11 Diagnostics and servicing
Code
Text message
F080
General software error
F110
Switching points too close together
F111
Switching points interchanged
F260
Error in the calibration
F261
Error in the instrument settings
Cause
General software error
Selected switching points too close together
Switching point 1 is smaller than switching point 2
Checksum error in the calibration values
Checksum error in the configuration values
Rectification
Restart
Increase the distance between the switching points
Select switching point 1 to greater than switching point 2
Send instrument for repair
Carry out a reset
Function check
Code
Text message
C700
Simulation active
Cause
A simulation is active
Rectification
Finish simulation
Wait for the automatic end after
60 mins.
Out of specification
Code
Text message
S600
Impermissible electronics temperature
S604
Switching output overloaded
Maintenance
Code
Text message
M504
Error at a device interface
M510
No communication with the main controller
Cause
Temperature of the electronics in the non-specified range
Cause
Rectification
Check ambient temperature
Insulate electronics
Overload or short circuit at output 1 or 2 Electrical connection, check load resistance
Rectification
Interference of the internal communication to Bluetooth
Fault in internal communication with the display
Restart
Send instrument for repair
Restart
Send instrument for repair
11.5 Software update
The device software is updated via Bluetooth.
The following components are required:
VEGABAR 38 • Two-wire 4 … 20 mA 45
advertisement
Related manuals
advertisement
Table of contents
- 4 1 About this document
- 4 1.1 Function
- 4 1.2 Target group
- 4 1.3 Symbols used
- 5 2 For your safety
- 5 2.1 Authorised personnel
- 5 2.2 Appropriate use
- 5 2.3 Warning about incorrect use
- 5 2.4 General safety instructions
- 5 2.5 Installation and operation in the USA and Canada
- 7 3 Product description
- 9 3.2 Principle of operation
- 12 3.3 Adjustment
- 12 3.4 Packaging, transport and storage
- 13 3.5 Accessories
- 14 4 Mounting
- 14 4.1 General instructions
- 16 4.2 Process pressure measurement
- 18 4.3 Level measurement
- 19 5 Connecting to power supply
- 19 5.1 Preparing the connection
- 19 5.2 Connection procedure
- 22 5.3 Wiring plan - Two-wire 4 … 20 mA
- 23 5.4 Switch-on phase
- 24 6 Access protection
- 24 6.1 Bluetooth radio interface
- 24 6.2 Protection of the parameterization
- 25 6.3 Storing the codes in myVEGA
- 26 7 Set up with the integrated display and adjustment unit
- 26 7.1 Adjustment system
- 27 7.2 Measured value and menu item display
- 28 7.3 Parameter adjustment
- 35 8 Setup with smartphone/tablet (Bluetooth)
- 35 8.1 Preparations
- 35 8.2 Connecting
- 36 8.3 Sensor parameter adjustment
- 37 9 Setup with PC/notebook (Bluetooth)
- 37 9.1 Preparations
- 37 9.2 Connecting
- 38 9.3 Parameter adjustment
- 39 10 Menu overview
- 39 10.1 Display and adjustment unit (on site)
- 40 10.2 VEGA Tools app and DTM (Bluetooth)
- 42 11 Diagnostics and servicing
- 42 11.1 Maintenance
- 42 11.2 Rectify faults
- 43 11.3 Diagnosis, fault messages
- 43 11.4 Status messages according to NE
- 45 11.5 Software update
- 46 11.6 How to proceed if a repair is necessary
- 47 12 Dismount
- 47 12.1 Dismounting steps
- 47 12.2 Disposal
- 48 13.1 Radio licenses
- 48 13.2 Approvals for Ex areas
- 48 13.5 Conformity
- 48 13.6 NAMUR recommendations
- 49 13.7 Environment management system
- 50 14 Supplement
- 50 14.1 Technical data
- 58 14.2 Dimensions
- 63 14.3 Industrial property rights
- 63 14.4 Licensing information for open source software
- 63 14.5 Trademark