MPower MP100 User Manual
Add to My manuals24 Pages
MPower MP100 is a portable personal gas monitor with a large LCD screen for easy viewing of gas concentrations. It can display real-time readings, as well as STEL, TWA, peak and min values. It has configurable high, low, STEL, and TWA alarm thresholds, and can be used to monitor a variety of gases, including CO, H2S, SO2, O2, and VOCs. The MP100 is easy to use, with two-key operation and a simple menu system. It is also durable and has a long battery life.
advertisement
MP100
User’s Guide
Rev.F
Sep 2018
MP100 User’s Guide
Contents
Read Before Operating ................................................................................................................... 2
General Information ....................................................................................................................... 3
User Interface .................................................................................................................................. 3
Display .............................................................................................................................................. 4
Using MP100 .................................................................................................................................... 5
Turn the unit on ....................................................................................................................... 5
Warm up sequence .................................................................................................................. 5
Normal mode ........................................................................................................................... 6
Real time reading ............................................................................................................ 6
STEL screen ..................................................................................................................... 6
TWA screen ...................................................................................................................... 6
Peak screen ...................................................................................................................... 6
Min screen (for Oxygen sensor only) ............................................................................. 7
Turn the unit off .............................................................................................................. 7
Config mode ..................................................................................................................................... 7
Enter Config mode .................................................................................................................. 7
Sensor calibration.................................................................................................................... 8
Calibration adapter ................................................................................................................. 8
Zero calibration ....................................................................................................................... 9
Span calibration ...................................................................................................................... 9
Bump test ............................................................................................................................... 10
Change alarm limit ............................................................................................................... 10
Change bump/cal interval .................................................................................................... 11
Change span value ................................................................................................................ 12
Change display unit .............................................................................................................. 12
Vibrator enable/disable ......................................................................................................... 12
Power on zero enable/disable ............................................................................................... 13
Fast power on enable/disable ............................................................................................... 13
Reset config ............................................................................................................................ 13
Exit Config mode ................................................................................................................... 14
Maintenance .................................................................................................................................. 14
Replace the battery ............................................................................................................... 14
Replace the sensor filter ........................................................................................................ 15
Replace the sensor ................................................................................................................. 15
Ordering information ................................................................................................................... 17
Specifications ................................................................................................................................. 18
Sensor configuration ..................................................................................................................... 19
Alarm signal summary ................................................................................................................. 20
Trouble shooting ............................................................................................................................ 22
1
MP100 User’s Guide
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining or servicing this product. The product will perform as designed only if it is used, maintained and serviced in accordance with the manufacturer’s instructions.
Warning
⚫
Never operate the monitor when the cover is removed.
⚫
Remove the monitor cover and battery only in area known as non-hazardous.
⚫ Use only mPower’s lithium battery part number 1.17.02.0002 (3.6V, 2700mAH, AA size) or part No. ER14505 cell manufactured by EVE Energy Co., LTD
⚫
This instrument has not been tested in an explosive gas/air atmosphere having an oxygen concentration greater than 21%.
⚫
Substitution of components will impair suitability for intrinsic safety.
⚫
Substitution of components will void warranty.
⚫
It is recommended to bump test with a known concentration gas to confirm the instrument is functioning properly before use.
⚫
Before use, ensure that the ESD film on the display is not damaged or peeling.
Avertissement:
⚫
N'utilisez jamais le moniteur lorsque le couvercle est enlevé.
⚫
Retirer le couvercle du moniteur et la batterie uniquement dans une zone connue comme non dangereuse.
⚫
Utilisez uniquement la batterie au lithium de mPower, pièce No. 1.17.02.0002 (3.6V, 2700mAH, taille AA) ou celle de EVE Énergie Cie., Lt é , pièce No. ER14505.
⚫
Cet instrument n'a pas été testé dans une atmosphère explosive gaz / air ayant une concentration en oxygène supérieure à 21%.
⚫
La substitution de composants compromettra l'aptitude à la sécurité intrinsèque.
⚫
La substitution des composants annulera la garantie.
⚫
Il est recommandé de tester avec un gaz de concentration connu pour confirmer que l'instrument fonctionne correctement avant de l'utiliser.
⚫ Avant de l’utiliser, assurez-vous que le film ESD sur l'écran n'est pas endommagé ou épluché.
Proper Product Disposal at The End Of Life
The Waste Electrical and Electronic Equipment(WEEE) directive (2002/96/EC) is intended to promote recycling of electrical and electronic equipment and their components at end of life . This symbol (crossed-out wheeled bin) indicates separate collection of waste electrical and electronic equipment in the EU countries. This product may contain one or more Nickel-metal hydride (NiMH),
Lithium-ion, or Alkaline batteries. Specific battery information is given in this user guide. Batteries must be recycled or disposed of properly. At the end of its life, this product must undergo separate collection and recycling from general or household waste. Please use the return and collection system available in your country for the disposal of this product.
2
MP100 User’s Guide
General Information
UNI (MP100) is a single sensor, portable, personal toxic gas monitor. It displays gas concentration continuously on a big segment LCD. It also monitors STEL, TWA, peak and min
(for O
2
only) value of the gas and can be displayed on demand.
High, Low, STEL&TWA alarm threshold values are configurable. The shell is made of high strength, durable material. Two key operation, simple to use. Sensor and battery can be replaced easily. Calibration is also very convenient.
User Interface
1.
Audible Alarm Port
2.
LED alarm window
3.
LCD
4.
Left Key (Confirm/Number increasing)
5.
Right Key (Power/ Cursor moving)
6.
Alligator clip
7.
Sensor
8.
Vibrator
3
MP100 User’s Guide
Display
1.
Gas name, includes: CO, H
2
S, SO
2
, O
2
, VOC
2.
Question mark
3.
OK
4.
Gas unit, includes: x10 -6 , ppm, %, mg/m 3 ,
µ
mol/mol
5.
Battery
6.
HIGH, LOW, STEL, TWA alarm
7.
Span calibration
8.
Zero calibration
9.
Number
4
Using MP100
Turn the unit on
MP100 User’s Guide
Press and hold the Right Key for 3 seconds, until LCD displays green LED flashes, then release the button, the unit is powered on.
, buzzer beeps,
Warm up sequence
After powered on, the unit enter warm up and self-test sequence, show the firmware version as follows:
If the sensor is not able to be identified by the instrument or is not installed into the instrument, then the screen switches between following values will be shown accordingly:
High alarm threshold value: and .
Otherwise, the
Low alarm threshold value:
STEL (short-term exposure limit) alarm threshold value:
5
MP100 User’s Guide
TWA (time-weighted average) alarm threshold value:
Normal mode
Real time reading
The unit enters normal mode, start monitoring gas concentration and display on the LCD screen.
User can check some other values like STEL,TWA, PEAK, and MIN (for O
2
only) by pressing the
Right Key
.
*Unit will return to real time reading from any other screen when there’s no key action for 60 seconds.
STEL screen
TWA screen
Peak screen
Press the
Left Key
to clear peak value
6
MP100 User’s Guide
Press the Left Key again, peak value is cleared.
Min screen (for Oxygen sensor only)
Press the
Left Key
to clear min value
Press the
Left Key
again, minimum value is cleared
Turn the unit off
In normal display mode, press and hold the Right Key, then the unit will display a 5 second count down, LEDs will flash and buzzer will beep once per second.
Config mode
In Config mode, user can do calibration & change parameters for the unit. In general, use the Left
Key to increase the number or confirm, use the Right Key to move the cursor or move to the next programming item.
Enter Config mode
Press and hold the Left Key and the Right Key together for 3 seconds, the unit enters Config mode.
The Config mode is password protected, LCD displays to prompt enter password. The screen displays
7
, with one digit flashing.
MP100 User’s Guide
To input password, use the Left Key to increase the number, use the Right Key to move cursor.
Once all four digits are input, the cursor will move to “OK”, use the Left Key to finish password input and enter the Config mode.
If the digit input is mistaken, use the Right Key to move cursor between four digits and “OK” mark, to change the input.
*MP100 preset password is 0000
Sensor calibration
Before the unit can monitor gas correctly, it needs to know the metric, this is done by zero calibration and span calibration.
Calibration adapter
1.
Calibration adapter is used to apply gas to the unit during calibration.
2.
Before span calibration (in the following section), attach the Calibration Adapter over the inlet port on the front of MP100 by pressing it into place.
3.
Open the gas cylinder valve, then press the Left Key to start the calibration count down.
8
MP100 User’s Guide
4.
Take off the Calibration Adapter and complete calibration.
When monitoring, never operate the MP100 with the Calibration Adaptor attached. The
MP100’s sensor operates by diffusion. If the Calibration Adapter is attached during normal operation, inconsistent and lower-than-normal readings will occur because of decreased concentration of the gas being monitored.
Zero calibration
Zero calibration is to set the base line for the sensor, it is done in fresh air. When LCD displays , press the Left Key to start zero calibration. The unit will start a 15 second count-down, after the count-down is finished, zero calibration result will be displayed on the LCD, pass or fail .
If user does not want to do zero calibration, during the 15 seconds count, press the Right Key,
LCD displays , zero calibration is aborted.
Span calibration
Span calibration is to set the metric of the sensor to the gas, it is done with a known concentration gas. When LCD displays , apply the known gas to the unit, then press the Left
Key to start span calibration. Unit will start count-down, the count-down time depends on
9
MP100 User’s Guide different sensors (normally 60 seconds), after count-down is finished, span calibration result will be displayed on the LCD, pass or fail .
If user does not want to do span calibration during count-down, press the Right Key, LCD displays , span calibration is aborted.
Bump test
Bump test is to check if sensor and alarm devices are working properly, it is done with a known concentration gas. When LCD displays , apply the known gas to the unit, then press the Left Key to start bump test. Unit will start count-down, the count-down time depends on different sensors (normally 45 seconds), after count-down is finished, bump test result will be displayed on the LCD, pass or fail .
If user does not want to do bump test during count-down, press the Right Key, LCD displays , bump test is aborted.
Change alarm limit
All the preset alarm limits, High, Low, STEL & TWA can be changed. When LCD displays:
, , , to change the corresponding alarm limit, the value change process is similar.
, press the Left Key
. First the current setting value is displayed, with the first digit flashing:
10
MP100 User’s Guide
Use the Left Key to increase the current digit,
Use the Right Key to move cursor to the next digit:
, cycle from 0 to 9.
After all digits are done, use the Right Key to move cursor to “OK” symbol, press the Left Key to confirm the changing.
* MP100 will show “Err”, if the input data is invalid as follows,
- Low alarm setting is bigger than high alarm setting.
- High alarm setting is smaller than low alarm setting.
- Input data is bigger than measuring range.
Change bump/cal interval
The bump and cal interval can also be changed. When LCD switches between:
, and , and , press the Left
Key to change the corresponding interval range, the value change process is similar.
First the current setting value is displayed, with the first digit flashing: .
Use the Left Key to increase the current digit, , cycle from 0 to 9.
Use the Right Key to move cursor to the next digit:
After all digits are done, use the Right Key to move cursor to “OK” symbol,
11
,
MP100 User’s Guide press the Left Key to confirm the changing.
* MP100 will show “Err”, if the input data is out of valid range:0~180 day(s).
Change span value
The span calibration preset value can also be changed, the change process is similar with the alarm limit. But the new span will not take effect until user successfully completes a span calibration next time.
* MP100 will show “Err”, if the input data is invalid as follows:
- Span setting is smaller than 5% measuring range or bigger than measuring range.
- For Oxygen sensor, span setting is bigger than 19.0 .
Change display unit
MP100 supports different gas units, to change gas unit, when LCD switches between and , press the Left Key to change gas unit. The supported units for current sensor are all displayed on the LCD, the current selected unit is blinking. Use the
Right Key to change unit, use the Left Key to confirm selected gas unit.
Vibrator enable/disable
The vibrator consumes a lot of power, it can be disabled to save battery power, to extend the battery life.
When LCD switches between and , press the Left Key to change the vibrator enable/disable status. The current vibrator status is displayed on the LCD, switching between and
12
if the vibrator is enabled, or
MP100 User’s Guide switching between and , if the vibrator is disabled. Use the
Right Key to change the enable/disable status, and use the Left Key to confirm the change.
Power on zero enable/disable
Sensor base line may have some changes due to the environment (temperature, humidity), that will require a zero calibration. MP100 can do zero calibration every time the unit is powered on; this feature can be enabled/disabled.
When LCD switches between and , press the Left Key to go to change power on zero calibrations enable/disable status. The current enable/disable status is displayed on the LCD, use the Right Key to change, use the Left Key to confirm the change.
Fast power on enable/disable
If fast startup is enabled, the screens showing High/Low/STEL/TWA alarm threshold value will be skipped during warm up sequence.
When LCD switches between and , press the
Left Key
to change fast startup enable/disable status. The status is displayed on the LCD, switching between and if the fast startup is enabled, or switching between and if the fast startup is disabled. Use the Right Key to change the enable/disable status, use the Left Key to confirm the change.
Reset config
If the unit parameter is incorrect and user does not know how to set them back, user can use reset config to make all the parameters back to factory default.
13
MP100 User’s Guide
When LCD switches between and enter config reset menu, press the Left Key to confirm config reset.
Exit Config mode
, press the Left Key to
When LCD displays , press the Left Key exit from Config mode back to normal mode.
Maintenance
Maintenance should be performed only by a qualified person who has proper training and fully understands the contents of the manual.
Replace the battery
When the battery’s charge is low, LCD displays triggered once every minute. User needs to replace the battery.
, battery low alarm will be
When battery is dead, LCD displays , battery dead alarm will be triggered once every second. User needs to replace the battery.
To replace battery:
1 ) Turn off the MP100.
2 ) Place the MP100 face down on a soft surface.
3 ) Use a T10 Torx screwdriver to loosen each of the four screws by turning them counterclockwise.
4 ) Remove the top cover after carefully unplugging the buzzer connector.
5 ) Slide the battery out of its compartment.
6 ) Place the new battery into the compartment with its “+” end oriented toward the “+” on the
14
MP100 User’s Guide printed circuit board.
7 ) Plug in the buzzer connector and reinstall the top cover.
8 ) Install the screws in back cover. Be careful to not overtighten the screws.
Replace the sensor filter
A “peel-and-stick” filter should be used on the MP100 in order to keep debris from fouling the sensor. Sheets of 5 filters are available. When the filter appears dirty, replace it with a new one and dispose of the dirty filter.
1) Turn off the MP100.
2) Place the MP100 face down on a soft surface.
3) Use a T10 Torx screwdriver to loosen each of the four screws by turning them counterclockwise.
4) Remove the top cover after carefully unplugging the buzzer connector.
5) Peel a filter from the sheet and center it over the sensor. Gently press down.
6) Plug in the buzzer connector and reinstall the top cover.
7) Install the screws in back cover. Be careful to not overtighten the screws.
Replace the sensor
MP100 models are designed so that you can easily change the sensor.
1) Turn off the unit.
2) Place the MP100 face down on a soft surface.
3) Use a T10 Torx screwdriver to loosen each of the four screws by turning them counterclockwise.
4) Remove the top cover after carefully unplugging the buzzer connector.
5) Replace the old sensor with a new one. Make sure the pins are not bent or corroded . Align the pins to the corresponding holes and push the sensor straight in . The sensor should fit flush against the printed circuit board.
6) Plug in the buzzer connector and reinstall the top cover.
7) Install the screws in back cover. Be careful to not overtighten the screws.
Sensors are not interchangeable. Use only mPower sensors, and use only the sensor type specified for your MP100 monitor. Use of non-mPower components will void the warranty and can compromise the safe performance of this product.
15
MP100 User’s Guide
*Filter need to be replaced when it becomes discolored, clogged with particles or draws in liquid, otherwise it will contaminate sensor and void warranty.
Caution
⚫
Change battery only in area known to be non-hazardous.
⚫
Use only mPower battery, PN: 1.17.02.0002 or part No. ER14505 cell manufactured by
EVE Energy Co., LTD.
Attention
⚫
Changez la batterie uniquement dans une zone connue pour être non dangereuse.
⚫
Utilisez uniquement la batterie au lithium de mPower, pièce No. 1.17.02.0002 (3.6V,
2700mAH, taille AA) ou celle deEVE Énergie Cie., Lt
é
, pièce No. ER14505.
16
MP100 User’s Guide
Ordering information
Kit Package
⚫
MP100 instrument (with sensor, AA size lithium battery and alligator clip)
⚫
Quick start guide
⚫
Calibration adapter
Calibration Kits ( Optional )
⚫
Test gas (for sensor as specified)
⚫
Gas regulator with flow controller
⚫
Hard transport case
⚫
Multi-unit charging and calibration station
Year of manufacturing
To identify the year of manufacturing, please refer to the serial number of the instrument.
The fifth to sixth digit in the serial number indicates the year of manufacturing. 00~99 indicates the manufacturing year is 2000 to 2099.
17
MP100 User’s Guide
Specifications
Size
Weight
Temperature
Humidity
Alarm type
Alarm signal
Calibration
3.46 x 2.44 x 1.3 in.
(88 x 62 x 33mm)
4.4 oz. (125g)
-4 ℉ ~ 122 ℉ (-20 ℃ ~ 50 ℃ )
5 ~ 95% relative humidity(non-condensing)
High alarm, low alarm, STEL alarm & TWA alarm adjustable.
Over range alarm, battery low alarm.
Audible: 90dB @ 30cm
Visual: Bright red LED
Sense: Built in vibrator
2 point calibration, zero and span, power on zero (need user confirm).
Up to 10 alarm events
20 seconds(only for CO/H
2
S/O
2
)
Event log
Response time(T90)
IP rating
EMI/RFI
Certifications
Battery
Sensor
Warranty
IP67
Compliant with EMC 2004/108/EC
Intrinsic Safety:
UL/CUL:
Class I, Group A, B, C, D
Class II, Group E,F,G
Class III ,Hazardous Locations
T4, -20° C≤ Tamb≤ +50° C
Replaceable AA size Lithium battery,
3 years if used within specs
CO/H
2
S expected operating life 5 years in air, others 2 years
2 year warranty for CO, H
2
S, O
2
and SO
2
, 1 year for the others.
18
MP100 User’s Guide
NH
3
Cl
2
ClO
2
HCN
NO
NO
2
PH
3
SO
2
Sensor Range
(ppm)
CO
0-1999
0-1000
0-500
0~50
H
2
S
ETO
0-100
0~200
0~1000
0-50
0-10
0-1
0-100
0-250
0-20
0-20
0-20
0-100
0-200
O
3
HF
0-1
0-10
HCl 0-15
CH
3
SH 0-10
C
2
H
4
O 0-20
Sensor Range
(%)
O
2
0 - 25
0 - 30
Sensor configuration
Resolution
(ppm)
1
1
1
0.1
0.1
0.1
1
1
0.1
0.01
1
1
0.1
0.01
0.1
0.1
0.1
0.02
0.1
0.1
0.1
0.1
Resolution
(%)
0.1
0.1
Span Low High STEL TWA Panel
(ppm) (ppm) (ppm) (ppm) (ppm) Ring
100
100
35
35
200
200
100
100
35 Red
35
T90(s)
50 35 200 100 35
25 10 20 15 10 Light
25 10 20 15 10
25 10 20 15 10
Blue 20
20
25 10 20 15 10 30
50 25 50 35 25 Orange 150
20
20
20
20
10 0.5 5 1 0.5 Orange 30
0.5 0.2 0.5 0.3 0.1 Orange 120
10 4.7 50 4.7 4.7 Orange 200
25 25 50 25 25 Orange 30
5
5
5
10
10
1
1
2
1
5
10
2
10
5
10
1 1 Orange 30
1 0.3 Orange 60
5
5
5
2
1
1
Orange 15
Orange 120
Orange 120
1 0.1 0.2 0.1 0.1 Orange 60
6 2 6 6 3 Orange 90
10 2 5 5 1 Orange 90
5 0.5 5 0.5 0.5 Orange 35
5 2 10 5 2 Orange 120
Span
(%)
18
18
Low
(%)
19.5
19.5
High
(%)
23.5
23.5
STEL TWA
(%) (%)
-
-
Panel
Ring
- Dark
- Blue
T90(s)
20
20
Caution
Use only mPower sensors.
Call for Availability.
19
MP100 User’s Guide
Alarm signal summary
LCD Display Reason/Alarm Signals
Over range :
Buzzer 3 beeps per second
LED 3 flashes per second
1 vibration per second
“OVER” and “500” 1 flash per second
High alarm :
Buzzer 3 beeps per second
LED 3 flashes per second
1 vibration per second
“ HIGH ” 2 flashes per second
Low alarm
:
Buzzer 2 beeps per second
LED 2 flashes per second
1 vibration per second
“ LOW ” 2 flashes per second
STEL alarm :
Buzzer 1 beeps per second
LED 1 flash per second
1 vibration per second
“ STEL ” 2 flashes per second
TWA alarm :
Buzzer 1 beep per second
LED 1 flash per second
1 vibration per second
“ TWA ” 2 flashes per second
Negative drift alarm :
Buzzer 1 beep per second
LED 1 flash per second
1 vibration per second
20
MP100 User’s Guide
Bump over due alarm:
Buzzer 1 beep per minute
LED 1 flash per minute
1 vibration per minute
Cal over due alarm:
Buzzer 1 beep per minute
LED 1 flash per minute
1 vibration per minute
Battery empty alarm :
Buzzer 1 beep per minute
LED 1 flash per minute
1 vibration per minute
1 flash per minute
Sensor error
:
Buzzer 1 beep per second
LED 1 flash per second
“SEN Err”1 flash per second
Battery low alarm :
Buzzer 1 beep per second
LED 1 flash per second
“bAT LoW”1 flash per second
21
MP100 User’s Guide
Trouble shooting
Problem
Can not turn on unit
Possible reason
Battery not installed
Depleted or defective battery
Reading abnormally low Calibration Adapter is attached.
Incorrect calibration.
Buzzer, vibration inoperative
LED, or alarm
Bad buzzer, LEDs, or vibration alarm.
Blocked alarm port
Solution
Install battery.
Replace battery.
Remove Calibration Adapter.
Calibrate the MP100.
Call authorized service center.
Unblock alarm port.
22
MP100 User’s Guide
mPower Electronics Inc.
3046 Scott Blvd
Santa Clara, CA 95054
www.mpowerinc.com
Tech Support: [email protected]
23
advertisement
Related manuals
advertisement
Table of contents
- 3 Read Before Operating
- 4 General Information
- 4 User Interface
- 5 Display
- 6 Using MP
- 6 Turn the unit on
- 6 Warm up sequence
- 7 Normal mode
- 7 Real time reading
- 7 STEL screen
- 7 TWA screen
- 7 Peak screen
- 8 Min screen (for Oxygen sensor only)
- 8 Turn the unit off
- 8 Config mode
- 8 Enter Config mode
- 9 Sensor calibration
- 9 Calibration adapter
- 10 Zero calibration
- 10 Span calibration
- 11 Bump test
- 11 Change alarm limit
- 12 Change bump/cal interval
- 13 Change span value
- 13 Change display unit
- 13 Vibrator enable/disable
- 14 Power on zero enable/disable
- 14 Fast power on enable/disable
- 14 Reset config
- 15 Exit Config mode
- 15 Maintenance
- 15 Replace the battery
- 16 Replace the sensor filter
- 16 Replace the sensor
- 18 Ordering information
- 19 Specifications
- 20 Sensor configuration
- 21 Alarm signal summary
- 23 Trouble shooting