4.3.1.6 Maximum Z verification test

This test checks that the needle does not collide with the bottles on the rack tray. Select the rack type (reagent, paediatric, 30 mm or 15 mm), the position of the rack on the tray and the position of the bottle or well on the rack.

Press the Start button to move to the selected position and check if the needle collides with the bottle or well or if there is space between the needle and the bottle.

Repeat the process in the positions required by the user.

4.3.2. Diaphragm pumps and electrovalves test

tested and the test that is to be carried out. The devices that can be tested independently are:

-

-

-

The analyzer uses a 3-way electrovalve to manage the dispensing operations. The washing system of the needle uses a 2-way electrovalve and two diaphragm pumps. The screen makes it possible to choose the device to be

3-channel electrovalve of the dispensing pump.

2-channel electrovalve of the washing system.

Washing system diaphragm pumps

71

advertisement

Key Features

  • Throughput of up to 150 tests per hour
  • Performs a wide range of clinical chemistry assays
  • Compact and easy-to-use
  • Advanced features and capabilities

Related manuals

Frequently Answers and Questions

What is the throughput of the biosystems A15?
The biosystems A15 has a throughput of up to 150 tests per hour.
What types of assays can the biosystems A15 perform?
The biosystems A15 can perform a wide range of clinical chemistry assays, including routine chemistries, immunoassays, and specialized tests.
Is the biosystems A15 easy to use?
Yes, the biosystems A15 is a compact and easy-to-use analyzer, making it ideal for small to medium-sized laboratories.

advertisement

Table of contents