Specification. Additel 673, 686 and 673 Series

Add to My manuals
70 Pages

advertisement

Specification. Additel 673, 686 and 673 Series | Manualzz

1.3

Technical Specifications

Table 5 Technical Specifications

Protection level

Pressure units

Pressure overload

IP67

Pa, hPa,kPa, MPa, psi, bar, mbar, torr , gf/cm2 , kgf/cm2 , inH2O@4°C , inH2O@20°C , mmH2O@4°C , mmH2O@20°C , ftH2O@4°C , ftH2O@4°C ,inHg@0°C , mmHg@0°C ,mtorr, Ib/ft2,tsi,psf, inH2O@68F, ftH2O@68F , Different pressure ranges have different units. Users can custom 5 pressure units.

When the pressure measurement value exceeds 120% FS, the outer circle of the screen will turn red, and flash and the speaker sounds an alarm rhythmically

Measurement speed User can set the measure speed , default 3 times/second

Display digit User can set it to 4, 5 and 6, default is 5

Working environment

Ambient temperature: (-10~50)°C

Relative humidity: < 95% non-condense

Atmospheric pressure: (86~106)kPa

Warm up time 15 Mins

Barometric measure Range: ( 60~110 ) kPa.a

, Accuracy : ±55Pa

Communication

Real-time clock

Storage capacity

BLE ( built-in ), USB ( master and slave ) WIFI ( optional , RS232 ( optional )

Switchable between 24/12 hours and set time zones

8 GB

8

Display

Power supply

Charging

Battery life

Circular TFT LCD capacitive touch screen

1. 6600mAh Lithium-ion rechargeable battery, power consumption around 2.5W

2. Special TYPE-C USB adapter for power supply, maximum charging current 2A (battery power for best accuracy);

Charging time is about 5 hours, provided by dedicated 5V adapter

24V power supply loading will affect the battery life.

Dimension 175mm×118mm×41mm

Weight 715g

Electrical Connections

Circuit, voltage and switches : standard φ4mm ;

DC24Voutput : standard φ4mm ;

Charging : standard TYPE-C USB connection

Table 6 Electrical Specification (only for ADT673)

Current Measurement ±30.0000mA; Accuracy: ±(0.01%RD+0.005%FS); Resolution:0.0001mA; Input impedance: 25Ω.

Voltage

Measurement

Switch Test

±30.0000V; Accuracy: ±(0.01%RD+0.005%FS); Resolution:0.0001V; Input impedance <1MΩ.

Mechanical, NPN and PNP type supported.

Output Internal loop power: 24V±1V, maximum output circuit: 30mA.

9

Loop Transmitter

Measurement

The internal 250ΩHART resistance is switchable , three types of HART transmitter connections are supported:

1) offline calibration mode: internal power supply, internal resistance;

2) communicator: external power supply, external resistance;

3) online calibration mode: external power supply, internal resistance;

1.4

Features

High accuracy pressure measurement, wide range temperature compensation; mA, Voltage and switch measurements; (Only for ADT673)

DC24V power supply, no external power supply is needed to verify the transmitter; (Only for ADT673)

Based on SDC625 full-function HART, HART pressure transmitter can be easily calibrated;

Built-in atmospheric pressure module, providing accurate reference of ambient pressure, convenient for gauge/absolute pressure synthesis;

4, 5, 6 digits display user selectable resolution;

Analog dial indication, which greatly facilitates dial gauges’ calibration;

8 GB large capacity storage;

The powerful UI function, the pioneering use of a circular TFT LCD capacitive touch screen for real-time display, and supports USB, Bluetooth, WIFI or RS232 communication;

Smart ACloud service, you can connect remotely through mobile phone or PC, not limited by space distance (only when equipped with WIFI module);

Large capacity removable rechargeable lithium-ion battery.

10

advertisement