User Notice. OTIS OI-315 TOCSIN3
Add to My manuals24 Pages
OTIS OI-315 TOCSIN3 personal monitor is a safety device designed to detect toxic gas in ambient air and alert you with visual, vibrating, and audible alarms. It features adjustable alarm set points, data logging, user-replaceable sensor, LCD display with LED backlight, and is water/dust proof. Wear it on your person in the field to continuously monitor for gas exposure and ensure your safety.
advertisement
![User Notice. OTIS OI-315 TOCSIN3 | Manualzz User Notice. OTIS OI-315 TOCSIN3 | Manualzz](http://s1.manualzz.com/store/data/073506704_1-3208840cc34dd38d0be1c5c63425cb7d-360x466.png)
User Notice
The OI-315 should be used in the range of the applicable temperature, humidity and pressure that are listed in the Specifications section of this operation manual. Using the device beyond this range may cause the device to malfunction.
Gas concentration measurements by the sensor can vary according to the environment
(temperature, pressure and humidity). Therefore, calibration of the OI-315 should be performed in the same (or similar) environment of the device’s actual use.
If t h e temperature changes sharply during use of the instrument (e.g., indoors vs. outdoors), the value of the measured gas concentration can suddenly change. Please use the
OI-315 after the gas concentration value has stabilized.
Severe vibration or shock to the device may cause a sudden reading change. Please use the
OI-315 after the value of gas concentration has stabilized. Excessive shock to the OI-315 can cause the device and/or sensor to malfunction.
Approval Label Notes
SENKO
0575
DNV A EX
K -GA2BO-
II 1 G
--20C ≤ Ta ≤ +50C in use nin ion ead manual i
NKO Ltd
Osanyeonggio
-
Oes -don
S/N:
CE marking: Electromagnetic Compatibility
(Directive 89/336/EEC)
Explosion Protection
(Directive 94/9/EEC)
KS marking: Korean Gas Safety for
Explosion Protection
This device is intended to be used in hazardous area Zone 0 within a temperature range of
-20 C to +50 C, where gases of explosion group IIC and temperature class T4 may be present.
S/N: Serial Number
20
advertisement
Related manuals
advertisement
Table of contents
- 2 Product Overview
- 5 Introduction
- 5 Complete System Diagram
- 5 Complete System
- 6 LCD Display Symbols
- 7 Power On
- 7 Power Off
- 8 Basic Operation
- 8 Gas Measure Mode
- 8 LCD Backlight On/Off
- 8 Viewing Date and Time
- 9 Setting Date and Time
- 10 Alarms and Indicators
- 10 Setting Alarms
- 10 Default Alarm Set Points
- 11 Alarm Displays
- 11 Alarm Indicators
- 11 Returning to Normal Operating Mode Following a Low Alarm
- 12 Returning to Normal Operating Mode Following a High Alarm
- 12 Indication of Peak Value and STEL / TWA Value
- 12 Battery Notifications
- 13 Data Logging (Alarm Data)
- 14 Fresh Air Calibration
- 15 Standard Gas Calibration
- 17 Battery Replacement
- 18 Sensor Element Replacement
- 20 User Notice
- 20 Approval Label Notes
- 21 Specifications