Fault – Cause – Remedy

8 Fault – Cause – Remedy

Fault / message

Not possible to switch on the instrument

The instrument cannot be charged.

Notice: "Instrument failure" or "Sensor has bad contact"

Instrument displays potentially individually incomprehensible characters

Instrument turns on continually

The instrument is defective

Flow too low!

No valid span calibration

Cause

Power pack is fully discharged.

Temperature too high or too low

Power pack is not connected correctly or bad contact in the power-pack connector.

Wrong language selected

Remedy

Charge the power pack.

Note: In this case, the instrument display initially remains blank.

Observe temperature range.

Reconnect the power pack and secure it with the screws.

Select the correct language with the program

Dräger CC Vision. Alternatively, navigate

"blindly" through the menu structure. Language

64).

option)

Instrument \ (last item) Language.

Electronic fault (Watchdog)

An internal fault has been detected.

#2: Instrument fault

#129: Electronic fault

#130: Fault when switching on the flow sensor

Hose is blocked

Switch the instrument off then on again, or remove and reconnect the power pack. If the fault is still displayed, contact DrägerService.

Switch the instrument off then on again, or remove and reconnect the power pack. If the fault is still displayed, contact DrägerService.

Inspect the hose for kinks and blockages.

If necessary, remove the pump adapter and inspect the membrane in the sensor grille for dirt and drops of water.

Ensure that the hose is not kinked or submerged during use.

#9 Pump faulty Replace the pump.

#107: Calibration was executed without test gas Repeat calibration with test gas.

Dräger X-am 7000 73

Fault – Cause – Remedy

Fault / message

Sensor drift

IR-sensor faulty

No valid zero-point calibration

Sensor defective

Function not possible: O

2

deficiency.

Date and time not valid

Data logger 90 % full / data logger full

Sensor warming up

Cause

#104: Long-term drift of the sensor.

Remedy

Carry out fresh air calibration / zero-point calibration.

The instrument has found a fault in the IR sensor.

#118: IR sensor faulty.

Switch the instrument off then on again,

remove and reconnect the power pack.

If necessary, replace the IR sensor. If the fault is still displayed, contact DrägerService.

There is no valid zero-point calibration.

Repeat zero-point calibration with a suitable zero gas.

#106: Zero-point calibration was not completed successfully (e.g. no suitable zero gas used/

CAT sensor was exposed to nitrogen)

Carry out zero-point calibration with a suitable zero gas.

#123: Fault during calibration Repeat the calibration procedure or disconnect and reconnect the power pack. If the fault still exists, change the sensor or inform

DrägerService.

The instrument has detected a fault in the sensor.

#108: Inconsistent sensor data.

#133: Measuring signal is outside of the permitted range.

Switch the device off and on again. If the fault

still displayed on the instrument, contact

DrägerService.

The CatEx sensor needs oxygen for correct operation. If insufficient oxygen is available, the sensor cannot measure correctly.

Use an IR-Ex sensor in atmospheres with O

2 deficiency.

The date and/or time are set to invalid values Set the date and time correctly.

%) full Read out the contents of the data logger

(with

The warming-up time of the sensors may vary, depending on the ambient conditions

Wait until the sensor has warmed up.

74 Dräger X-am 7000

Fault – Cause – Remedy

Fault / message

Increased measuring failure

Sensor near end of life

Sensor not supported

Cause

The measuring error is higher during the warming-up phase. Calibration cannot yet be executed.

Remedy

Info Mode provides further details.

Measurements can still be carried out under these conditions.

The sensor no longer has sufficient reserves for calibration

#100: The sensor is not supported by the instrument.

The sensor cannot be used.

For details, contact DrägerService.

Config error, Channelinit required

No sensor or bad contact No sensor in the active sensor slot Plug in a sensor or deactivate the channel.

#100: The sensor has been removed or there is a bad sensor in the sensor connector

Switch the instrument off and on again.

If necessary, remove the sensor and insert

83.

Temperature out of range, incr. meas. failure The sensor was/is too cold

Temperature out of range, incr. meas. failure The sensor was/is too warm

No valid TC span calibration

Channel error

The CatEx sensor has been calibrated in the

Vol% range (>100 LEL) with a CH

4 concentration which was too low. (e.g. with

2 Vol.-% CH

4

)

An error has been detected in the measuring channel.

#109: An error occurred during loading of the sensor data

The sensor must be calibrated or adjusted.

The sensor must be calibrated or adjusted.

Repeat calibration in the TC range

(>100

50

LEL) with a concentration of about

Vol.-%.

Switch the device off and on again. If necessary, remove and reconnect the power pack. If the fault is still displayed on the instrument, contact

DrägerService.

Dräger X-am 7000 75

Fault – Cause – Remedy

Fault / message

Checksum error: instrument init.

Internal com error with subsystem 3

Invalid fresh air calibration

Invalid mixed gas calibration

Incorrect sensor test

Low IR signal

Cause Remedy

Note: All settings made with the menu will reset to the factory settings.

Carry out instrument initialisation, switch the instrument off and on again or remove and reconnect the power pack.

The data connection to the power pack is opencircuit.

(It may be that the screws are not tightened correctly)

Remove the power pack and connect it correctly.

The air in which fresh-air calibration was carried out was not free of contamination. (e.g.

Repeat fresh-air calibration in clean air.

CO content too high)

N

2

was used for fresh air calibration Repeat fresh-air calibration in clean air.

The results of mixed-gas calibration were invalid

(possibly a sensor at the end of its life)

Inspect the sensors and replace any faulty ones.

Mixed gas calibration for O

2

sensor was carried out with a mixed gas which did not contain O

2

Use a suitable test gas and compare the concentration information on the test gas cylinder with the concentration specified for

The sensor self-test has detected a fault in

sensor

Switch the device off and on again. If the fault

still displayed on the instrument, contact

DrägerService.

The instrument has found a fault in the IR sensor.

#126: IR signal too low

Switch the device off and on again.

83.

If the fault is still displayed on the instrument, contact DrägerService.

76 Dräger X-am 7000

Fault – Cause – Remedy

Fault / message

High IR signal

Sensitivity low. Please check lamp

Perform bump test

Cause

The instrument has found a fault in the IR sensor.

#127: High IR signal

Switch the device off and on again.

83.

If the fault is still displayed on the instrument, contact DrägerService.

#132: Span falls short of warning setpoint

Remedy

# 137: Bump test failed

Clean the lamp on the PID sensor, see sensor data sheet. If the device continues to display this error, replace the lamp of the PID sensor, see sensor data sheet.

A new bump test is required each time the

Automatic bump test does not start exited.

Repeat bump test, ensure that gas is continuously supplied during the test. If required, re-adjust the sensors.

Rectify the channel or instrument fault and repeat the bump test.

Configure the Dräger X-am 7000.

Dräger X-am 7000 were not configured according to Dräger CC Vision

There is no activated printer with correctly positioned IR interface in front of theBumpTest

Station

Switch on the printer and position the IR interface of the printer in front of the fiberoptic cable of the

BumpTest Station.

The manual Bumptest cannot be started via the "Quick Menu"

One of the channels activated for the Bumptest is in WarmUp or Error mode or an instrument fault has occurred

Wait until the WarmUp time has elapsed, rectify the fault and restart the Bumptest.