11 Diagnostics and servicing

Code

Text message

F080

General software error

F110

Switching points too close together

F111

Switching points interchanged

F260

Error in the calibration

F261

Error in the instrument settings

Cause

General software error

Selected switching points too close together

Switching point 1 is smaller than switching point 2

Checksum error in the calibration values

Checksum error in the configuration values

Rectification

Restart

Increase the distance between the switching points

Select switching point 1 to greater than switching point 2

Send instrument for repair

Carry out a reset

Function check

Code

Text message

C700

Simulation active

Cause

A simulation is active

Rectification

Finish simulation

Wait for the automatic end after

60 mins.

Out of specification

Code

Text message

S600

Impermissible electronics temperature

S604

Switching output overloaded

Maintenance

Code

Text message

M504

Error at a device interface

M510

No communication with the main controller

Cause

Temperature of the electronics in the non-specified range

Cause

Rectification

Check ambient temperature

Insulate electronics

Overload or short circuit at output 1 or 2 Electrical connection, check load resistance

Rectification

Interference of the internal communication to Bluetooth

Fault in internal communication with the display

Restart

Send instrument for repair

Restart

Send instrument for repair

11.5 Software update

The device software is updated via Bluetooth.

The following components are required:

VEGABAR 38 • Two-wire 4 … 20 mA 45

advertisement