Adjustable range:

Measured gas

Short gas name

Long gas name see the appropriate sensor data sheet

4 characters

10 characters

Full-scale deflection see the appropriate sensor data sheet

Response factor

Molecular weight

0.1 to 20.0

2 to 250 g/mol

A2

A1 greater than 0, less than full-scale deflection greater than 0, less than A2

Determination of the response factor of a customer-specific gas

 Configure PID sensor to isobutylene (IBUT).

Adjust zero point and sensitivity.

Apply gas to the sensor using customer gas of known concentration

(e.g. 150 ppm) and wait for stable measurement value.

Make a note of the displayed measurement value (e.g. 120 ppm) and determine the response factor as follows.

RF = (desired value / actual value) = (150 ppm / 120 ppm) = 1.25

Turn off the gas flow and set the sensor to the freely configurable gas (VOC1, VOC2,…).

After the short and the long gas names have been entered, enter the response factor of the gas to be measured by pressing the » « and » « keys and confirm the setting by pressing the » « key.

M OK

7.3.5

Adjustable calibration interval

Use this function to enter the calibration interval setting of the sensor.

 Select the menu items »

Interval

Settings \ Sensors

OK

\ CA-CH4 \ Cal.-

« in this order, pressing the » « key after each selection.

 Set the calibration interval using the » « and » « keys and confirm by pressing the » « key.

M

OK

Only one calibration interval setting is entered for each EC, IR and

PID sensor.

7.3.6

Mean value period

This function can be used to set the mean value period for exposure evaluation.

 Select the menu items »

Param.

\ Mean value period key after each selection.

Settings \ Sensors \ EC–CO \ Expo-

« in this order, pressing the »

OK

«

 Select the desired configuration from the list and, if more than one configuration has been enabled for activation, press the » key (» «), see » Select status

« page 69.

OK

«

 Use the » « and » « keys to set the mean value period and

M confirm the setting with the » « key.

OK

Adjustable range: 1 to 60 minutes.