Replace the sensor filter. MPower MP100
Add to My manuals24 Pages
MPower MP100 is a portable personal gas monitor with a large LCD screen for easy viewing of gas concentrations. It can display real-time readings, as well as STEL, TWA, peak and min values. It has configurable high, low, STEL, and TWA alarm thresholds, and can be used to monitor a variety of gases, including CO, H2S, SO2, O2, and VOCs. The MP100 is easy to use, with two-key operation and a simple menu system. It is also durable and has a long battery life.
advertisement
MP100 User’s Guide printed circuit board.
7 ) Plug in the buzzer connector and reinstall the top cover.
8 ) Install the screws in back cover. Be careful to not overtighten the screws.
Replace the sensor filter
A “peel-and-stick” filter should be used on the MP100 in order to keep debris from fouling the sensor. Sheets of 5 filters are available. When the filter appears dirty, replace it with a new one and dispose of the dirty filter.
1) Turn off the MP100.
2) Place the MP100 face down on a soft surface.
3) Use a T10 Torx screwdriver to loosen each of the four screws by turning them counterclockwise.
4) Remove the top cover after carefully unplugging the buzzer connector.
5) Peel a filter from the sheet and center it over the sensor. Gently press down.
6) Plug in the buzzer connector and reinstall the top cover.
7) Install the screws in back cover. Be careful to not overtighten the screws.
Replace the sensor
MP100 models are designed so that you can easily change the sensor.
1) Turn off the unit.
2) Place the MP100 face down on a soft surface.
3) Use a T10 Torx screwdriver to loosen each of the four screws by turning them counterclockwise.
4) Remove the top cover after carefully unplugging the buzzer connector.
5) Replace the old sensor with a new one. Make sure the pins are not bent or corroded . Align the pins to the corresponding holes and push the sensor straight in . The sensor should fit flush against the printed circuit board.
6) Plug in the buzzer connector and reinstall the top cover.
7) Install the screws in back cover. Be careful to not overtighten the screws.
Sensors are not interchangeable. Use only mPower sensors, and use only the sensor type specified for your MP100 monitor. Use of non-mPower components will void the warranty and can compromise the safe performance of this product.
15
MP100 User’s Guide
*Filter need to be replaced when it becomes discolored, clogged with particles or draws in liquid, otherwise it will contaminate sensor and void warranty.
Caution
⚫
Change battery only in area known to be non-hazardous.
⚫
Use only mPower battery, PN: 1.17.02.0002 or part No. ER14505 cell manufactured by
EVE Energy Co., LTD.
Attention
⚫
Changez la batterie uniquement dans une zone connue pour être non dangereuse.
⚫
Utilisez uniquement la batterie au lithium de mPower, pièce No. 1.17.02.0002 (3.6V,
2700mAH, taille AA) ou celle deEVE Énergie Cie., Lt
é
, pièce No. ER14505.
16
advertisement
Related manuals
advertisement
Table of contents
- 3 Read Before Operating
- 4 General Information
- 4 User Interface
- 5 Display
- 6 Using MP
- 6 Turn the unit on
- 6 Warm up sequence
- 7 Normal mode
- 7 Real time reading
- 7 STEL screen
- 7 TWA screen
- 7 Peak screen
- 8 Min screen (for Oxygen sensor only)
- 8 Turn the unit off
- 8 Config mode
- 8 Enter Config mode
- 9 Sensor calibration
- 9 Calibration adapter
- 10 Zero calibration
- 10 Span calibration
- 11 Bump test
- 11 Change alarm limit
- 12 Change bump/cal interval
- 13 Change span value
- 13 Change display unit
- 13 Vibrator enable/disable
- 14 Power on zero enable/disable
- 14 Fast power on enable/disable
- 14 Reset config
- 15 Exit Config mode
- 15 Maintenance
- 15 Replace the battery
- 16 Replace the sensor filter
- 16 Replace the sensor
- 18 Ordering information
- 19 Specifications
- 20 Sensor configuration
- 21 Alarm signal summary
- 23 Trouble shooting