) ends. This should be assembled as shown in the following diagram.

Assembling an FMA 3115PW Power Adapter Package

(the FMA 3230PW Power Adapter Package is similar)

A 0-5VDC analog output proportional to the flow rate may be made available by connecting the cable assembly to the connection hub. This output may then be connected to a display, data acquisition system or voltmeter with an impedance of greater than 2.5 kΩ (kilo ohms).

The RED wire of the cable assembly provides the signal output and should be connected to the positive terminal of the display, data acquisition system or voltmeter.

The COPPER/BARE wire of the cable assembly is the signal negative

(ground) and should be connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.

Caution: Do not short the output signal wires or allow them to contact the power wires at any time. DAMAGE WILL RESULT!

M-4270/0707, pg. 12 of 29

C.

Operation

1. Warm Up

Before applying power to the unit check all tubing and electrical connections. Once correct installation is verified, you may switch on the power. The unit should then be allowed to warm up for 5 minutes.

2. Verification of Zero

Flow through the unit should be stopped by sealing or capping the inlet of the sensor. It is not adequate to only stop flow by turning off the gas supply or closing a valve as there may be a leak in the system. This would give a false reading.

After 5 minutes, the zero should be stable when there is no flow through the unit. If after 10-15 minutes the output is still not zero volts (within

±0.5 volts) the unit should be adjusted as detailed in section C part 4.

It should be noted that power supply voltage variations and changes in ambient temperature can have an effect on zero readings.

3. Flow Readings

Each sensor is factory calibrated for a specific flow range and gas (or gas mixture). The calibration gas and flow range are shown on the unit’s label and calibration certificate.

By monitoring the voltage output signal it is possible to determine the flow rate of the gas. Units are configured so that an output signal of 5.0VDC is provided when the maximum flow (i.e. Full Scale flow) is passing through the unit. The output signal is linear and scaleable enabling calculation of flow rates with in the sensor’s range. For example:

For a flow range of 0-500sccm:

At 500sccm the output signal would be 5VDC

If the output signal is 3.5VDC then the flow rate would be:

500 ÷ 5 × 3.5 = 350sccm

If the maximum flow rate is exceeded non-linear and inaccurate readings will result.

Units may be used for gases other than the calibration gas. In this case a

“K Factor” would need to be applied and a corrected value calculated using the following formula:

M-4270/0707, pg. 13 of 29

Q

1

/ Q

2

= K

1

/ K

2

Q

1

= (K

Q

1

Q

2

K

1

K

2

is the flow rate of the new gas

is the flow rate of the original calibration gas

is the K factor of the new gas

is the K factor of the original calibration gas

1

/ K

2

) Q

2

If K

2

is larger than K does not exceed 5(K

1

then linear results will only be achieved if the unit

1

/ K

2

)VDC for the full scale output.

The accuracy of readings using K factors is not as good as that achieved for the calibration gas. The accuracy obtained (typically ±3% for K factors similar to the calibration gas) depends on the gas being used and the flow rate.

For a list of common K Factors see Section J.

Example 1

For a 0-200sccm unit calibrated for air the flow at 5.0VDC would be

200sccm. The K factor for air is 1. If the unit is used with Helium (K factor

1.454 relative to air) then the flow at 5VDC (i.e. the maximum flow) would be (1.454/1)200 = 290.8 sccm

Example 2

For a 0-50.0 l/min unit calibrated for Argon the flow at 5.0VDC would be

50.0l/min. The K factor for Argon is 1.45. If the unit is used with Carbon

Dioxide (K factor 0.74) then the flow rate 5.0VDC would be

(0.74/1.45)50.0 = 25.5l/min

4. Zero Adjustments

The zero should be checked as detailed in section C part 2. If an adjustment is needed the Zero Potentiometer should be carefully turned using a small flat head screwdriver until the output (VDC) becomes zero.

Caution: Do NOT adjust the Gain Potentiometer when adjusting the zero or the unit will need to be recalibrated.

Care should be taken to only make small adjustments to the zero potentiometer. If too much of an adjustment is made and difficulties are being experienced in achieving a zero reading then turn the potentiometer fully anti-clockwise and begin making small clockwise adjustments until a zero reading is obtained.

M-4270/0707, pg. 14 of 29

Making Zero Adjustments Using a Small Flathead Screwdriver

5. Recalibration

If recalibration is required please contact Omega Engineering.

D.

Maintenance and Product Care

1. General

Inlet filters should be periodically checked and cleaned or replaced as necessary.

Regularly check all electrical and process connections for damage or deterioration.

If the sensor is to be stored, keep both the inlet and outlet ports sealed.

Do not allow any liquid or moisture to enter the sensor or damage will occur.

2. Returning Units for Repair or Recalibration

To return a unit for repair or recalibration please contact the Omega

Engineering Customer Service Department. An Authorized Return (AR) number will then be issued. The AR number should then be noted on the outside of the package and on any correspondence. Further details may be found on page 28 of this manual.

M-4270/0707, pg. 15 of 29

E.

Specifications

Accuracy

(including linearity)

Repeatability

Pressure Rating

Pressure Sensitivity

Temperature Rating

Temperature

Sensitivity

Leak Integrity

Wetted Materials

O-Ring Material

Fitting Material

Recommended

Filtration

Compatible gases

0-5 VDC Output

Signal

Warm-Up Time

Integrated Display

Typical Power

Consumption

Peak Power

Consumption

Electrical Connections

Settling Time

Reliability

FMA 3100 FMA 3300 FMA 3100ST

±1.5% Full Scale*

FMA 3300ST

±0.25% Full Scale*

150 psig (10.3 bar) 500 psig (34.5 bar)

±0.02% Full Scale* per psi (per 69 mbar)

Operating Range: 5 to 55ºC

Recommended Range (for best performance) : 10 to 40ºC

Storage Range: 0 to 70ºC

±0.15% F.S.* or less per ºC

1x10 -7 sccs of He

Aluminum

304 Stainless Steel

316 Stainless Steel

303 Stainless Steel

304 Stainless Steel

316 Stainless Steel

Epoxy

Viton ® N/A

Choose from acetal, brass, or stainless steel

20 microns or less

Optional inline filters available

Clean, dry gases compatible with wetted materials

Impedance of greater than 2.5 KΩ

N/A

Less than 5 minutes

3½ digit N/A

Standard: 12 VDC @ 100 mA (12.5-15 VDC)

“E” Suffix: 24 VDC @ 80 mA (22-25 VDC)

Standard: 12 VDC @ 180 mA (12.5-15 VDC)

“E” Suffix: 24 VDC @ 160 mA (22-25 VDC)

Integrated 36” (92 mm) cable, terminated with:

Standard: 6-pin Mini-DIN male (PS/2 Style)

D1 Option: 9-pin D-Sub male

D2 Option: 15-pin D-Sub male

Typically <1 second for 97% of final value

100,000 Hours MTBF (testing ongoing)

3½ digit

89/336/EEC (EN 55011 & EN 50082-1)

73/23/EEC Low Voltage Directive

Ratings IP10 (NEMA 1)

*Specifications from 10-100% of rated flow. Linearity is best fit straight line. All calibrations performed with air unless otherwise stated on calibration certificate.

M-4270/0707, pg. 16 of 29

F.

Dimensions

ALL DIMENSIONS IN INCHES (MILLIMETERS IN BRACKETS)

FMA 3100 - 0-500 sccm Flow Ranges and Below

¼” Stainless Fittings Shown

M-4270/0707, pg. 17 of 29

FMA 3100 - 0-1000sccm Up To 0-10 L/min &

FMA 3100ST - All Ranges Up To 0-10 L/min

¼” Stainless Fittings Shown

M-4270/0707, pg. 18 of 29

FMA 3100 / 3100ST - 0-20 L/min up to 0-100 L/min

3/8” Stainless Fittings Shown

M-4270/0707, pg. 19 of 29

FMA 3100 – 0-200 L/min and 0-500 L/min

½” Stainless Fittings Shown

M-4270/0707, pg. 20 of 29

FMA 3300 – Flow Ranges Up To 0-500sccm

¼” Stainless Fittings Shown

M-4270/0707, pg. 21 of 29

FMA 3300 - 0-1000sccm Up To 0-10 L/min &

FMA 3300ST - All Ranges Up To 0-10 L/min

¼” Stainless Fittings Shown

M-4270/0707, pg. 22 of 29

FMA 3300 / 3300ST - 0-20 L/min up to 0-100 L/min

3/8” Stainless Fittings Shown

M-4270/0707, pg. 23 of 29

FMA 3100 – 0-200 L/min and 0-500 L/min

½” Stainless Fittings Shown

M-4270/0707, pg. 24 of 29

G.

Gas K Factors

Gas

Chemical

Symbol

K Factor

Acetylene C

2

H

2

0.589

Argon Ar 1.438

Butane C

4

H

10

0.260

Carbon Dioxide CO

2

0.739

Deuterium D

2

1.000

Ethylene C

2

H

4

0.598

Freon 11

Freon 12

Freon 13

CCL

3

F 0.330

CCL

2

F

2

0.354

CCLF

3

0.385

Freon 14

Freon 22

CF

4

0.420

CHCLF

2

0.460

Germane GeH

4

0.570

Helium He 1.458

Hydrogen H

2

1.011

Krypton Kr 1.440

Methane CH

4

0.721

Neon Ne 1.443

Nitric Oxide

Nitrous Oxide

NO

Nitrogen N

2

1.000

N

2

0.990

O 0.710

Oxygen O

2

0.991

Ozone O

3

0.446

Propane C

3

H

8

0.383

Sulfur Dioxide SO

2

0.690

Xenon Xe 1.437

These K Factors are given for reference only and are not intended as a recommendation of application suitability. Accuracy and response will be affected depending on the gas and flow range. Check the compatibility of all wetted materials before using any gas other than the calibration gas for the unit.

M-4270/0707, pg. 25 of 29

H.

Troubleshooting Guide

Symptom

No response

Inaccurate readings

Possible Cause Method of Correction

Unit wired incorrectly

Loose connection

Damaged connector pins

Blocked flow path

Piping leak before sensor

Insufficient power

Output load resistance too low

Flow too low for the unit

Unit damaged or faulty

Particles in flow path

Flow path obscured

Check wiring is according to

Section B5

Check all connectors and wiring

Contact Omega Engineering

Check flow path for obstructions.

Check all piping and connections.

Check the power supply output and increase if necessary

Ensure the voltmeter or data acquisition system or display has an impedance of greater than of 2.5kohm

Ensure that the flow being measured is within the capabilities of the unit

Contact Omega Engineering

Add filtration before the sensor.

Remove any debris or blockage in the flow path eg.

PTFE tape.

Unit calibrated for a different gas

Check calibration certificate and apply a “K” Factor to readings if necessary.

Gas composition is variable Contact Omega Engineering

Fittings have been changed Replace the factory installed fittings

Moisture in gas

Zero drift

Ensure gas is clean and dry

Insufficient warm-up period Allow the unit to warm-up for at least 5 minutes.

Verify the zero and adjust as necessary as explained in

Section C

The gain potentiometer has been adjusted

Unit needs recalibration

Flow too high for the unit

Contact the Omega

Engineering

Contact the Omega

Engineering

Ensure that the flow being measured is within the capabilities of the unit

M-4270/0707, pg. 26 of 29

Symptom

Inaccurate Readings

Possible Cause

Output load resistance too low

Insufficient power

Ambient temperature too high or too low

Gas temperature too high or too low

Problems with rezeroing

Unit damaged or faulty

Gas flow through unit not completely stopped

Severe fluctuations in the ambient temperature e.g. unit in direct sunlight

Unstable power supply

Insufficient warm-up period

Method of Correction

Ensure the voltmeter or data acquisition system or display has an impedance of greater than of 2.5kohm

Check the power supply output and increase if necessary

Place the unit in a suitable environment

Ensure the gas temperature is within the recommended operating range

Contact Omega Engineering

Ensure there is no flow through the unit. The easiest way to do this is to plug both the inlet and outlet.

Carry out the rezero procedure in a stable environment

Check the stability and suitability of the power source

Allow the unit to warm-up for at least 5 minutes.

M-4270/0707, pg. 27 of 29

M-4270/0707, pg. 28 of 29

M-4270/0707, pg. 29 of 29