advertisement
4
Calibration
This chapter provides procedures to perform all adjustments required to calibrate the
Spot LXi to conform to Welch Allyn specifications. Calibration requires qualified personnel to open the device housing.
WARNING Electric shock hazard. There are no user-serviceable parts inside Spot
LXi other than battery replacement (see “Battery Replacement” on page 79). An operator may only perform maintenance procedures specifically described in this manual. For service, refer the device to an Authorized Service Center.
Note
Always disconnect the sealed lead-acid battery in the Spot Vital Signs LXi before performing any calibration function.
Gather the following tools to have available during the procedures.
Description
Blood Pressure Test Volume Repair Fixture
Spot LXi Repair Software
Serivce Manual, Spot LXi
Tester, Calibration, 9600 Plus
Blood Pressure Y-Tube, No Fittings 1/8 Tube
Pliers
Wire Cutter
Tweezers
#2 Phillips Screwdriver
T10 Torx Driver
Cable Tie Tool
*Setra Pressure Meter
*Netech Pressure Meter
Masimo SpO2 Tester
Nellcor SRC-MAX SpO
2
Tester
Cable, USB A to 5 pin mini
Digital Volt Meter with 4 1/2 Digit Display
Power Supply: 0-20 Vdc adjustable with 0-3A output
Part Number
401028
401710
704432
01802-110
5082-183
2270-01
2000 in
1593
SRC-MAX
Qty Source
1 Welch Allyn
1 Welch Allyn
1 Welch Allyn
1 Welch Allyn
1 Welch Allyn
1 Tool/Supply Store
1 Tool/Supply Store
1 Tool/Supply Store
1 Tool/Supply Store
1 Tool/Supply Store
1 Tool/Supply Store
1 Setra +1 800 257 3872
1 Netech +1 800 547 6557
1 Masimo +1 800 326 4890
1 Nellcor +1 800 635 5267
1 Electronics Supply Store
2 Electronics Supply Store
1 Electronic Supply Store
33
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 9 1 - Safety Summary
- 9 Introduction
- 9 Warnings and Cautions
- 9 General Warnings
- 10 Blood Pressure Warnings
- 12 Temperature Warnings
- 13 Warnings
- 14 General Cautions
- 15 Blood Pressure Cautions
- 15 Temperature Cautions
- 15 Cautions
- 16 Electrostatic Discharge (ESD)
- 17 Symbols
- 17 Safety Symbols
- 17 Button Symbols
- 18 Connection Symbols
- 18 Agency Symbols
- 19 2 - Overview
- 19 Purpose and Scope
- 19 Other Applicable Documents
- 20 Contents Checklist
- 20 Possible Attachments
- 21 Service
- 21 Technical Assistance
- 21 Field Replacement Units
- 21 Service Loaners
- 21 Service Intervals
- 22 Spot Vital Signs LXi Configurations
- 22 Controls
- 24 Display Window
- 26 Connections
- 27 Braun ThermoScan PRO 4000 Lock
- 27 Blood Pressure Hose and Cuff
- 27 Thermometer
- 28 Sensor
- 28 Quick Reference Card
- 29 AC Power Transformer
- 29 Battery
- 29 Standby Mode
- 31 3 - Functional Overview
- 31 Self Test
- 32 Internal Configuration Mode
- 36 Functional Verification
- 36 Blood Pressure Functional Check
- 37 Temperature Functional Check
- 40 Functional Check
- 41 4 - Calibration
- 42 Connections
- 42 Voltage Calibration
- 43 Blood Pressure Calibration
- 43 Configuration Calibration
- 45 5 - Troubleshooting
- 45 Error Codes
- 47 Battery Voltage Check
- 47 Window Display Check
- 47 Blood Pressure Calibration Check
- 48 Temperature Functional Check
- 48 Functional Check
- 48 Functional Testing Procedures
- 49 Current Test
- 49 Noise Levels
- 49 Button Test
- 49 Interface Test
- 50 Pneumatic Tests
- 50 Temperature Test
- 50 Overpressure Test
- 51 Service Work Checklist
- 51 Spot Vital Signs
- 53 6 - Disassembly and Repair
- 56 Key Pad Disassembly
- 57 Circuit Board Disassembly
- 57 Masimo
- 58 Nellcor
- 59 Blood Pressure Circuit Board Disassembly
- 61 LCD Inverter Ballast Board Disassembly
- 61 LCD Disassembly
- 62 Pump Disassembly
- 62 RS232-Communication Circuit Board Disassembly
- 63 Fan and Power Circuit Board Disassembly
- 63 Thermometry Circuit Board Disassembly
- 65 SureTemp Plus
- 67 Braun ThermoScan PRO
- 69 7 - Technical Overview
- 69 System Description
- 69 Battery System
- 69 Battery Charger
- 69 “CPU I/O” Power Supply
- 69 CPU Core Power Supply
- 70 +5V Supply
- 70 Clock/Calendar Power
- 70 Mod F NIBP Power
- 70 and Thermometer Power
- 70 LCD Power
- 71 External Load (RS-232) Power
- 71 SureTemp Plus Thermometer Interface
- 71 Braun ThermoScan PRO 4000 Thermometer Interface
- 71 Description
- 72 Bar Code Scanner Description
- 73 Printer Description
- 74 Interconnect Diagram
- 75 8 - Field Replaceable Units
- 79 9 - Specifications
- 79 Performance
- 79 Blood Pressure Accuracy
- 80 Temperature Specifications
- 81 Specifications
- 81 Masimo Sensor Accuracy Guide
- 81 Masimo Patents
- 82 Nellcor Sensor Accuracy Guide
- 82 Nellcor Patents
- 83 Mechanical
- 83 Electrical
- 83 Patents
- 83 Battery
- 83 Environmental
- 84 Wireless Radio
- 84 Guidance and Manufacturer’s Declaration
- 84 Emissions and Immunity Information
- 87 10 - Maintenance and Service
- 87 Cleaning
- 87 Spot Vital Signs LXi
- 87 Blood Pressure Cuff
- 87 Blood Pressure Hose and Cable
- 87 SureTemp Plus Thermometer
- 88 Braun ThermoScan PRO 4000 Thermometer
- 88 Sensors
- 88 Printer
- 89 Battery Replacement
- 89 Spot Vital Signs LXi
- 90 Braun ThermoScan PRO
- 90 Printer
- 91 A - Repair Test Specifications
- 91 General Unit Test
- 91 A-D Noise Test
- 91 Leak Test
- 91 Inflation Test
- 91 Dump Test
- 91 Pneumatic Accuracy Test
- 92 Valve Control Test
- 92 Voltage Calibration
- 92 Blank Mode Current Test
- 92 Back Light (Idle) Current Test
- 92 Valve/Pump Mode Current Test
- 92 Interface Test
- 92 Temperature Option Requirements
- 92 Accuracy Test
- 93 Temperature Probe Test
- 93 Option Requirements
- 93 Functional Test
- 93 Mode Current Test
- 93 Fail Safe Test
- 93 Over Pressure Test
- 93 Over 15 mmHg
- 95 B - Supplies and Accessories
- 95 Blood Pressure
- 96 Temperature
- 97 Pulse Oximetry
- 97 Masimo Accessories
- 98 Nellcor Accessories
- 99 Miscellaneous
- 99 Service Contracts
- 101 C - Miscellaneous Mounting Accessories
- 101 Wall Mount Kit
- 103 Mobile Stand Kit
- 105 DPAC Mounting Accessory
- 106 Printer Mounting Accessory
- 107 Scanner Mounting Accessory
- 109 Spot LXi
- 109 Accessories