5.1.4.6. Changing the rotor motor. biosystems A15

Add to My manuals
131 Pages

advertisement

5.1.4.6. Changing the rotor motor. biosystems A15 | Manualzz

1

2

5.1.4.6. Changing the rotor motor

a) Remove the rotor completely. Remove the insulation.

b) Remove the rotor covers.

c) Remove the nuts (1).

d) Remove the motor from below. Disconnect the hose.

e) Fit the new motor on the unit.

f) Perform the miss test on the rotor motor and the rotor positioning test in the service program to check that the gap between gears is correct and that the system functions correctly.