Service menu. Gima 33520, 33519
Hirdetés
Hirdetés
ICON DESCRIPTION
To test oximetry
To test FVC spirometry
To test VC spirometry
To test MVV spirometry
To run a bronchodilator test (POST)
To check the alarms and thresholds set during an oximetry test
To check the oximetry sensor is properly connected during testing
To check the presence of a signal during the oximetry test
To print a test
To delete a test
To cancel an operation and return to the previous step
2.5
Service menu
To access the menu, press the icon .
To view headings, enter the following code using the virtual numeric keypad:
1 2 2 3 3 3
In this screen, you can also change the brightness of the display by tapping the orange cursor vertically. The brightness will change in real time.
Spirolab
User manual
The service menu will display the following heading sequence:
Device
Spirometry
Calibration
Oximetry
References
Delete archive
Tap the desired heading to select it.
Device
The section is divided into:
General
Date and time
Info
Printer
Select the desired settings by tapping the white box; the selection is highlighted by an orange box.
For the brightness of the display, move the orange cursor vertically until the desired brightness.
To set the correct date and time, tap the box to bring up the cursor; use the numeric keypad on the lower right to set the correct value.
The "Info" section displays the hardware and software versions of the device's most significant components, in addition to the serial number.
In the "Printer" section, you can set the print header on thermal paper. Tap the white space under "print header" to bring up the virtual numeric keyboard. You can enter characters until the string is complete.
Rev.1.3 Page 15 of 34
Spirometry
The section is divided into:
Parameters
Outcome
Incentives
In the “Parameters” section, you can select the values that will be displayed at the end of a test; these will be stated on the printout.
Here too, the selection is highlighted by an orange box.
In the “Results” section, you can choose what to include in the printout from among the following items:
Best values
Interpretation of spirometry results
F/V and V/t curves
Only F/V EXPIRATORY curve
In the "Incentive" section, you can configure the paediatric incentive to improve spirometric performance.
The incentive may be enabled or disabled; should it be enabled, by selecting "ON" on the left, you must enter the age threshold up to which the incentive is displayed; use the numeric keypad on the bottom to set this value.
Calibration
For calibration, see paragraph 2.5.1 below.
Oximetry
In this section, you can configure all alarms related to an oximetry test. By setting the desired alarms to "ON", you will activate a box where you can set the relevant threshold value using the numeric keypad.
Spirolab
User manual
WARNING
If the maximum value of a parameter is less than or equal to the minimum value, the setting process will not proceed. A beep will sound and the device will automatically return to the screen for setting the minimum value.
References
This section allows you to set the theoretical value used between:
Adults Children
ERS
Knudson
USA
ERS
MC-Barcelona
JRS-JAPAN
Knudson
Knudson
Knudson
Zapletal
Zapletal
JRS-JAPAN
Rev.1.3 Page 16 of 34