Sensor Element Replacement. OTIS OI-315 TOCSIN3
Add to My manuals24 Pages
OTIS OI-315 TOCSIN3 personal monitor is a safety device designed to detect toxic gas in ambient air and alert you with visual, vibrating, and audible alarms. It features adjustable alarm set points, data logging, user-replaceable sensor, LCD display with LED backlight, and is water/dust proof. Wear it on your person in the field to continuously monitor for gas exposure and ensure your safety.
advertisement
1.
Remove the five machine screws located on the rear cover of the device.
2.
Separate the rear cover from the front cover by gently pulling them away from each other.
3.
Using your thumb and forefinger, pull the battery out of the terminal board.
4.
Slide a new battery into the terminal board.
5.
Verify that the “+” and “-” sides of the battery are facing the correct direction
(and touching the battery contacts on the terminal board).
6.
Place the front and rear covers together, then replace the five machine screws.
Sensor Element Replacement
Sensor replacement is recommended every two years (or whenever the “X” icon appears on the display indicating that either the automatic sensor test on start-up, or the “fresh air” calibration process, has failed).
Required Items:
• Phillips-head Screwdriver
• Replacement Sensor Element = Senko SS series
• Replacement Filters
NOTE: Replacement of components can hinder intrinsic safety function.
NOTE: Replacement sensor elements should be purchased from the same retailer that the device was originally purchased from.
NOTE: The device must be recalibrated following battery or sensor element replacement.
NOTE: Before disassembling, power off the device according to the instructions in this operation manual.
18
4
5
6
7
1
2
3
8
9
Label
Calibration cap
Front cover
Sensor element
PCB (terminal board)
Battery
Rear cover
Belt clip
Machine screw
1.
Remove the five machine screws located on the rear cover of the device.
2.
Separate the rear cover from the front cover by gently pulling them away from each other.
3.
Carefully separate the PCB (terminal board) from the front cover.
4.
Using your thumb and forefinger, gently pull the sensor element off of the terminal board.
5.
Align the sensor element prongs of a new sensor element with the eyelets on the terminal board. Slide the senor element into place.
6.
Place the terminal board back into the front cover.
7.
Place the front and rear covers together, then replace the five machine screws.
19
advertisement
Related manuals
advertisement
Table of contents
- 2 Product Overview
- 5 Introduction
- 5 Complete System Diagram
- 5 Complete System
- 6 LCD Display Symbols
- 7 Power On
- 7 Power Off
- 8 Basic Operation
- 8 Gas Measure Mode
- 8 LCD Backlight On/Off
- 8 Viewing Date and Time
- 9 Setting Date and Time
- 10 Alarms and Indicators
- 10 Setting Alarms
- 10 Default Alarm Set Points
- 11 Alarm Displays
- 11 Alarm Indicators
- 11 Returning to Normal Operating Mode Following a Low Alarm
- 12 Returning to Normal Operating Mode Following a High Alarm
- 12 Indication of Peak Value and STEL / TWA Value
- 12 Battery Notifications
- 13 Data Logging (Alarm Data)
- 14 Fresh Air Calibration
- 15 Standard Gas Calibration
- 17 Battery Replacement
- 18 Sensor Element Replacement
- 20 User Notice
- 20 Approval Label Notes
- 21 Specifications