Safety Devices to DIN12876

Low level switch

Adjustable over-temperature switch (also valid for units with heating)

Classification of Laboratory Thermostats and Baths

II

III

Technical requirement Designation d

Over-temperature c NFL

Adjustable over-temperature cut-off flammable b FL

Adjustable over-temperature cut-off and extra low-level switch a Normally water; other fluids only when they are non-flammable in the event of a single

failure. b The thermal fluid must have a flame point ≥ 65 °C, this means that ethanol can only

be used under constant supervision. c The over-temperature protection can for example be provided by a fluid sensor or

a suitable over temperature switch. d Determined by the manufacturer.

Your temperature control unit is designated a Class III FL.

Mechanical Over-temperature Switch

This temperature control unit is equipped with a mechanical over-temperature switch.

For setting the over-temperature protection refer to chapter setting the overtemperature switch.

10

advertisement

Key Features

  • Advanced functionality with programmer and interface
  • Compact design with small footprint
  • High cooling capacity of 1.8 kW
  • Temperature range from -20 to 120 °C
  • Natural refrigerant for environmentally friendly operation

Related manuals

Frequently Answers and Questions

What is the temperature range of the MPC UC015w?
The MPC UC015w has a temperature range from -20 to 120 °C.
What is the cooling capacity of the MPC UC015w?
The MPC UC015w has a cooling capacity of 1.8 kW.
What type of refrigerant does the MPC UC015w use?
The MPC UC015w uses a natural refrigerant for environmentally friendly operation.