Battery Replacement. OTIS OI-315 TOCSIN3
Add to My manuals24 Pages
OTIS OI-315 TOCSIN3 personal monitor is a safety device designed to detect toxic gas in ambient air and alert you with visual, vibrating, and audible alarms. It features adjustable alarm set points, data logging, user-replaceable sensor, LCD display with LED backlight, and is water/dust proof. Wear it on your person in the field to continuously monitor for gas exposure and ensure your safety.
advertisement
Battery Replacement
When the battery level is low (one bar remains on the battery icon), an alarm will sound repeatedly at five minute intervals. When the battery is nearly exhausted, a unique combination of audible, LED, and LCD backlight alerts will display for approximately 10 seconds (before the unit turns off completely).
Required Items:
• Phillips-head Screwdriver
• Replacement Battery = SB-AA02 (Lithium ½ AA Battery, 3.6V/1.2Ah)
NOTE: Replacement of the battery in dangerous regions with potential for explosion is strictly prohibited. Replacement of components can hinder intrinsic safety function.
NOTE: Replacement battery kits should be purchased from the same retailer that the device was originally purchased from.
NOTE: The device must be recalibrated following battery replacement.
NOTE: Before disassembling, power off the device according to the instructions in this operation manual.
1
2
6
7
8
3
4
5
9
Label
Calibration cap
Front cover
Sensor element
PCB (terminal board)
Battery
Rear cover
Belt clip
Machine screw
17
advertisement
Related manuals
advertisement
Table of contents
- 2 Product Overview
- 5 Introduction
- 5 Complete System Diagram
- 5 Complete System
- 6 LCD Display Symbols
- 7 Power On
- 7 Power Off
- 8 Basic Operation
- 8 Gas Measure Mode
- 8 LCD Backlight On/Off
- 8 Viewing Date and Time
- 9 Setting Date and Time
- 10 Alarms and Indicators
- 10 Setting Alarms
- 10 Default Alarm Set Points
- 11 Alarm Displays
- 11 Alarm Indicators
- 11 Returning to Normal Operating Mode Following a Low Alarm
- 12 Returning to Normal Operating Mode Following a High Alarm
- 12 Indication of Peak Value and STEL / TWA Value
- 12 Battery Notifications
- 13 Data Logging (Alarm Data)
- 14 Fresh Air Calibration
- 15 Standard Gas Calibration
- 17 Battery Replacement
- 18 Sensor Element Replacement
- 20 User Notice
- 20 Approval Label Notes
- 21 Specifications