Zero calibration. MPower MP100
Add to My manuals24 Pages
MPower MP100 is a portable personal gas monitor with a large LCD screen for easy viewing of gas concentrations. It can display real-time readings, as well as STEL, TWA, peak and min values. It has configurable high, low, STEL, and TWA alarm thresholds, and can be used to monitor a variety of gases, including CO, H2S, SO2, O2, and VOCs. The MP100 is easy to use, with two-key operation and a simple menu system. It is also durable and has a long battery life.
advertisement
![Zero calibration. MPower MP100 | Manualzz Zero calibration. MPower MP100 | Manualzz](http://s3.manualzz.com/store/data/065085398_1-a02422d93080c874e9c0245cb9e87467-360x466.png)
MP100 User’s Guide
4.
Take off the Calibration Adapter and complete calibration.
When monitoring, never operate the MP100 with the Calibration Adaptor attached. The
MP100’s sensor operates by diffusion. If the Calibration Adapter is attached during normal operation, inconsistent and lower-than-normal readings will occur because of decreased concentration of the gas being monitored.
Zero calibration
Zero calibration is to set the base line for the sensor, it is done in fresh air. When LCD displays , press the Left Key to start zero calibration. The unit will start a 15 second count-down, after the count-down is finished, zero calibration result will be displayed on the LCD, pass or fail .
If user does not want to do zero calibration, during the 15 seconds count, press the Right Key,
LCD displays , zero calibration is aborted.
Span calibration
Span calibration is to set the metric of the sensor to the gas, it is done with a known concentration gas. When LCD displays , apply the known gas to the unit, then press the Left
Key to start span calibration. Unit will start count-down, the count-down time depends on
9
advertisement
Related manuals
advertisement
Table of contents
- 3 Read Before Operating
- 4 General Information
- 4 User Interface
- 5 Display
- 6 Using MP
- 6 Turn the unit on
- 6 Warm up sequence
- 7 Normal mode
- 7 Real time reading
- 7 STEL screen
- 7 TWA screen
- 7 Peak screen
- 8 Min screen (for Oxygen sensor only)
- 8 Turn the unit off
- 8 Config mode
- 8 Enter Config mode
- 9 Sensor calibration
- 9 Calibration adapter
- 10 Zero calibration
- 10 Span calibration
- 11 Bump test
- 11 Change alarm limit
- 12 Change bump/cal interval
- 13 Change span value
- 13 Change display unit
- 13 Vibrator enable/disable
- 14 Power on zero enable/disable
- 14 Fast power on enable/disable
- 14 Reset config
- 15 Exit Config mode
- 15 Maintenance
- 15 Replace the battery
- 16 Replace the sensor filter
- 16 Replace the sensor
- 18 Ordering information
- 19 Specifications
- 20 Sensor configuration
- 21 Alarm signal summary
- 23 Trouble shooting