Bump test. MPower MP100
Add to My manuals24 Pages
MPower MP100 is a portable personal gas monitor with a large LCD screen for easy viewing of gas concentrations. It can display real-time readings, as well as STEL, TWA, peak and min values. It has configurable high, low, STEL, and TWA alarm thresholds, and can be used to monitor a variety of gases, including CO, H2S, SO2, O2, and VOCs. The MP100 is easy to use, with two-key operation and a simple menu system. It is also durable and has a long battery life.
advertisement
MP100 User’s Guide different sensors (normally 60 seconds), after count-down is finished, span calibration result will be displayed on the LCD, pass or fail .
If user does not want to do span calibration during count-down, press the Right Key, LCD displays , span calibration is aborted.
Bump test
Bump test is to check if sensor and alarm devices are working properly, it is done with a known concentration gas. When LCD displays , apply the known gas to the unit, then press the Left Key to start bump test. Unit will start count-down, the count-down time depends on different sensors (normally 45 seconds), after count-down is finished, bump test result will be displayed on the LCD, pass or fail .
If user does not want to do bump test during count-down, press the Right Key, LCD displays , bump test is aborted.
Change alarm limit
All the preset alarm limits, High, Low, STEL & TWA can be changed. When LCD displays:
, , , to change the corresponding alarm limit, the value change process is similar.
, press the Left Key
. First the current setting value is displayed, with the first digit flashing:
10
advertisement
Related manuals
advertisement
Table of contents
- 3 Read Before Operating
- 4 General Information
- 4 User Interface
- 5 Display
- 6 Using MP
- 6 Turn the unit on
- 6 Warm up sequence
- 7 Normal mode
- 7 Real time reading
- 7 STEL screen
- 7 TWA screen
- 7 Peak screen
- 8 Min screen (for Oxygen sensor only)
- 8 Turn the unit off
- 8 Config mode
- 8 Enter Config mode
- 9 Sensor calibration
- 9 Calibration adapter
- 10 Zero calibration
- 10 Span calibration
- 11 Bump test
- 11 Change alarm limit
- 12 Change bump/cal interval
- 13 Change span value
- 13 Change display unit
- 13 Vibrator enable/disable
- 14 Power on zero enable/disable
- 14 Fast power on enable/disable
- 14 Reset config
- 15 Exit Config mode
- 15 Maintenance
- 15 Replace the battery
- 16 Replace the sensor filter
- 16 Replace the sensor
- 18 Ordering information
- 19 Specifications
- 20 Sensor configuration
- 21 Alarm signal summary
- 23 Trouble shooting