Thermo T 6 P, T 12 P, T 20 P Drying Oven Operating instructions

The T 6 P, T 12 P, and T 20 P Drying Ovens are designed for laboratory use. They can be used for various applications such as drying, ageing, analysis, fusing, burning in, oxidizing, reducing, preheating, and hot-air sterilization. The ovens are equipped with a programmed controller and can be set to operate at temperatures between room temperature + approx. 10°C and 250 °C. The ovens are also designed for continuous use and feature an independent overtemperature protection device, which means they do not require constant supervision.

PDF
Document
Thermo T 6 P, T 12 P, T 20 P Drying Ovens Operating Instructions | Manualzz

Advertisement

Advertisement

Valid: 09.2005 / 50042749 C
Laboratory Drying Ovens
Models T 6 P, T 12 P, T 20 P
Operating instructions
GB
USA
1a
Below is a list of the international Thermo marketing organizations.
Postal ad dress Germany
Thermo Electron LED GmbH
D – 63505 Langenselbold
Robert-Bosch-Strasse 1
Postal address USA
Thermo Electron Corporation.
275 Aiken Road
Asheville, NC 28804
USA
Enquiries from Germany
Phone
Sales 0800 1 536376
Service 0800 1 112110
Fax
Sales/Service 0800 1 112114
E-Mail
[email protected]
Enquiries from North America
Phone + 1 800-879 7767
Fax + 1 828-658 0363
E-Mail
[email protected]
Enquiries from Europe, Middle East
and Africa
Phone + 49(0) 6184 / 90-6940
Fax + 49(0) 6184 / 90-7474
E-Mail
[email protected]
Enquiries from Latin America
Phone + 1 828-658 2711
Fax + 1 828-645 9466
E-Mail
[email protected]
Enquiries from Asia Pacific
Phone + 1 852-2711 3910
Fax + 1 852-2711 3858
E-Mail
[email protected]
Internet: www.thermo.com
2a
Thermo Electron LED GmbH, 63505 Langenselbold, Germany
function line® is a registered trademark of Thermo Electron LED GmbH.
If translated versions of this manual are used, the German version is the authoritative basis.
Subject to technical modifications.
This instruction manual applies for the following models:
Order No. Model Equipment
50 042 294 T 6 P Drying oven, programmed controller, 1/PE AC, 230 V; 50/60 Hz
50 042 296 T 6 P Drying oven, programmed controller, 1/PE AC, 120 V; 50/60 Hz
50 042 304 T 12 P Drying oven, programmed controller, 1/PE AC, 230 V; 50/60 Hz
50 043 407 T 12 P Drying oven, programmed controller, 1/PE AC, 208 V; 50/60 Hz
50 042 310 T 20 P Drying oven, programmed controller, 1/PE AC, 230 V; 50/60 Hz
50 043 309 T 20 P Drying oven, programmed controller, 1/PE AC, 208 V; 50/60 Hz
The safety concerning the protection of persons, environment and material to be treated
mainly depends on the behavior of the operating personnel of these units.
Please read and observe the following instructions carefully before starting the unit in
order to avoid faults and resulting damage, especial ly adverse health effects.
2b
Information according EN 61010
AMBIENT OPERATING CONDITIONS
For indoor use only
For altitudes up to 2000 m above sea level
Temperature range from 5 °C to 40 °C
Maximum relative humidity: 80 %, for temperatures up to 31 °C, decreasing linearly to
50 % relative humidity at 40 °C
Main power supply: Voltage fluctuations not to exceed +/- 10 % of the nominal value
ELECTRICAL DATA
Overvoltage category: II
Contamination level: 2
WEEE Compliance:
This product is required to comply with the European Union`s Waste Electrical & Electronic
Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Electron has contracted with one or more recycling/disposal companies in each EU
Member State, and this product should be disposed of or recycled through them. Further
information on Thermo Electron`s compliance with these Directives, the recyclers in your
country, and information on Thermo Electron products wich may assist the detection of
substances subject to the RoHS Directive are available at www.thermo.com/WEEERoHS .
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
/

Advertisement

Key features

  • Temperature Range: Room temperature + 10°C to 250°C

  • Programmable temperature control

  • Independent overtemperature protection

  • Continuous use

  • Hot-air sterilization capability

  • Time-switch function

  • Fixed program segments

  • Programmed controller function

Frequently asked questions

No, the oven cannot be used to dry or heat substances that may release combustible gases or vapors into the atmosphere. It is also unsuitable for the heat treatment of combustible dusts or fibrous materials.

You must use a draft breaker to connect the oven to an extraction system. All flue ducts must be made of fireproof materials and thermally insulated.

The oven is equipped with an electronic upper-limit cut-out device (TWB) of protection class 2 as defined by DIN 12 880 Part 1. It is electrically and functionally independent of the temperature control system. If the unit heating system malfunctions, the TWB will shut down the unit heating across all poles as soon as the temperature exceeds the preset value.

The fixed program contains a maximum of four program segments and a status command. Each program segment is assigned to a typical temperature / time response. 10 selectable sets of parameters are permanently stored in the database for each program segment.

The programmed controller function enables the arbitrary definition of temperature / time processes. You may select up to 9 program segments. Operating temperature and period can be selected for each segment.

User questions

E F

Patricia E.

Where can I find the electrical schematic for the T6P oven?

The electrical schematic for the T6P oven is available as an approved spare part or accessory. You can locate the order numbers for the circuit diagrams in the 'Approved spare parts / accessories' section of the manual. For 1/PE AC, 230 V/208V, the order number is 50 041 987. For 1/PE AC, 120 V, the order number is 50 040 640.

U R

James U.

What is the recommended relative humidity for operating the device?

The recommended relative humidity for operation is between 60 and 70%, avoiding condensation. The maximum relative humidity is 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C.

Preparing document for printing…