Alarms and Indicators. OTIS OI-315 TOCSIN3
Add to My manuals24 Pages
OTIS OI-315 TOCSIN3 personal monitor is a safety device designed to detect toxic gas in ambient air and alert you with visual, vibrating, and audible alarms. It features adjustable alarm set points, data logging, user-replaceable sensor, LCD display with LED backlight, and is water/dust proof. Wear it on your person in the field to continuously monitor for gas exposure and ensure your safety.
advertisement
Setting Alarms
1.
Power off the device.
2.
To enter Alarm Set Mode, press and hold the blue arrow button for approximately two seconds. The device will enter 1 st Alarm Set Mode.
3.
To change the Low Alarm setting, briefly press the green power button. The display will indicate the current three-digit Low Alarm setting and the first digit will flash. Press the blue arrow button to increase the digit by one (per press).
NOTE: Each unit value can be set between “0” and “9.”
4.
Once the desired value is displayed on the screen, press the green power button to save the alarm value (and move to the 2nd Alarm Set Mode). Repeat these steps for the High Alarm.
5.
To return to Gas Measure Mode, press the blue arrow button once.
Default Alarm Set Points
Gas O
2
CO SO
2
H
2
H
2
S Cl
2
NH
3
NO
2
1 st 19% 30 ppm 2 ppm 100 ppm 10 ppm .5 ppm 25 ppm 3 ppm
2 nd 23% 60 ppm 5 ppm 500 ppm 20 ppm 1 ppm 35 ppm 5 ppm
TWA
STEL
N/A 30 ppm 2 ppm
N/A 200 ppm 5 ppm
N/A
N/A
10
10 ppm .5 ppm 25 ppm 3 ppm
15 ppm 1 ppm 35 ppm 5 ppm
advertisement
Related manuals
advertisement
Table of contents
- 2 Product Overview
- 5 Introduction
- 5 Complete System Diagram
- 5 Complete System
- 6 LCD Display Symbols
- 7 Power On
- 7 Power Off
- 8 Basic Operation
- 8 Gas Measure Mode
- 8 LCD Backlight On/Off
- 8 Viewing Date and Time
- 9 Setting Date and Time
- 10 Alarms and Indicators
- 10 Setting Alarms
- 10 Default Alarm Set Points
- 11 Alarm Displays
- 11 Alarm Indicators
- 11 Returning to Normal Operating Mode Following a Low Alarm
- 12 Returning to Normal Operating Mode Following a High Alarm
- 12 Indication of Peak Value and STEL / TWA Value
- 12 Battery Notifications
- 13 Data Logging (Alarm Data)
- 14 Fresh Air Calibration
- 15 Standard Gas Calibration
- 17 Battery Replacement
- 18 Sensor Element Replacement
- 20 User Notice
- 20 Approval Label Notes
- 21 Specifications