- Home
- Domestic appliances
- Climate control
- Temperature & humidity sensors
- Vega
- VEGABAR 38
- Operating instructions
Vega VEGABAR 38 Pressure sensor with switching function Operating instructions
Add to My manuals68 Pages
advertisement
Operating Instructions
Pressure sensor with ceramic measuring cell
VEGABAR 38
Two-wire 4 … 20 mA
Document ID: 57088
2
Contents
Contents
VEGABAR 38 • Two-wire 4 … 20 mA
Contents
Safety instructions for Ex areas:
Take note of the Ex specific safety instructions for Ex applications.
These instructions are attached as documents to each instrument with Ex approval and are part of the operating instructions.
Editing status: 2022-08-25
VEGABAR 38 • Two-wire 4 … 20 mA 3
4
1 About this document
1 About this document
1.1 Function
This instruction provides all the information you need for mounting, connection and setup as well as important instructions for maintenance, fault rectification, the exchange of parts and the safety of the user. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained personnel.
The contents of this manual must be made available to the qualified personnel and implemented.
1.3 Symbols used
Document ID
This symbol on the front page of this instruction refers to the Document ID. By entering the Document ID on www.vega.com you will reach the document download.
Information, note, tip: This symbol indicates helpful additional information and tips for successful work.
Note: This symbol indicates notes to prevent failures, malfunctions, damage to devices or plants.
Caution: Non-observance of the information marked with this symbol may result in personal injury.
Warning: Non-observance of the information marked with this symbol may result in serious or fatal personal injury.
Danger: Non-observance of the information marked with this symbol results in serious or fatal personal injury.
Ex applications
This symbol indicates special instructions for Ex applications.
•
List
The dot set in front indicates a list with no implied sequence.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Disposal
This symbol indicates special instructions for disposal.
VEGABAR 38 • Two-wire 4 … 20 mA
2 For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this documentation must be carried out only by trained, qualified personnel authorised by the plant operator.
During work on and with the device, the required personal protective equipment must always be worn.
2.2 Appropriate use
The VEGABAR 38 is a pressure transmitter for process pressure and hydrostatic level measurement.
You can find detailed information about the area of application in chapter " Product description ".
Operational reliability is ensured only if the instrument is properly used according to the specifications in the operating instructions manual as well as possible supplementary instructions.
2.3 Warning about incorrect use
Inappropriate or incorrect use of this product can give rise to application-specific hazards, e.g. vessel overfill through incorrect mounting or adjustment. Damage to property and persons or environmental contamination can result. Also, the protective characteristics of the instrument can be impaired.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing regulations and directives. The instrument must only be operated in a technically flawless and reliable condition. The operator is responsi ble for the trouble-free operation of the instrument. When measuring aggressive or corrosive media that can cause a dangerous situation if the instrument malfunctions, the operator has to implement suitable measures to make sure the instrument is functioning properly.
The safety instructions in this operating instructions manual, the national installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modifications are explicitly forbidden. For safety reasons, only the accessory specified by the manufacturer must be used.
To avoid any danger, the safety approval markings and safety tips on the device must also be observed.
2.5 Installation and operation in the USA and
Canada
This information is only valid for USA and Canada. Hence the following text is only available in the English language.
VEGABAR 38 • Two-wire 4 … 20 mA 5
2 For your safety
Installations in the US shall comply with the relevant requirements of the National Electrical Code (ANSI/NFPA 70).
Installations in Canada shall comply with the relevant requirements of the Canadian Electrical Code.
6 VEGABAR 38 • Two-wire 4 … 20 mA
Scope of delivery
Scope of this operating instructions
3 Product description
3 Product description
3.1 Configuration
The scope of delivery encompasses:
• VEGABAR 38 pressure transmitter
• Information sheet " Documents and software " with:
– Instrument serial number
– QR code with link for direct scanning
• Information sheet " PINs and Codes with:
– Bluetooth access code
" (with Bluetooth versions)
• Information sheet " Access protection with:
– Bluetooth access code
– Emergency Bluetooth unlock code
– Emergency device code
" (with Bluetooth versions)
The further scope of delivery encompasses:
• Documentation
– Quick setup guide VEGABAR 38
– Ex-specific " Safety instructions " (with Ex versions)
– Radio approvals (versions with Bluetooth)
– If necessary, further certificates
Note:
Optional instrument features are also described in this operating instructions manual. The respective scope of delivery results from the order specification.
This operating instructions manual applies to the following instrument versions:
• Hardware version from 1.0.0
Software version from 1.3.0
VEGABAR 38 • Two-wire 4 … 20 mA 7
8
3 Product description
Constituent parts
Type label
3
2
1
6
5
4
Fig. 1: Components of VEGABAR 38
1 Process fitting
2 Electronics housing
3 Ventilation/pressure compensation
4 Plug connector
5 LED illuminated ring
6 Display/adjustment unit
The type label contains the most important data for identification and use of the instrument.
VEGABAR 38 • Two-wire 4 … 20 mA
3 Product description
1
2
PP - 222 329 22B s/n 12345678
3
12V...35V
4...20mA
IP66/IP67/69, TYPE 6P
MWP 33bar (3300kPa)
0...2,5bar (0...250kPa) rel
AL203, FKM, 316L
+
2
3
1
4
-
2019 www.vega.com
D-77761 SCHILTACH,
Made in Germany
4
5
123456
Fig. 2: Layout of the type label (example)
1 Order/Serial number
2 Field for approvals
3 Technical data
4 Assignment
5 Bluetooth access code
6 QR code for device documentation
6
Documents and software Move to " www.vega.com
" and enter in the search field the serial number of your instrument.
There you can find the following information about the instrument:
• Order data
Documentation
Software
Alternatively, you can find all via your smartphone:
• Scan the QR-code on the type label of the device or
Enter serial number manually in the VEGA Tools app (available free of charge in the respective stores)
Application area
Measured products
3.2 Principle of operation
VEGABAR 38 is suitable for applications in virtually all industries. It is used for the measurement of the following pressure types.
• Gauge pressure
Absolute pressure
• Vacuum
Measured products are gases, vapours and liquids.
Depending on the process fitting and measurement setup, measured products can be also viscous or contain abrasive substances.
Measured variables The VEGABAR 38 is suitable for the measurement of the following process variables:
• Process pressure
VEGABAR 38 • Two-wire 4 … 20 mA 9
3 Product description
• Level
Fig. 3: Process pressure measurement VEGABAR 38
Measuring system pressure
The sensor element is the Mini-CERTEC ® measuring cell with robust ceramic diaphragm. The process pressure deflects the ceramic diaphragm and causes a capacitance change in the measuring cell.
This capacitance change is converted into an electrical signal and outputted as measured value via the output signal.
1
2
3
Measuring system temperature
Pressure types
Recessed installation
Fig. 4: Configuration of the Mini-CERTEC ® measuring cell
1 Process diaphragm
2 Glass joint
3 Base element
A temperature sensor in the electronics of the Mini-CERTEC ture value is e.g. output via Bluetooth.
® measuring cell measures the current process temperature. The tempera-
Relative pressure : the measuring cell is open to the atmosphere.
The ambient pressure is detected in the measuring cell and compensated. It thus has no influence on the measured value.
Absolute pressure : the measuring cell contains vacuum and is encapsulated. The ambient pressure is not compensated and does hence influence the measured value.
The recessed installation is particularly suitable for applications with gases, vapours and clear liquids. The measuring cell seals are positioned laterally as well as in front.
10 VEGABAR 38 • Two-wire 4 … 20 mA
3 Product description
Front flush installation
1
2
3
4
5
Fig. 5: Recessed installation of the measuring cell (example: manometer connection G½)
1 Measuring cell
2 Lateral measuring cell seal
3 Front measuring cell seal
4 Diaphragm
5 Process fitting
The front-flush installation is particularly suitable for applications with viscous or abrasive media and for buildup.
1
2
3
4
5
Fig. 6: Front-flush installation of the measuring cell (example: thread G1½)
1 Measuring cell
2 Lateral measuring cell seal
3 Front measuring cell seal
4 Process fitting
5 Diaphragm
Front-flush installation in hygienic fitting
The front-flush, hygienic installation of the measuring cell is particu larly suitable for food applications. The front seal is installed gap-free.
3
4
5
1
2
Fig. 7: Hygienic installation of the measuring cell (example: Clamp 2")
1 Measuring cell
2 Lateral measuring cell seal
3 Process fitting
4 Front measuring cell seal
5 Diaphragm
VEGABAR 38 • Two-wire 4 … 20 mA 11
3 Product description
Local adjustment
3.3 Adjustment
The integrated display and adjustment unit is used for on-site adjustment of VEGABAR 38.
For convenient reading and operation in any installation position of the device, the housing and display and adjustment unit can be rotated without tools.
2
1
Wireless adjustment
1 Rotating the housing by 330°
1 Rotating the display and adjustment unit by 270°
Devices with integrated Bluetooth module can be adjusted wirelessly via standard adjustment tools:
• Smartphone/tablet (iOS or Android operating system)
PC/notebook (Windows operating system)
Packaging
12
2
1
3
Fig. 8: Wireless connection to standard operating devices with integrated
Bluetooth LE
1 Sensor
2 Smartphone/Tablet
3 PC/Notebook
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test based on ISO 4180.
VEGABAR 38 • Two-wire 4 … 20 mA
3 Product description
Transport
Transport inspection
The packaging consists of environment-friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out in due consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Storage
Storage and transport temperature
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
• Not in the open
Dry and dust free
• Not exposed to corrosive media
Protected against solar radiation
• Avoiding mechanical shock and vibration
• Storage and transport temperature see chapter " Supplement -
Technical data - Ambient conditions "
• Relative moisture 20 … 85 %
3.5 Accessories
The instructions for the listed accessories can be found in the download area on our homepage.
Welded socket, threaded and hygienic adapter
Welded sockets are used to connect the devices to the process.
Threaded and hygienic adapters enable simple adaptation of devices with standard threaded fittings to process-side hygiene connections.
Mounting accessories The suitable mounting accessories for VEGABAR 38 includes siphons, blocking valves and measuring instrument holders.
VEGABAR 38 • Two-wire 4 … 20 mA 13
4 Mounting
Ambient conditions
Process conditions
Permissible process pressure (MWP) - Device
4 Mounting
4.1 General instructions
The instrument is suitable for standard and extended ambient conditions acc. to DIN/EN/IEC/ANSI/ISA/UL/CSA 61010-1. It can be used indoors as well as outdoors.
Note:
For safety reasons, the instrument must only be operated within the permissible process conditions. You can find detailed information on the process conditions in chapter " Technical data " of the operating instructions or on the type label.
Hence make sure before mounting that all parts of the instrument exposed to the process are suitable for the existing process conditions.
These are mainly:
• Active measuring component
Process fitting
Process seal
Process conditions in particular are:
• Process pressure
Process temperature
Chemical properties of the medium
• Abrasion and mechanical influences
The permissible process pressure range is specified by "MWP" (Maxi mum Working Pressure) on the type label, see chapter " Structure ".
The MWP takes the element of the measuring cell and processing fitting combination with the weakest pressure into consideration and may applied permanently. The specification refers to a reference temperature of +20 °C (+68 °F). It also applies when a measuring cell with a higher measuring range than the permissible pressure range of the process fitting is installed order-related.
In addition, a temperature derating of the process fitting, e. g. with flanges, can limit the permissible process pressure range according to the respective standard.
Note:
In order to prevent damage to the device, a test pressure may only exceed the specified MWP briefly by 1.5 times at reference tempera ture. The pressure stage of the process fitting as well as the overload resistance of the measuring cell are taken into consideration here
(see chapter " Technical Data ").
Protection against moisture
Protect your instrument against moisture ingress through the following measures:
• Use a suitable connection cable (see chapter " power supply ")
Connecting to
• Tighten the cable gland or plug connector
14 VEGABAR 38 • Two-wire 4 … 20 mA
Ventilation and pressure compensation
4 Mounting
• When mounting horizontally, rotate the housing so that the cable gland or plug connector point downward
• Lead the connection cable downward in front of the cable entry or plug connector
This applies mainly to outdoor installations, in areas where high humidity is expected (e.g. through cleaning processes) and on cooled or heated vessels.
Make sure that the degree of contamination specified in chapter "
Technical data " meets the existing ambient conditions.
Ventilation and pressure compensation for VEGABAR 38 are provided by an air-permeable, moisture-blocking filter element.
Note:
In case of horizontal mounting, turn the housing so that the filter ele ment points downward after the instrument is installed. This provides better protection against buildup.
Screwing in
1
Fig. 9: Position of the filter element
1 Filter element
For effective ventilation, the filter element must always be free of buildup.
Devices with threaded fitting are screwed into the process fitting with a suitable wrench via the hexagon.
See chapter " Dimensions " for wrench size.
Warning:
The housing or the electrical connection may not be used for screwing in! Depending on the device version, tightening can cause damage, e. g. to the rotation mechanism of the housing.
Permissible process pressure (MWP) - Mounting accessory
Temperature limits
The permissible process pressure range is stated on the type label.
The instrument should only be operated with these pressures if the mounting accessory used also fulfils these values. This should be en sured by suitable flanges, welded sockets, tension rings with Clamp connections, sealings, etc.
Higher process temperatures often mean also higher ambient temperatures. Make sure that the upper temperature limits stated in chapter " Technical data " for the environment of the electronics housing and connection cable are not exceeded.
VEGABAR 38 • Two-wire 4 … 20 mA 15
4 Mounting
In gases
In vapours
2
1
Fig. 10: Temperature ranges
1 Process temperature
2 Ambient temperature
4.2 Process pressure measurement
Keep the following in mind when setting up the measuring system:
• Mount the instrument above the measuring point
Possible condensation can then drain off into the process line.
1
2
3
Fig. 11: Measurement setup for process pressure measurement of gases in pipelines
1 VEGABAR 38
2 Blocking valve
3 Pipeline
Keep the following in mind when setting up the measuring system:
• Connect via a siphon
16 VEGABAR 38 • Two-wire 4 … 20 mA
1
2
1
2
4 Mounting
3
4
3
4
In liquids
Fig. 12: Measurement setup for process pressure measurement of gases in pipelines
1 VEGABAR 38
2 Blocking valve
3 Siphon in U or circular form
4 Pipeline
A protective accumulation of water is formed through condensation in the pipe bends. Even in applications with hot steam, a medium temperature < 100 °C on the transmitter is ensured.
Keep the following in mind when setting up the measuring system:
• Mount the instrument below the measuring point
The effective pressure line is always filled with liquid and gas bubbles can bubble up to the process line.
3
2
1
Fig. 13: Measurement setup for process pressure measurement of liquids in pipelines
1 VEGABAR 38
2 Blocking valve
3 Pipeline
VEGABAR 38 • Two-wire 4 … 20 mA 17
4 Mounting
Measurement setup
4.3 Level measurement
Keep the following in mind when setting up the measuring system:
• Mount the instrument below the min. level
Do not mount the instrument close to the filling stream or emptying
• area
Mount the instrument so that it is protected against pressure shocks from the stirrer
Fig. 14: Measurement setup for level measurement
18 VEGABAR 38 • Two-wire 4 … 20 mA
Safety instructions
Voltage supply
Connection cable
M12 x 1 plug
Plug according to
ISO 4400
5 Connecting to power supply
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
• Carry out electrical connection by trained, qualified personnel authorised by the plant operator
• If overvoltage surges are expected, overvoltage arresters should be installed
Warning:
Only connect or disconnect in de-energized state.
The data for power supply are specified in chapter " Technical data ".
Note:
Power the instrument via an energy-limited circuit (power max. 100 W) acc. to IEC 61010-1, e.g.
• Class 2 power supply unit (acc. to UL1310)
SELV power supply unit (safety extra-low voltage) with suitable internal or external limitation of the output current
Keep in mind the following additional factors that influence the operat ing voltage:
• Lower output voltage of the power supply unit under nominal load
(e.g. with a sensor current of 20.5 mA or 22 mA in case of fault signal)
• Influence of additional instruments in the circuit (see load values in chapter " Technical data ")
Use cable with round cross section. Depending on the plug connection, you have to select the outer diameter of the cable respectively so that the seal effect of the cable gland is ensured.
Depending on the connection method or signal output, the device is connected with standard two, three or four-wire cable without shielding.
5.2 Connection procedure
This plug connection requires a complete confectioned cable with counter plug.
Proceed as follows:
1. Loosen the screw on the rear of the plug connector
2. Remove the plug connector and seal from VEGABAR 38
3. Remove the plug insert from the plug housing
VEGABAR 38 • Two-wire 4 … 20 mA 19
5 Connecting to power supply
1
3
2
Fig. 15: Loosen the plug insert
1 Cable gland
2 Plug insert
3 Plug housing
4. Remove approx. 5 cm of the cable mantle, strip approx. 1 cm insulation from the individual wires
5. Lead the cable through the cable gland into the plug housing
6. Connect the wire ends to the screw terminals according to the wiring plan
2
1
3
4
ISO 4400 plug with hinged cover
20
Fig. 16: Connection to the screw terminals
1 Cable gland
2 Plug housing
3 Plug insert
4 Plug seal
7. Snap the plug insert into the plug housing and insert the sensor seal
8. Plug the plug insert with seal to VEGABAR 38 and tighten the screw
The electrical connection is finished.
Proceed as follows:
1. Loosen the screw in the cover of the plug connector
2. Open the cover and remove it
3. Press the plug insert downwards
4. Loosen the screws of the strain relief and cable entry
VEGABAR 38 • Two-wire 4 … 20 mA
5 Connecting to power supply
2 3 1
3
1
2
4
Fig. 17: Loosen the plug insert
1 Plug insert
2 Strain relief
3 Cable gland
4 Plug housing
5. Remove approx. 5 cm of the cable mantle, strip approx. 1 cm insulation from the individual wires
6. Lead the cable through the cable gland into the plug housing
7. Connect the wire ends to the screw terminals according to the wiring plan
1
2
3
4
5
Fig. 18: Connection to the screw terminals
1 Cable gland
2 Cover
3 Plug housing
4 Plug insert
5 Plug seal
8. Snap the plug insert into the plug housing and insert the sensor seal
Information:
Note the correct arrangement, see illustration
9. Tighten the screws on the strain relief and cable entry
10. Hook in the cover and push onto the plug connection, tighten cover screw
11. Plug the plug insert with seal to VEGABAR 38 and tighten the screw
VEGABAR 38 • Two-wire 4 … 20 mA 21
5 Connecting to power supply
Plug according to
ISO 4400 with IDC method of termination
The electrical connection is finished.
For this plug version you can use standard cable with round wire cross-section. The inner conductors do not have to be stripped. The plug connects the conductors automatically when screwing in. Cable diameter 5.5 … 8 mm, protection IP67.
5
1 2 3 4
Plug according to
ISO 4400
Fig. 19: Connection, valve plug ISO 4400 with IDC crimping technology
1 Compression nut
2 Cable
3 Seal ring
4 Terminal insert
5 Plug housing
5.3 Wiring plan - Two-wire 4 … 20 mA
+
1
3
2 1
-
Fig. 20: Wiring plan - two-wire 4 … 20 mA - plug according to ISO 4400
1 Voltage supply and signal output
2
3
Contact, plug connector Function/Polarity
1 Voltage supply, signal output/+
Voltage supply, signal output/-
Free
Electrically connected with metal housing
22 VEGABAR 38 • Two-wire 4 … 20 mA
M12 x 1 plug
5 Connecting to power supply
+
2
3
1
4
1
-
Fig. 21: Wiring plan - two-wire 4 … 20 mA - M12 x 1 plug
1 Voltage supply and signal output
3
4
Contact, plug connector Function/Polarity
1 Voltage supply, signal output/+
2 Voltage supply, signal output/-
Voltage supply, signal output/- 1)
Free
5.4 Switch-on phase
After switching on, the device first carries out a self-check:
• Internal check of the electronics
The output signal jumps to the set fault current
The current measured value is then output on the signal cable.
1) Internal bridge between contact 2 and 3
VEGABAR 38 • Two-wire 4 … 20 mA 23
6 Access protection
Bluetooth access code
Emergency Bluetooth unlock code
Device code
Emergency device code
6 Access protection
6.1 Bluetooth radio interface
Devices with a Bluetooth radio interface are protected against unwanted access from outside. This means that only authorized persons can receive measured and status values and change device settings via this interface.
A Bluetooth access code is required to establish Bluetooth communication via the adjustment tool (smartphone/tablet/notebook).
This code must be entered once when Bluetooth communication is established for the first time in the adjustment tool. It is then stored in the adjustment tool and does not have to be entered again.
The Bluetooth access code is individual for each device. It is printed on the device housing with Bluetooth. In addition, it is supplied with the device in the information sheet " PINs and Codes " In addition, the
Bluetooth access code can be read out via the display and adjustment unit, depending on the device version.
The Bluetooth access code can be changed by the user after the first connection is established. If the Bluetooth access code is entered incorrectly, the new entry is only possible after a waiting period has elapsed. The waiting time increases with each further incorrect entry.
The emergency Bluetooth access code enables Bluetooth communication to be established in the event that the Bluetooth access code is no longer known. It can't be changed. The emergency Bluetooth access code can be found in information sheet " Access protection ".
If this document is lost, the emergency Bluetooth access code can be retrieved from your personal contact person after legitimation.
The storage and transmission of Bluetooth access codes is always encrypted (SHA 256 algorithm).
6.2 Protection of the parameterization
The settings (parameters) of the device can be protected against unwanted changes. The parameter protection is deactivated on delivery, all settings can be made.
To protect the parameterization, the device can be locked by the user with the aid of a freely selectable device code. The settings (parameters) can then only be read out, but not changed. The device code is also stored in the adjustment tool. However, unlike the Bluetooth access code, it must be re-entered for each unlock. When using the adjustment app or DTM, the stored device code is then suggested to the user for unlocking.
The emergency device code allows unlocking the device in case the device code is no longer known. It can't be changed. The emergency device code can also be found on the supplied information sheet " Access protection ". If this document is lost, the emergency device code can be retrieved from your personal contact person after legitimation.
24 VEGABAR 38 • Two-wire 4 … 20 mA
6 Access protection
The storage and transmission of the device codes is always encrypted (SHA 256 algorithm).
6.3 Storing the codes in myVEGA
If the user has a " myVEGA " account, then the Bluetooth access code as well as the device code are additionally stored in his account under
" PINs and Codes ". This greatly simplifies the use of additional adjust ment tools, as all Bluetooth access and device codes are automatically synchronized when connected to the " myVEGA " account
VEGABAR 38 • Two-wire 4 … 20 mA 25
7 Set up with the integrated display and adjustment unit
Function
7 Set up with the integrated display and adjustment unit
7.1 Adjustment system
The instrument is operated via the three keys of the integrated display and adjustment unit. The respective menu items are shown on the LC display. You can find the function of the individual keys in the below overview.
Certain settings are only possible to a limited extent or not possible with the integrated display and adjustment unit. For these settings, we recommend using the adjustment app or PACTware with corresponding DTM.
Display and adjustment elements
1
Key functions
26
2
Fig. 22: Integrated display and adjustment unit
1 LC display
2 Adjustment keys
Key
[•]
[+]
[-]
[+] and [-] simultaneously
Function
Entry to the menu level
Jump to selected menu item
Edit parameter
Select editing position
Save value
Switching between the individual measured value windows
Navigation in the menu items, forwards
Change parameter values upwards
Switching between the individual measured value windows
Navigation in the menu items, backwards
Change parameter values downwards
Jump to next higher menu
Interrupt input
VEGABAR 38 • Two-wire 4 … 20 mA
Time functions
Measured value indication
7 Set up with the integrated display and adjustment unit
When the [+] and [-] keys are pressed quickly, the edited value, or the cursor, changes one value or position at a time. If the key is pressed longer than 1 s, the value or position changes continuously.
Simultaneous pressing of the [+] and [-] keys causes a return to the measured value indication.
Approx. 60 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not confirmed with [Ο] will not be saved.
7.2 Measured value and menu item display
Measured value images are displayed according to the following presentation:
1
0.3000 bar
22.0°C
Sensor 12345678
2
3
4
Menu item display
Fig. 23: Indication of measured value with additional displays
1 Measured value as bargraph
2 Measured value as digital value with unit
3 Measuring cell temperature
4 Sensor-TAG
The menu items are displayed according to the following diagram:
1
SPN
1.0000 bar
SPAN
2
3
Fig. 24: Display menu item
1 Menu item code acc. to VDMA 24574-1
2 Actual parameter value
3 Menu item name
VEGABAR 38 • Two-wire 4 … 20 mA 27
7 Set up with the integrated display and adjustment unit
Selection language
Zero
7.3 Parameter adjustment
7.3.1 Main menu
With the first setup, the instrument offers you a selection of the menu languages. The selection you are making here can be changed any time under " Extended functions ", " Menu language ".
The menu item Zero (initial value) defines the pressure value at the output current 4 mA.
Information:
The zero adjustment has no influence on the value of the span adjust ment.
Span
Menu item code:
• ZEO
Parameter:
• Pressure value
The menu item Span (final value) defines the pressure value at the output current 20 mA.
Extended functions, diagnosis
Menu item code:
• SPN
Parameter:
• Pressure value
These menu items allow access to the menu areas " Extended functions " or " Diagnosis ".
28 VEGABAR 38 • Two-wire 4 … 20 mA
7 Set up with the integrated display and adjustment unit
Reaction when malfunction occurs
Damping
Offset correction
Menu item code:
•
EF
DIA
7.3.2 Extended functions
In this menu item you define the behaviour of the current output in the event of failures.
Menu item code:
• FER
Parameter:
•
≤ 3.6 mA
≥ 21 mA
To damp process-dependent measured value fluctuations, set an integration time in this menu item.
Due to the set damping, the 4 … 20 mA output as well as the switching output react in case of a sudden increased of the measured variable with a time-delayed slope curve.
Menu item code:
• DAM
Parameter:
• Time value
The installation position of the device can shift the measured value minimally (offset). The offset correction compensates this measured value shift. The measured value that should currently be displayed is entered (manual offset correction). With relative pressure transmitters, an automatic offset to 0.0000 bar can alternatively be carried out.
Note:
With automatic offset correction, the current measured value must not be influenced by product coverage or static pressure.
VEGABAR 38 • Two-wire 4 … 20 mA 29
7 Set up with the integrated display and adjustment unit
Accept value
The position correction can be repeated as often as necessary. However, if the sum of the corrective values exceeds 20 % of the nominal measuring range, then no position correction is possible.
Menu item code:
• OFS
Parameter:
• Pressure value
In this menu item ( live adjustment ) you can accept the current measured value as the value for the 4 mA adjustment (LRV) or the 20 mA adjustment (URV). 2) .
Display lighting
Pressure unit
Menu item code:
• LRV
• URV
Parameter:
• Pressure value
In this menu item you switch the background lighting for the display off or on.
Menu item code:
• DIS
Parameter
•
On
Off
Note:
When the background lighting is on, the 360° status display is deactivated or vice versa.
In this menu item the adjustment unit of the device is defined. The se lection made determines the displayed unit in the menu items " Zero/
Span " and " Offset correction " as well as " Accept value ".
30
2) LRV: Lower Range Value, URV: Upper Range Value
VEGABAR 38 • Two-wire 4 … 20 mA
7 Set up with the integrated display and adjustment unit
Unit temperature
Menu language
Bluetooth access code
Menu item code:
• UNI
The following units are available: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, mmH
2
O, inH
2
O
In this menu item the temperature unit of the device is defined. The selection made determines the unit for the measuring cell temperature shown on the display.
Menu item code:
• TMP
Parameter:
•
°C
°F
This menu item enables the setting of the requested national language for the display.
Menu item code:
German, English, French, Spanish, Portuguese, Italian, Dutch, Russian, Chinese, Turkish
In this menu item, you can change the factory-preset Bluetooth access code to your personal Bluetooth access code.
Note:
The individual preset Bluetooth access code of the device can be found on the supplied information sheet " PINs and Codes ". If this is changed by the user and is no longer available, access is only possible via the emergency Bluetooth unlock code on the information sheet " Emergency unlock codes " also supplied.
Menu item code:
• BT
VEGABAR 38 • Two-wire 4 … 20 mA 31
7 Set up with the integrated display and adjustment unit
Protection of the parameterization
In this menu item you safeguard the sensor parameters by entering a 6-digit device code against unauthorized or unintentional modifica tions.
-
Reset
With protected parameter adjustment, the individual menu items can be selected and displayed, however the parameters can no longer be modified.
The sensor operation can also be enabled in any menu item by entering the device code. The parameter adjustment remains open until you return to the measured value display. This takes place automatically after 60 min.
Menu item code:
• COD
Parameter:
• Numerical value
Note:
The factory set device code is " 000000 ". If this is changed by the user and is no longer available, access is only possible via the emergency device unlock code on the information sheet " Emergency unlock codes " also supplied.
Note:
With protected parameter adjustment, adjustment via the VEGA Tools app as well as PACTware/DTM and other systems is also blocked.
During a reset, parameter settings made by the user are reset to the values of the basic setting or the delivery status (see chapter " Menu overview ") 3) .
32
Menu item code:
• RES
Parameter:
• Basic settings
4 Language and Bluetooth access code are not reset.
VEGABAR 38 • Two-wire 4 … 20 mA
Status
7 Set up with the integrated display and adjustment unit
• Delivery status 4)
Basic settings: Resets the parameter settings to the default values of the respective device. The order-related settings are not transferred to the current parameters after this reset.
Delivery status: Resets the parameter settings to the delivery status.
Information:
The current status of the access protection, the Bluetooth access code and the device code are not reset.
7.3.3 Diagnostics
In this menu item, the device status is displayed.
Parameter modification counter
Menu item code:
• STA
In the event of an error, the error code, e.g. F017, and an error description, e.g. " Adjustment span too small " are displayed.
This menu item displays the number of parameter changes made.
Peak indicator
Menu item code:
• PCO
In this menu item, the min. and max. values for pressure, measuring cell temperature and electronics temperature are displayed.
Menu item code:
• LO
VEGABAR 38 • Two-wire 4 … 20 mA 33
7 Set up with the integrated display and adjustment unit
• HI
Information:
To reset the peak indicator, the VEGA Tools app or PACTware/DTM is required.
Sensor information This menu item displays the hardware and software status as well as the serial number of the device.
Menu item code:
• INF
Parameter:
•
HW
SW
SN
Extended functions -
Simulation
In this menu item you simulate switching states of the transistor outputs or current values of the 4 … 20 mA output. This allows the signal path to be tested, e.g. via downstream display instruments or the input card of the control system. The simulation values are: Pressure, current, switching status.
Menu item code:
• SIM
Parameter:
•
Numerical value
Open
Closed
Note:
Without manual deactivation, the sensor terminates the simulation automatically after 60 minutes.
34 VEGABAR 38 • Two-wire 4 … 20 mA
System requirements
Connecting
Authenticate
Enter Bluetooth access code
8 Setup with smartphone/tablet (Bluetooth)
8 Setup with smartphone/tablet (Bluetooth)
8.1 Preparations
Make sure that your smartphone/tablet meets the following system requirements:
• Operating system: iOS 8 or newer
Operating system: Android 5.1 or newer
• Bluetooth 4.0 LE or newer
Download the VEGA Tools app from the " Apple App Store ", " Google Play Store " or " Baidu Store " to your smartphone or tablet.
8.2 Connecting
Start the adjustment app and select the function " Setup ". The smartphone/tablet searches automatically for Bluetooth-capable instruments in the area.
The message " Connecting … " is displayed.
The devices found are listed and the search is automatically continued.
Select the requested instrument in the device list.
As soon as the Bluetooth connection to a device is established, the
LED display of the device in question flashes blue 4 times.
When establishing the connection for the first time, the operating tool and the sensor must authenticate each other. After the first correct authentication, each subsequent connection is made without a new authentication query.
For authentication, enter the 6-digit Bluetooth access code in the next menu window. You can find the code on the outside of the device housing and on the information sheet " Pins and Codes " in the device packaging.
Connected
Fig. 25: Enter Bluetooth access code
Note:
If an incorrect code is entered, the code can only be entered again after a delay time. This time gets longer after each incorrect entry.
The message " Waiting for authentication " is displayed on the smartphone/tablet.
After connection, the sensor adjustment menu is displayed on the respective adjustment tool.
VEGABAR 38 • Two-wire 4 … 20 mA 35
8 Setup with smartphone/tablet (Bluetooth)
Change device code
Enter parameters
If the Bluetooth connection is interrupted, e.g. due to a too large distance between the two devices, this is displayed on the adjustment tool. The message disappears when the connection is restored.
Parameter adjustment of the device is only possible if the parameter protection is deactivated. When delivered, parameter protection is deactivated by default and can be activated at any time.
It is recommended to enter a personal 6-digit device code. To do this, go to menu " Extended functions ", " Access protection ", menu item "
Protection of the parameter adjustment ".
8.3 Sensor parameter adjustment
The sensor adjustment menu is divided into two halves:
On the left you'll find the navigation section with the menus " Setup ", "
Display ", " Diagnosis " and others.
The selected menu item, recognisable by the colour change, is displayed in the right half.
36
Fig. 26: Example of an app view - Setup measured values
VEGABAR 38 • Two-wire 4 … 20 mA
System requirements
Activate Bluetooth connection
Connecting
Authenticate
Enter Bluetooth access code
9 Setup with PC/notebook (Bluetooth)
9 Setup with PC/notebook (Bluetooth)
9.1 Preparations
Make sure that your PC/notebook meets the following system requirements:
• Operating system Windows 10
DTM Collection 10/2020 or newer
• Bluetooth 4.0 LE or newer
Activate the Bluetooth connection via the project assistant.
Note:
Older systems do not always have an integrated Bluetooth LE. In these cases, a Bluetooth USB adapter is required. Activate the
Bluetooth USB adapter using the Project Wizard.
After activating the integrated Bluetooth or the Bluetooth USB adapter, devices with Bluetooth are found and created in the project tree.
9.2 Connecting
Select the requested device for the online parameter adjustment in the project tree.
As soon as the Bluetooth connection to a device is established, the
LED display of the device in question flashes blue 4 times.
When establishing the connection for the first time, the operating tool and the device must authenticate each other. After the first correct authentication, each subsequent connection is made without a new authentication query.
For authentication, enter in the next menu window the 6-digit
Bluetooth access code:
Fig. 27: Enter Bluetooth access code
VEGABAR 38 • Two-wire 4 … 20 mA 37
9 Setup with PC/notebook (Bluetooth)
Connected
Change device code
Prerequisites
You can find the code on the outside of the device housing and on the information sheet " PINs and Codes " in the device packaging.
Note:
If an incorrect code is entered, the code can only be entered again after a delay time. This time gets longer after each incorrect entry.
The message " Waiting for authentication " is displayed on the PC/ notebook.
After connection, the device DTM appears.
If the connection is interrupted, e.g. due to a too large distance between device and adjustment tool, this is displayed on the adjustment tool. The message disappears when the connection is restored.
Parameter adjustment of the device is only possible if the parameter protection is deactivated. When delivered, parameter protection is deactivated by default and can be activated at any time.
It is recommended to enter a personal 6-digit device code. To do this, go to menu " Extended functions ", " Access protection ", menu item "
Protection of the parameter adjustment ".
9.3 Parameter adjustment
For parameter adjustment of the instrument via a Windows PC, the configuration software PACTware and a suitable instrument driver
(DTM) according to FDT standard are required. The latest PACTware version as well as all available DTMs are compiled in a DTM Collection. The DTMs can also be integrated into other frame applications according to FDT standard.
38
Fig. 28: Example of a DTM view - Adjustment current output
VEGABAR 38 • Two-wire 4 … 20 mA
Measured value indication
10 Menu overview
10 Menu overview
10.1 Display and adjustment unit (on site)
Menu item
Measured value indication
Measured value window 1
Pressure
Measured value window 2
Pressure, measuring cell temperature, sensor TAG
Main menu
Menu item
Zero 4 mA
Span 20 mA
Extended functions
Diagnostics
Code acc. to
VDMA 24574-1
ZEO
SPN
EF
DIA
Basic settings
-
-
Measuring range begin
Measuring range end
Extended functions
Menu item
Damping
Offset correction
Reaction when malfunctions occur
Accept value 4 mA
Accept value 20 mA
Display lighting
Pressure unit
Unit temperature
Menu language
Bluetooth access code
Protection of the parameterization
Reset
UNI
TMP
LG
BT
COD
RES
Code acc. to
VDMA 24574-1
DAM
OFS
FER
LRV
URV
DIS
-
-
Basic settings
1 s
≤ 3.6 mA
On mbar
°C
English
On
-
Off
Diagnostics
Menu item
Status
Parameter modification counter
Min. value peak indicator
Max. value peak indicator
Sensor information
Code acc. to
VDMA 24574-1
STA
PCO
LO
HI
INF, HW, SW -
Delivery status
-
-
Last values
VEGABAR 38 • Two-wire 4 … 20 mA 39
10 Menu overview
Menu item Code acc. to
VDMA 24574-1
SIM -
Delivery status
Simulation
10.2 VEGA Tools app and DTM (Bluetooth)
Start image (app)
Menu item
Start image
Device information
Device name, software version, serial number
Actual measured values
Pressure, output current, measuring cell temperature, electronics temperature, sensor TAG
Device status
OK, error indication
Basic functions
Menu item
Measurement loop name
Parameter (code acc. to VD-
MA 24574-1)
Measurement loop name
Editing section
19 alphanumeric characters/ special characters
Basic setting
Sensor
Current output
360° status indication (acc. to NAMUR NE 107)
360° status display (free signalling)
Zero 4 mA (ZEO)
Span 20 mA (SPN)
Reaction in case of fault (FER)
Brightness (LED)
Signalling
Brightness (LED)
Signalling
Fault
Flashing
Operating status 5)
Upper limit
Colour selection
Flashing
Measuring range begin … measuring range end
≤ 3.6 mA, ≥ 21 mA
0 %, 10 %, 20 % … 100 %
Acc. to NAMUR NE 107
0 %, 10 %, 20 % … 100 %
Acc. to NAMUR NE 107, switching output, free signalling
Red, yellow, green …, individual colour selection, no signalling
Yes, no
Operating states 1, 2, 3, 4, 5
Measuring range begin minus
20 % … measuring range end plus 20 %
Red, yellow, green …, individual colour selection, no signalling
Yes, no
No
1
0.000 bar
Green
No
Measuring range begin
Measuring range end
≤ 3.6 mA
100 %
Acc. to NA-
MUR NE 107
100 %
Acc. to NA-
MUR NE 107
Red
Note:
With the 360° status indication switched on, the lighting of the display and adjustment unit is deactivated or vice versa.
40
5) Signalling of process pressure ranges by colour and flashing
VEGABAR 38 • Two-wire 4 … 20 mA
Extended functions
Menu item Parameter (code acc. to VD-
MA 24574-1)
Damping
Offset correction
Integration time (DAM)
Activate thermo-shock suppression
Offset correction (OFS)
Adjustment with medium Apply min. pressure on the sensor
Indication
Apply max. pressure on the sensor
Lighting (DIS) 6)
Menu language (LG)
Presentation
Units
Access protection
Pressure unit (UNI)
Temperature (TMP)
Bluetooth access code
Editing section
0 … 9.0 s
Yes, no
Execute, automatic correction
Accept 4 mA (LRV)
Accept 20 mA (URV)
Measured value indication 1, measured value indication 2 mbar, bar, Pa, kPa, MPa, psi, mmH2O, mmHg, inH2O, inHg
°C, °F
Protection of the parameterization
Reset
10 Menu overview
-
Basic setting
Reset
Diagnostics
Menu item
Status
0 s
No
0.000 bar
On
German
Measured value indication 1 bar
-
°C
Device-specific access code
Deactivated
Peak indicator
Measured values
Measured value memory (DTM)
Simulation
Sensor information
Sensor features (DTM)
Parameter
Device status, parameter change counter
Process pressure, measuring cell temperature, electronics temperature
Measured values, output, additional measured values
Indication
Actual values
Actual values, min. values, max. values
Trend curve
Pressure, current output
Sensor characteristics
Actual values pressure, current output, electronics temperature, measuring cell temperature
Last values
Simulated values
Device name, serial number, hardware version, software version, factory calibration date, Device Revision, measuring range begin, measuring range end
Features of the instrument version
6) When the display lighting is on, the 360° status display is deactivated or vice versa.
VEGABAR 38 • Two-wire 4 … 20 mA 41
11 Diagnostics and servicing
Maintenance
Precaution measures against buildup
Cleaning
Reaction when malfunction occurs
Causes of malfunction
11 Diagnostics and servicing
11.1 Maintenance
If the device is used properly, no special maintenance is required in normal operation.
In some applications, product buildup on the diaphragm can influence the measuring result. Depending on the sensor and application, take precautions to ensure that heavy buildup, and especially a hardening thereof, is avoided.
The cleaning helps that the type label and markings on the instrument are visible.
Take note of the following:
• Use only cleaning agents which do not corrode the housings, type
• label and seals
Use only cleaning methods corresponding to the housing protection rating
11.2 Rectify faults
The operator of the system is responsible for taking suitable measures to rectify faults.
The device offers maximum reliability. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
• Sensor
Process
• Voltage supply
Signal processing
Fault rectification
Reaction after fault rectification
24 hour service hotline
The first measures are:
• Evaluation of fault messages
Checking the output signal
Treatment of measurement errors
A smartphone/tablet with the adjustment app or a PC/notebook with the software PACTware and the suitable DTM offer you further com prehensive diagnostic possibilities. In many cases, the causes can be determined in this way and the faults eliminated.
Depending on the reason for the fault and the measures taken, the steps described in chapter " Setup " must be carried out again or must be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550 .
The hotline is also available outside normal working hours, seven days a week around the clock.
42 VEGABAR 38 • Two-wire 4 … 20 mA
11 Diagnostics and servicing
Since we offer this service worldwide, the support is provided in
English. The service itself is free of charge, the only costs involved are the normal call charges.
4 … 20 mA signal
Error Cause
4 … 20 mA signal not stable Fluctuating measured value
4 … 20 mA signal missing Electrical connection faulty
Voltage supply missing
Current signal greater than
22 mA, less than 3.6 mA
11.3 Diagnosis, fault messages
Connect a multimeter in the suitable measuring range according to the wiring plan. The following table describes possible errors in the current signal and helps to eliminate them:
Operating voltage too low, load resistance too high
Sensor electronics defective
Rectification
Set damping
Check connection, correct, if necessary
Check cables for breaks; repair if necessary
Check, adapt if necessary
Replace device or send in for repair depending on device version
LED illuminated ring
Colour 8)
Green
Yellow
Red
The 360° status indication on the device (see chapter " Configuration ") shows the following:
• Device status
Operating status 7)
This enables simple on-site diagnosis without tools, see the following table:
Permanent light voltage supply on, operation without failure voltage supply on, operation with failure
Flashing
Message acc. to NE 107 " Maintenance required " available
-
Message acc. to to NE 107 " Function check ", " Out of specification " or " Simulation state " is displayed
Note:
For devices with M12 x 1 stainless steel plug, the 360° status indication is not available.
11.4 Status messages according to NE 107
The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650. In addition to the status messages in the following tables there are more detailed error messages available under the menu item " Diagnostics " via the respective adjustment module.
7) Signalling of process pressure ranges by colour and flashing, adjustable via
VEGA Tools app or PACTware/DTM.
8) Delivery status; adjustable via VEGA Tools app or PACTware/DTM
VEGABAR 38 • Two-wire 4 … 20 mA 43
11 Diagnostics and servicing
Status messages The status messages are divided into the following categories:
• Failure
Function check
Out of specification
• Maintenance required and explained by pictographs:
1 2
Fig. 29: Pictographs of the status messages
1 Failure - red
2 Out of specification - yellow
3 Function check - orange
4 Maintenance required - blue
3 4
Malfunction (Failure):
Due to a malfunction in the instrument, a fault signal is output.
This status message is always active. It cannot be deactivated by the user.
Function check:
The instrument is being worked on, the measured value is temporarily invalid (for example during simulation).
This status message is inactive by default.
Out of specification:
The measured value is unreliable because an instrument specification was exceeded (e.g. electronics temperature).
This status message is inactive by default.
Maintenance required:
Due to external influences, the instrument function is limited. The measurement is affected, but the measured value is still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).
This status message is inactive by default.
Failure
Code
Text message
F013 no measured value available
F017
Adjustment span too small
F036 no operable sensor software
Cause
Failed or interrupted software update
Rectification
Hardware error in the area of the measuring cell
Adjustment not within specification
Send instrument for repair
Change adjustment
Repeat software update
44 VEGABAR 38 • Two-wire 4 … 20 mA
11 Diagnostics and servicing
Code
Text message
F080
General software error
F110
Switching points too close together
F111
Switching points interchanged
F260
Error in the calibration
F261
Error in the instrument settings
Cause
General software error
Selected switching points too close together
Switching point 1 is smaller than switching point 2
Checksum error in the calibration values
Checksum error in the configuration values
Rectification
Restart
Increase the distance between the switching points
Select switching point 1 to greater than switching point 2
Send instrument for repair
Carry out a reset
Function check
Code
Text message
C700
Simulation active
Cause
A simulation is active
Rectification
Finish simulation
Wait for the automatic end after
60 mins.
Out of specification
Code
Text message
S600
Impermissible electronics temperature
S604
Switching output overloaded
Maintenance
Code
Text message
M504
Error at a device interface
M510
No communication with the main controller
Cause
Temperature of the electronics in the non-specified range
Cause
Rectification
Check ambient temperature
Insulate electronics
Overload or short circuit at output 1 or 2 Electrical connection, check load resistance
Rectification
Interference of the internal communication to Bluetooth
Fault in internal communication with the display
Restart
Send instrument for repair
Restart
Send instrument for repair
11.5 Software update
The device software is updated via Bluetooth.
The following components are required:
VEGABAR 38 • Two-wire 4 … 20 mA 45
11 Diagnostics and servicing
• Instrument
Voltage supply
• PC/notebook with PACTware/DTM and Bluetooth USB adapter
Current instrument software as file
You can find the current instrument software as well as detailed infor mation on the procedure in the download area of our homepage.
Caution:
Instruments with approvals can be bound to certain software versions.
Therefore make sure that the approval is still effective after a software update is carried out.
You can find detailed information in the download area on our home page.
11.6 How to proceed if a repair is necessary
You can find an instrument return form as well as detailed information about the procedure in the download area of our homepage. By doing this you help us carry out the repair quickly and without having to call back for needed information.
Proceed as follows in case of repair:
• Print and fill out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
• sheet outside on the packaging
Ask the agency serving you to get the address for the return shipment. You can find the agency on our homepage.
46 VEGABAR 38 • Two-wire 4 … 20 mA
12 Dismount
12 Dismount
12.1 Dismounting steps
To remove the device, carry out the steps in chapters " Mounting " and
" Connecting to power suplly " in reverse.
Warning:
When dismounting, pay attention to the process conditions in vessels or pipelines. There is a risk of injury, e.g. due to high pressures or temperatures as well as aggressive or toxic media. Avoid this by taking appropriate protective measures.
12.2 Disposal
Pass the instrument on to a specialised recycling company and do not use the municipal collecting points.
Remove any batteries in advance, if they can be removed from the device, and dispose of them separately.
If personal data is stored on the old device to be disposed of, delete it before disposal.
If you have no way to dispose of the old instrument properly, please contact us concerning return and disposal.
VEGABAR 38 • Two-wire 4 … 20 mA 47
13 Certificates and approvals
48
13 Certificates and approvals
13.1 Radio licenses
Bluetooth
The Bluetooth radio module in the device has been tested and approved according to the current edition of the applicable countryspecific norms or standards.
The confirmations as well as regulations for use can be found in the document " Radio licenses " supplied or on our homepage.
13.2 Approvals for Ex areas
Approved versions for use in hazardous areas are available or in preparation for the device or the device series.
You can find the relevant documents on our homepage.
13.3 Approvals as overfill protection
Approved versions for use as part of an overfill protection system are available or in preparation for the device or the device series.
The corresponding approvals can be found on our homepage.
13.4 Food and pharmaceutical certificates
Versions for use in the food and pharmaceutical industries are available or in preparation for the device or the device series.
The corresponding certificates can be found on our homepage.
13.5 Conformity
The device complies with the legal requirements of the applicable country-specific directives or technical regulations. We confirm con formity with the corresponding labelling.
The corresponding conformity declarations can be found on our homepage.
Due to the design of its process fittings, the device does not subject of EU pressure device directive if it is operated at process pressures
≤ 200 bar.
13.6 NAMUR recommendations
NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in field instrumentation.
The device fulfils the requirements of the following NAMUR recom mendations:
• NE 21 – Electromagnetic compatibility of equipment
NE 43 – Signal level for fault information from measuring transduc-
• ers
NE 53 – Compatibility of field devices and display/adjustment components
VEGABAR 38 • Two-wire 4 … 20 mA
13 Certificates and approvals
• NE 107 – Self-monitoring and diagnosis of field devices
For further information see www.namur.de.
13.7 Environment management system
Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system with the goal of continuously improving company environmental protection. The environment management system is certified according to DIN EN ISO 14001.
Help us to meet these requirements and observe the environmental instructions in the chapters " Packaging, transport and storage ", "
Disposal " of this operating instructions.
VEGABAR 38 • Two-wire 4 … 20 mA 49
14 Supplement
14 Supplement
14.1 Technical data
Note for approved instruments
The technical data in the respective safety instructions which are included in delivery are valid for approved instruments (e.g. with Ex approval). These data can differ from the data listed herein, for example regarding the process conditions or the voltage supply.
All approval documents can be downloaded from our homepage.
Materials and weights
Materials, wetted parts
Process fitting
Diaphragm
Measuring cell seal
316L, PVDF, Duplex steel (1.4462), PEEK
Sapphire-ceramic ® (> 99.9 % Al
2
O
3
ceramic)
FKM (VP2/A), EPDM (A+P 70.10-02), FFKM (Perlast
G74S)
Seal for process fitting (in the scope of delivery)
Ʋ Thread G½ (EN 837), thread G½ inside G¼ (ISO 228-1), thread G½ inside 11.4 mm (ISO 228-1), thread
M20 x 1.5 (EN 837), thread G1½
(DIN 3852-A)
Klingersil C-4400
Ʋ Further thread versions
Materials for applications in foodstuffs
Depending on the configuration
Surface quality, hygienic fittings, typ.
Ʋ Process fitting
Ʋ Ceramic diaphragm
Materials, non-wetted parts
Electronics housing
Illuminated ring
M12 x 1 plug connector
Ʋ Contact support
Ʋ Contacts
Weight
R
R a a
< 0.76 µm
< 0.5 µm
316L und PBT/PC
PC
PBT/PC
CuZn, nickel layer and 0.8 µm gold-plated approx. 0.25 kg (0.55 lbs)
Torques
Max. torque for process fitting (examples)
Ʋ Thread G½, inside G¼ (ISO 228-1),
PVDF
Ʋ Clamp
Ʋ Thread ½ NPT inside 6 mm, G1
(DIN 3852-E) PEEK, G1½ (DIN 3852-
A-B) PEEK
Ʋ Varivent
5 Nm (3.688 lbf ft)
5/10 Nm (3.688/7.376 lbf ft)
10 Nm (7.376 lbf ft)
20 Nm (14.75 lbf ft)
50 VEGABAR 38 • Two-wire 4 … 20 mA
14 Supplement
Ʋ Thread G½ (ISO 228-1), G¾
(DIN 3852-E), M30 x 1.5, Ingold, NPT connections
Ʋ SMS, collar socket DIN 11851,
DIN 11864-1, Form A
Ʋ Thread G½ (EN 837), G½ (DIN 3852-
A), G1 (ISO 228-1), G1½ (DIN 3852-
A)
Ʋ Thread G1 with conus
30 Nm (22.13 lbf ft)
40 Nm (29.50 lbf ft)
50 Nm (36.88 lbf ft)
100 Nm (73.76 lbf ft)
Input variable
The specifications are only an overview and refer to the measuring cell. Limitations due to the material and version of the process fitting as well as the selected pressure type are possible. The specifications on the nameplate apply. 9)
Nominal measuring ranges and overload capability in bar/kPa
Nominal range
Gauge pressure
0 … +0.1 bar/0 … +10 kPa
0 … +0.4 bar/0 … +40 kPa
0 … +1 bar/0 … +100 kPa
0 … +2.5 bar/0 … +250 kPa
0 … +5 bar/0 … +500 kPa
0 … +10 bar/0 … +1000 kPa
0 … +25 bar/0 … +2500 kPa
0 … +60 bar/0 … +6000 kPa
-0.05 … +0.05 bar/-5 … +5 kPa
-0.2 … +0.2 bar/-20 … +20 kPa
-0.5 … +0.5 bar/-50 … +50 kPa
-1 … 0 bar/-100 … 0 kPa
-1 … +1.5 bar/-100 … +150 kPa
Absolute pressure
0 … 0.1 bar/0 … 10 kPa
0 … 1 bar/0 … 100 kPa
0 … 2.5 bar/0 … 250 kPa
0 … +5 bar/0 … +500 kPa
0 … 10 bar/0 … 1000 kPa
0 … 25 bar/0 … 2500 kPa
0 … 60 bar/0 … 6000 kPa
Overload capacity, max. pressure Overload capacity, min. pressure
+15 bar/+1500 kPa
+30 bar/+3000 kPa
+35 bar/+3500 kPa
+50 bar/+5000 kPa
+65 bar/+6500 kPa
+90 bar/+9000 kPa
+130 bar/+13000 kPa
+200 bar/+20000 kPa
+15 bar/+1500 kPa
+20 bar/+2000 kPa
+35 bar/+3500 kPa
+35 bar/+3500 kPa
+40 bar/+4000 kPa
15 bar/1500 kPa
35 bar/3500 kPa
50 bar/5000 kPa
65 bar/+6500 kPa
90 bar/9000 kPa
+130 bar/+13000 kPa
+200 bar/+20000 kPa
-0.2 bar/-20 kPa
-0.8 bar/-80 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-0.2 bar/-20 kPa
-0.4 bar/-40 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
-1 bar/-100 kPa
0 bar abs.
0 bar abs.
0 bar abs.
0 bar abs.
0 bar abs.
0 bar abs.
0 bar abs.
9) Data on overload capability apply for reference temperature.
VEGABAR 38 • Two-wire 4 … 20 mA 51
14 Supplement
Nominal measuring ranges and overload capacity in psi
Nominal range
Gauge pressure
0 … +1.5 psig
0 … +5 psig
0 … +15 psig
0 … +30 psig
0 … +75 psig
0 … +150 psig
0 … +300 psig
0 … +900 psig
-0.7 … +0.7 psig
-3 … +3 psig
-7 … +7 psig
-14.5 … 0 psig
-14.5 … +20 psig
Absolute pressure
0 … 15 psi
0 … 30 psi
0 … +75 psi
0 … 150 psi
0 … 300 psi
0 … 900 psi
Overload capacity, max. pressure Overload capacity, min. pressure
+225 psig
+435 psig
+525 psig
+725 psig
+950 psig
+1300 psig
+1900 psig
+2900 psig
+225 psig
+290 psi
+525 psig
+525 psig
+580 psig
525 psi
600 psi
975 psi
1350 psi
1500 psi
+2900 psi
0 psi
0 psi
0 psi
0 psi
0 psi
0 psi
-3 psig
-12 psig
-14.51 psig
-14.51 psig
-14.51 psig
-14.51 psig
-14.51 psig
-14.51 psig
-3 psig
-6 psig
-14.51 psig
-14.51 psig
-14.51 psig
Adjustment ranges
Specifications refer to the nominal measuring range, pressure values lower than -1 bar cannot be set
Zero/Span adjustment:
Ʋ Zero
Ʋ Span
Maximum permissible Turn Down
-20 … +95 %
-120 … +120 %
Unlimited (recommended 20 : 1)
Switch-on phase
Start-up time with operating voltage U
B
Starting current for run-up time
≤ 2 s
≤ 3.6 mA
Output variable - two-wire 4 … 20 mA
Output signal
Connection technology
Range of the output signal
Signal resolution
4 … 20 mA - passive
Two-wire
3.8 … 20.5 mA
5 µA
52 VEGABAR 38 • Two-wire 4 … 20 mA
Fault signal, current output (adjustable) ≤ 3.6 mA, ≥ 21 mA, last measured value
Max. output current 21.5 mA
Load
Transferred measured value
See load resistance under Power supply
Pressure
Dynamic behaviour output
Dynamic characteristics - Current output
14 Supplement
100 %
90 %
1 2
10 % t
T t
S t
A t t
Fig. 30: Behaviour the current output in case of sudden change of the process variable. t
T
S
: jump response time
1 Process variable
2 Output signal
: dead time; t
A
: rise time;
Size
Dead time
Rise time (10 … 90 %)
Step response time (ti: 0 s, 10 … 90 %)
Time
≤ 2 ms
≤ 4 ms
≤ 6 ms
Damping (63 % of the input variable) 0 … 9 s, adjustable
Reference conditions and influencing variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature
Ʋ Relative humidity
+15 … +25 °C (+59 … +77 °F)
45 … 75 %
Ʋ Air pressure
Determination of characteristics
Characteristic curve
Reference installation position
Influence of the installation position
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Limit point adjustment according to IEC 61298-2
Linear upright, diaphragm points downward
< 0.2 mbar/20 Pa (0.003 psig)
Deviation (according to IEC 60770)
Applies to the 4 … 20 mA current output and refers to the adjusted span. Turn down (TD) is the relation nominal measuring range/adjusted span.
VEGABAR 38 • Two-wire 4 … 20 mA 53
14 Supplement
Accuracy class
0.3 %
Non-linearity, hysteresis and repeatability with TD 1 : 1 up to 5 : 1
< 0.3 %
Non-linearity, hysteresis and repeatability with 5 : 1
< 0.06 % x TD
Influence of the medium or ambient temperature
Average temperature coefficient of the zero signal
Ʋ In the compensated temperature range 10)
Ʋ Outside the compensated temperature range typ.
< 0.15 %/10 K
0.3 %/10 K
1,5 %
1,2 %
-40°C 20°C 100°C
-1,2 %
-1,5 %
Fig. 31: Temperature error with TD 1 : 1
Long-term stability (according to DIN 16086)
Specifications refer to the set span. Turn down (TD) is the ratio: nominal measuring range/set span.
Time period
One year
Two years
Five years
Ten years
Long-term drift zero signal and output span
< 0.1 % x TD
< 0.15 % x TD
< 0.2 % x TD
< 0.4 % x TD
Ambient conditions
Ambient temperature device
Ambient temperature display
Storage and transport temperature
-40 … +80 °C (-40 … +176 °F)
-25 … +80 °C (-13 … +176 °F)
-40 … +80 °C (-40 … +176 °F)
Mechanical environmental conditions
Vibrations (oscillations)
Impacts (mechanical shock)
Class 4M8 acc. to IEC 60271-3-4 (5 g at 4 … 200 Hz)
Class 6M4 acc. to IEC 60271-3-6 (50 g, 2.3 ms)
10) 0 … +100 °C (+32 … +212 °F)
54 VEGABAR 38 • Two-wire 4 … 20 mA
14 Supplement
Impact resistance
Ʋ Plug according to ISO 4400
Ʋ M12 x 1 plug
IK07 acc. to IEC 62262
IK06 acc. to IEC 62262
Process conditions
Process temperature
Measuring cell seal
FKM
EPDM
FFKM
Process temperature with process fitting
316L, Duplex steel PVDF PEEK 11)
VP2/A -20 … +130 °C (-
4 … +266 °F)
A+P 70.10-02 -40 … +130 °C (-
40 … +266 °F)
Perlast G74S -15 … +130 °C
(+5 … +266 °F)
-20 … +80 °C (-
4 … +176 °F) 12)
-20 … +100 °C (-
4 … +212 °F)
Temperature derating
2
1
Fig. 32: Temperature derating VEGABAR 38
1 Process temperature
2 Ambient temperature
2
1
Fig. 33: Temperature derating VEGABAR 38, with activated Bluetooth communication
1 Process temperature
2 Ambient temperature
SIP process temperature 13)
14 Applied for OL approval
12) Process pressures > 5 bar: 20 … +60 °C (-4 … +140 °F)
13) SIP = Sterilization in place
VEGABAR 38 • Two-wire 4 … 20 mA 55
14 Supplement
Device configuration suitable for vapour, i.e. measuring cell seal EPDM or FFKM (Perlast G74S), previous CIP cleaning up to max. +80 °C (+176 °F): 14)
SIP process temperature with vapour stratification up to
Ʋ 15 minutes +150 °C (+302 °F)
Ʋ
Ʋ
30 minutes
1 hour
Process pressure
Permissible process pressure
+140 °C (+284 °F)
+135 °C (+275 °F) see specification " MWP " on the type label 15)
Indication
Measured value and menu display
Ʋ Graphic-capable LC display, with lighting
Ʋ Max. indicating range
Status indication
Adjustment
Adjustment elements
PC/Notebook
Smartphone/Tablet digital and quasianalogue indication
-99999 … 99999
LED illuminated ring (green-yellow-red)
3 x keys for menu adjustment
PACTware/DTM
Adjustment app
Measuring cell temperature
Range
Resolution
Deviation
Output of the temperature values via 16)
Bluetooth interface
Bluetooth standard
Frequency
Max. emitted power
Max. number of participants
Effective range typ. 17)
Electromechanical data
Angled plug connector
Ʋ Version
Ʋ Cable gland
Ʋ Screw terminals for cable crosssection up to
-40 … +130 °C (-40 … +266 °F)
< 0.2 K
±3 K
Display and adjustment unit, Bluetooth, IO-Link
Bluetooth 5.0
2.402 … 2.480 GHz
+2.2 dBm
1
25 m (82 ft)
4-pin according to ISO 4400
M16 x 1.5 (for cable diameter 4.5 … 10 mm)
1.5 mm² (AWG 15)
15) MWP: Maximum Working Pressure
16) Depending on the instrument version
17) Depending on the local conditions
56 VEGABAR 38 • Two-wire 4 … 20 mA
14 Supplement
Angle plug connector with flap lid
Ʋ Version
Ʋ Cable gland
Ʋ Screw terminals for cable crosssection
Round plug connector
4-pin according to ISO 4400
PG 11 (for cable diameter 4.5 … 11 mm) up to 1.5 mm² (AWG 15)
4-pole with M12 x 1 screw connection
Voltage supply
Operating voltage U
B
Operating voltage U
B
18)
Reverse voltage protection
Permissible residual ripple
Ʋ for U
N
Ʋ for U
N
12 V DC (12 V < U
B
24 V DC (18 V < U
Load resistor
B
< 18 V)
< 35 V)
Ʋ Calculation
Ʋ Example - for U
B
= 24 V DC
12 … 35 V DC
15 … 35 V DC
Integrated
≤ 0.7 V
≤ 1.0 V
(U
B
- U eff eff min
(16 … 400 Hz)
(16 … 400 Hz)
)/0.022 A
(24 V - 12 V)/0.022 A = 545 Ω
Electrical protective measures
Potential separation
Protection rating
Connection technology
Electronics potential free up to 500 V AC
Protection according to
EN 60529/IEC 529
IP66/IP67
Protection according to NEMA/
UL 50E
Type 6P M12 x 1 plug
Plug according to ISO 4400
ISO 4400 plug with hinged cover
Plug according to ISO 4400 with
IDC method of termination
IP65 Type 4X
Altitude above sea level
Protection class
Pollution degree
5000 m (16404 ft)
III
4
18) Illuminated display and adjustment unit or activated 360° status display
VEGABAR 38 • Two-wire 4 … 20 mA 57
14 Supplement
14.2 Dimensions
Connection technology
35 mm
(1.38")
40 mm (1.57"
66 mm (2.60"
36 mm
(1.42")
66 mm
(2.60")
1 2
Fig. 34: Connection technology VEGABAR 38
1 M12 x 1 plug connector
2 Plug connector according to ISO 4400
3 Plug connector according to ISO 4400 with hinged cover
36 mm (1.38"
3
40 mm
(1.57")
53 mm (2.09"
58 VEGABAR 38 • Two-wire 4 … 20 mA
14 Supplement
VEGABAR 38, threaded fitting not front-flush
ø 36 mm
(1.42")
ø 31,7 mm
(1.25")
SW 27 mm
(1.06")
G½
3 mm (0.12"
ø 3 mm
(0.12")
DU
ø 6 mm
(0.24")
23 mm (0.91"
95,5 mm (3.76"
SW 27 mm
(1.06")
M20x1,5
5 mm (0.20"
ø 3 mm
(0.12")
C2
ø 6 mm
(0.24")
25 mm (0.98"
97,5 mm (3.84"
SW 27 mm
(1.06")
G½
ø 11,4 mm
(0.45")
CS
20 mm (0.79"
92,5 mm (3.64"
SW 27 mm
(1.06")
ø 17,5 mm
(0.69")
20 mm (0.79"
G½
ø 11,4 mm
(0.45")
DQ
ø 17,5 mm
(0.69")
124 mm (4.88"
SW 27 mm
(1.06")
½ NPT
25 mm (0.98"
97,5 mm (3.84"
SW 27 mm
(1.06")
129 mm (5.08"
¼ NPT
25 mm (0.98"
¼ NPT
½ NPT
20 mm (0.79"
LF ø 6 mm
(0.24")
9U
Fig. 35: VEGABAR 38, threaded fitting not front-flush
DU Thread G½ (EN 837), manometer connection
C2 Thread M20 x 1.5 (EN 837), manometer connection
CS Thread G½, inside G¼ A (ISO 228-1), Duplex steel (1.4462)
LF Thread ½ NPT, inside ¼ NPT (ASME B1.20.1)
TI Thread ¼ NPT (ASME B1.20.1)
TJ Thread G¼ (ISO 228-1)
ø 3 mm
(0.12")
TI
13 mm (0.51"
) 85,5 mm (3.37")
SW 27 mm
(1.06")
G ¼
ø 3 mm
(0.12")
ø 5 mm
(0.20")
TJ
15 mm (0.59"
) 87,5 mm (3.44")
VEGABAR 38 • Two-wire 4 … 20 mA 59
14 Supplement
VEGABAR 38, threaded fitting front-flush
ø 36 mm
( 1.42")
ø 31,7 mm
( 1.25")
SW 27 mm
(1.06")
G½
ø27 mm
( 1.06")
ø 18,2 mm
( 0.72")
9L
SW 36 mm
(1.42")
G ¾
ø 32 mm
( 1.26")
N9
ø24 mm
( 0.94")
SW 41 mm
(1.61")
G 1
ø 40 mm
( 1.57")
C5
ø 30 mm
( 1.18")
SW 46 mm
(1.81")
G 1½
SW 46 mm
(1.81")
1½ NPT
ø 44 mm
( 1.74")
ø 55 mm
( 2.17")
DA
Fig. 36: VEGABAR 38, threaded fitting front-flush
9L Thread G½ (DIN 3852-A)
N9 Thread G¾ (DIN 3852-E)
C5 Thread G1 (ISO 228-1)
DA Thread G1½ (DIN 3852-A)
C9 Thread 1½ NPT (ASME B1.20.1)
8C Thread G1½ (DIN 3852-E), PEEK
ø 44 mm
( 1.74")
ø 55 mm
( 2.17")
C9
60
SW 46 mm
(1.81")
G 1½
ø 44 mm
( 1.74")
ø 60 mm
( 2.36")
8C
VEGABAR 38 • Two-wire 4 … 20 mA
14 Supplement
VEGABAR 38, threaded fitting front-flush with cone/extension
ø 36 mm
( 1.42")
ø 31,7 mm
( 1.25")
SW 41 mm
(1.61")
G 1
SW 41 mm
(1.61")
G 1
ø 40 mm
( 1.57")
DH
Fig. 37: VEGABAR 38, cone/extension fitting
DH Thread G1 (ISO 228-1), cone 40°
LX Thread G1 (ISO 228-1), hygienic design
AL Thread M30 x 1.5 (DIN 13)
ø 40 mm
( 1.57")
LX
SW 32 mm
(1.26")
M30x1,5
AL
ø 27 mm
( 1.08")
VEGABAR 38 • Two-wire 4 … 20 mA 61
14 Supplement
VEGABAR 38, hygienic fitting
ø 36 mm
( 1.42")
ø 31,7 mm
( 1.25")
ø 64 mm
( 2.52")
AR
ø 84 mm
( 3.31")
FB
ø 52 mm
( 2.05")
ø 25 mm
( 0.98")
E7
ø 78 mm
( 3.07")
ø 71 mm
( 2.80")
ø 84 mm
( 3.31")
FR EZ
Fig. 38: VEGABAR 38, hygienic fitting
AR Clamp 2" PN 40 ø 64 mm (DIN 32676, ISO 2852)
FB SMS DN 51 PN 6
E7 Ingold connection PN 10
FR Varivent N50-40 PN 25
EZ Collar socket DN 40 PN 40 (DIN 11851)
E2 Collar socket DN 40 PN 40 (DIN 11864-1, Form A)
62
ø 78 mm
( 3.07")
E2
VEGABAR 38 • Two-wire 4 … 20 mA
14 Supplement
14.3 Industrial property rights
VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad industrial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。
进一步信息请参见网站< www.vega.com
。
14.4 Licensing information for open source software
Open source software components are also used in this device. A documentation of these components with the respective license type, the associated license texts, copyright notes and disclaimers can be found on our homepage.
14.5 Trademark
All the brands as well as trade and company names used are property of their lawful proprietor/ originator.
VEGABAR 38 • Two-wire 4 … 20 mA 63
Notes
64 VEGABAR 38 • Two-wire 4 … 20 mA
Notes
VEGABAR 38 • Two-wire 4 … 20 mA 65
Notes
66 VEGABAR 38 • Two-wire 4 … 20 mA
Notes
VEGABAR 38 • Two-wire 4 … 20 mA 67
Printing date:
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing.
Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2022
VEGA Grieshaber KG
Am Hohenstein 113
77761 Schiltach
Germany
Phone +49 7836 50-0
E-mail: [email protected]
www.vega.com
advertisement
Related manuals
advertisement
Table of contents
- 4 1 About this document
- 4 1.1 Function
- 4 1.2 Target group
- 4 1.3 Symbols used
- 5 2 For your safety
- 5 2.1 Authorised personnel
- 5 2.2 Appropriate use
- 5 2.3 Warning about incorrect use
- 5 2.4 General safety instructions
- 5 2.5 Installation and operation in the USA and Canada
- 7 3 Product description
- 9 3.2 Principle of operation
- 12 3.3 Adjustment
- 12 3.4 Packaging, transport and storage
- 13 3.5 Accessories
- 14 4 Mounting
- 14 4.1 General instructions
- 16 4.2 Process pressure measurement
- 18 4.3 Level measurement
- 19 5 Connecting to power supply
- 19 5.1 Preparing the connection
- 19 5.2 Connection procedure
- 22 5.3 Wiring plan - Two-wire 4 … 20 mA
- 23 5.4 Switch-on phase
- 24 6 Access protection
- 24 6.1 Bluetooth radio interface
- 24 6.2 Protection of the parameterization
- 25 6.3 Storing the codes in myVEGA
- 26 7 Set up with the integrated display and adjustment unit
- 26 7.1 Adjustment system
- 27 7.2 Measured value and menu item display
- 28 7.3 Parameter adjustment
- 35 8 Setup with smartphone/tablet (Bluetooth)
- 35 8.1 Preparations
- 35 8.2 Connecting
- 36 8.3 Sensor parameter adjustment
- 37 9 Setup with PC/notebook (Bluetooth)
- 37 9.1 Preparations
- 37 9.2 Connecting
- 38 9.3 Parameter adjustment
- 39 10 Menu overview
- 39 10.1 Display and adjustment unit (on site)
- 40 10.2 VEGA Tools app and DTM (Bluetooth)
- 42 11 Diagnostics and servicing
- 42 11.1 Maintenance
- 42 11.2 Rectify faults
- 43 11.3 Diagnosis, fault messages
- 43 11.4 Status messages according to NE
- 45 11.5 Software update
- 46 11.6 How to proceed if a repair is necessary
- 47 12 Dismount
- 47 12.1 Dismounting steps
- 47 12.2 Disposal
- 48 13.1 Radio licenses
- 48 13.2 Approvals for Ex areas
- 48 13.5 Conformity
- 48 13.6 NAMUR recommendations
- 49 13.7 Environment management system
- 50 14 Supplement
- 50 14.1 Technical data
- 58 14.2 Dimensions
- 63 14.3 Industrial property rights
- 63 14.4 Licensing information for open source software
- 63 14.5 Trademark