3.2.5 Language. Additel 673, 686 and 673 Series

Add to My manuals
70 Pages

advertisement

3.2.5 Language. Additel 673, 686 and 673 Series | Manualzz

3.2.5

Language

The device provides a multi-language user interface. Use this menu to change from the provided languages.

After the language is selected, the device needs to be restarted for the changes to take effect.

3.2.6 Display Orientation

The direction of display is adjustable for different direction of mounting.

3.3 System Service

3.3.1 System Calibration

The device provides function of calibration. Users need to enter a password to enter the system calibration page, and the factory default password is 123456. The following information is general information of the calibration page.

3.3.1.1 General Information

On the right side of electrical calibration page, there will be following icons:

Table 19 Calibration icon description

Icon Description

Return

Electrical calibration

Pressure module calibration

Restore factory ● ●

Previous point ● ●

Next point ● ●

Barometric calibration

35

Save

Cancel

OK

Cancel zero

Calibration

3.3.1.2 Pressure Calibration

1. Cancel zeroing

Cancel the recorded zero data.

2. Multi-point calibration a. Connection

Connect the calibrator with external pressure source and reference pressure standard. b.

Perform calibration

◆ Click on to perform calibration.

◆ Connect the calibrator with external pressure source.

◆ After the pressure is stabilized, click on to go to next point;

◆ Adjust the gauge output value to the same as the setpoint being calibrated;

◆ If the measured value has big difference with the calibration point, there will be notice by RED;

◆ If the calibration point has big difference with the measured value, there will be a notice, but it does not influence taking effect.

36

◆ New calibration data takes effect immediately. The calibration date will be changed to the current system date. c.

Restore the factory calibration data

Click on to restore the factory calibration data.

3.3.1.3 Barometric Calibration

The calibrator provides single point or dual point calibration to the internal barometric module. Before performing the barometric calibration, remove the back cover and connect the air line to 4mm connector, as shown in the Figure 11.

Figure 11 Barometric calibration

37

1. Single point calibration

Input the external barometric standard to standard pressure value, confirm and click on SAVE to finish calibration.

2. Dual point calibration a.

Connection

Connect the calibrator with external pressure source and provide greater standard of pressure. b. Perform calibration

◆ Click on to perform calibration.

◆ Input the external pressure according to the calibrator’s notice. After the pressure is stabilized, click on to go to next point;

◆ If the measured value has a big difference with the standard value, there will be notice by RED, the result cannot be saved now;

◆ If the calibration point has a big difference with the measured value, there will be a notice, but it does not influence taking effect.

◆ New calibration data takes effect immediately. The calibration date will be changed to the current system date. c. Restore the factory calibration data

Click on to restore the factory calibration data.

3.3.1.4 Electrical Calibration (ADT673 only)

1.

Cancel zeroing

Cancel the recorded zeroing data.

2. Multi- point calibration

◆ Click on to perform calibration.

38