Programming via ASCII Commands. MICRO-EPSILON optoNCDT 2300

Add to My manuals
202 Pages

advertisement

Programming via ASCII Commands. MICRO-EPSILON optoNCDT 2300 | Manualzz

Operation optoNCDT 2300

6.3 Programming via ASCII Commands

As an added feature you can program the sensor via an ASCII interface, physically RS422 and / or Ethernet.

This requires, that the sensor must be connected either to a serial RS422 interface via a suitable interface

converter, see Chap. A 1

, or a plug-in-card to a PC / PLC. In addition, the Ethernet interface can be used via a suitable program, for example Telnet.

Pay attention in the programs used to the correct RS422 default setting or a valid Ethernet address.

Once connected, you can transfer the commands from the appendix, see Chap. A 6 , via the terminal or Telnet

to the sensor.

6.4 Timing, Measurement Value Flux

The sensor operates internally with real time cycles in a pipeline mode:

The sensor requires 5 cycles for measuring and processing without triggering.

1. Exposure: Charging the image detector in the receiver (measurement),

2. Reading: Reading out of the imaging device and converting into digital data,

3. Computing (2 cycles),

4. Synchronous output.

Each cycle takes about 20 µs at a measuring rate of 49.140 kHz. The measured value N is available after each cycle with a constant lag of 5 cycles in respect to the real time event. The delay time between detection and start of outputs is therefore 100 µs. The processing of the cycles occurs sequentially in time and parallel in space (pipelining). This guarantees a constant real time data stream. A measured value delayed by

5 cycles is output for each measurement cycle.

Averaging the measured values has no effect on the time behavior. Remember, however, that the sensor needs time for the averaging, until measured values are present according to the set averaging number N.

Depending on the nature of the averaging value and the number of averaged values, there are different settling times.

Page 48

Operation

Grey shaded fields require a selection.

Value

Dark-bordered fields require you to specify a value.

Diffuse displacement measurement

Linearization diffuse

measurement arrangement,

laser power 1 mW

Measurement program

Direct displacement measurement

Direct thickness measurement

Linearization direct measurement arrangement, laser power switchable

Displacement measurement highest peak, peak with the largest surface, 1. peak

Average selected displacement

Setting masters / zeroing selected peak

Statistics calculation for displacement

Displacement measurement on diffuse reflecting targets

Displacement measurement

on direct reflecting targets

Fig. 25 Adjustment possibilities of the optoNCDT 2300

Refractive index correction

Thickness measurement thickness, 1. & 2. peak

Average thickness & 1. & 2. displacement

Setting masters / zeroing thickness

Statistics calculation for thickness

Thickness measurement on direct

reflecting transparent targets optoNCDT 2300 Page 49

advertisement

Related manuals

advertisement

Table of contents