GSS GSS.remote RLC 8 Assembly Instruction Manual

Add to My manuals
20 Pages

advertisement

GSS GSS.remote RLC 8 Assembly Instruction Manual | Manualzz

RLC 8

C ontents

1 Safety regulations .......................................................................................3

2 General information ....................................................................................3

2.1 Packing contents ............................................................................3

2.2 Meaning of the symbols used ..........................................................3

2.3 Technical data ...............................................................................4

2.4 Description ...................................................................................5

2.5 PC / laptop system requirements .....................................................6

2.6 Overview of the management unit ...................................................7

3 Installation ..................................................................................................8

3.1 Potential equalisation (PE) ...........................................................8

3.2 Assembling a connection cable .......................................................9

3.3 Connecting components .................................................................9

4 Installing the software on a PC or laptop ....................................................10

3.1 Key Code (Activation Code) for the software ...................................10

3.2 Installing the software ...................................................................10

5 Connecting to a network ............................................................................14

5.1 LAN configuration .......................................................................14

5.2 Check the configuration at the control unit of the head-end station .....16

6 Firmware update .......................................................................................17

7 Remote access via Internet .........................................................................18

- 2 - RLC 8

1 s afety regulations

Assembly, installation and servicing should be carried out by authorised

electricians.

• Observe the safety regulations and notes in the assembly instructions for the cassettes and head-end station used as well as the relevant safety instructions e.g the standards EN/DIN EN 50083 resp. IEC/EN/DIN EN 60728.

Take action to prevent static discharge when working on the device!

2 g eneral information

2.1 P aCking Contents

1 RLC 8

1 network cable 2 m

8 RS-232 plug (9-pin Sub D)

8 RS-232 sockets (9-pin Sub D)

1 ribbon cable (9-pin)

1 Power supply unit

1 Brief assembly instruction

2.2 m eaning of the symbols used

—>

Important note

General note

Performing works

- 3 - RLC 8

2.3 t eChniCal data

The devices meet the following EU directives:

2011/65/EU, 2014/30/EU, 2014/35/EU

The product fulfils the guidelines and standards for CE labelling (page 30).

Connections

LAN 1000-BASE-T ............................................................. 1 RJ-45 socket

RS-232: .................................................................... 8 Sub D 9-pin plugs

DC socket 2,1 / 5,5 mm for connecting up to 8 head-end stations and/or a monitoring cassette PSCU 6000/HSCU 6000

............................................................1

Netzwork ports:

TCP port 60002 (alternative 35000 – 60100, 61000 – 65000)

HTML port 80

DNS port 53 UDP/TCP

Monitoring cassette PSCU 6000 / HSCU 6000

The following items are signalled:

– Level of the output signal measured via PSCU 6000 / HSCU 6000

– BIT – error rate at QAM output signals

– Present line synchronizing signal at analogous PAL output signals

– Present RDS station name at FM output signals

– Quality of the input signal (good, poor and no signal) at QPSK-PAL double transmodulators as well as QPSK- QAM transmodulators

Remote control:

– Remote control is possible at all current cassettes and head-end stations of the Standard- and Profi Line.

Required index levels for remote control:

PSW 1000 index level: ≥ V65

BE-Remote index level: ≥ V45

- 4 - RLC 8

2.4 d esCriPtion

The RLC 8 Management System will be connected to a PC or a network via the

LAN interface and enables communication between the PC and the connected head-end stations (LAN <-> 8 x RS232). Therefore the Windows software

PSW 1000 ist needed.

A connection to the Internet (and thereby remote control) is possible via a router.

Hardware concept:

- 5 - RLC 8

The basic units can be configured and remote controlled via Ethernet with software PSW 1000.

The most current version can be downloaded from "www.mygss.eu".

A key code (activation code) is required for the activation of the PSW 1000 software.

This can be obtained from your regional authorised distributor.

The RLC 8 has 8 RS-232 interfaces (Sub D connectors/male) to connect up to

8 headend stations (control units "BE-Remote") to a PC/network. The monitoring cassette PSCU 6000/HSCU 6000 is also supported.

Error messages from the PSCU 6000/HSCU 6000 monitoring cassette can be forwarded by e-mail.

Required index levels for remote control and remote updates:

– PSW 1000 index level: ≥ V65

– BE-Remote index level: ≥ V45

2.5 PC / l aPtoP system requirements

System requirements for the PSW 1000 software:

– PC or laptop with a Pentium processor,

– Windows 95*/98*/ME/2000/Vista/7 (*from Internet Explorer 5 on),

– At least 32 MB RAM, at least 50 MB free space on hard drive,

– LAN interface (RJ 45 socket)

– Internet access for downloads and remote maintenance.

- 6 - RLC 8

2.6 o verview of the m anagement unit

1 2 3 4 1

5 6

1

Mounting support

2

Potential equalisation screw

3

RS-232 interfaces

Connectors for the head-end stations or a PSCU 6000/HSCU 6000

LEDs "Rx":

LEDs "Tx":

4

DC socket +5 V

5

LAN socket

6

USB sockets (not used)

- 7 - RLC 8

3 i nstallation

– The device should only be installed in a room where the permissible ambient temperature range (0 °C … +40 °C ) can be maintained, even during fluctuations in climatic conditions.

– Mount the device on a non-combustible base.

– Use mounting material suitable for the wall properties.

• Attach four mounting screws at the installation site. The drilling distances are shown in the drawing below.

1 2 1

218 mm

• Hang the unit with the mounting supports 1 on the 4 screws

• Tighten the screws.

3.1 P otential equalisation

(Pe)

Make a potential equalisation (PE) according to IEC/EN/DIN EN 50083 resp.

IEC/EN/DIN EN 60728.

• Connect the PE connection terminal 2 to a PE rail (supplied by customer) using a PE wire (Cu 4mm2 - 9mm2).

- 8 - RLC 8

3.2 a ssembling a ConneCtion Cable

External components are connected to the management unit with 1:1 RS-232 cables. You can assemble these cable in the required lengths by using the supplied RS-232 plugs and the ribbon cable.

9 9

5 5

1 1

6

1:1- RS-232 cable

6

Plug Socket

• Cut off a section of cable to the required length.

• Insert the ends of the cable into the openings on the RS-232 plug and

RS-232 socket, making sure they are pushed in all the way.

• Firmly close the sliders on the blade contacts.

3.3 C onneCting ComPonents

• Connect up to 8 head-end stations (7 head-end stations/1 monitoring cassettes PSCU 6000/HSCU 6000) to the RS-232 interfaces made 1:1 RS-232 cables.

3 by using self-

• Connect the attached power supply unit to DC socket 4 .

After about 1 minute the RLC 8 is ready to use.

A combination with HRCU 8/PRCU 8 is not possible!

- 9 - RLC 8

4 i nstalling the software on a

PC or laPtoP

The basic units can be configured and remote controlled via Ethernet with software PSW 1000.

The most current version can be downloaded from "www.mygss.eu".

3.1 k ey

C ode

(a

Ctivation

C ode

) for the soft ware

A key code is required for the activation of the PSW 1000 software. This can be obtained from your regional authorised distributor.

3.2 i nstalling the soft ware

The PSW 1000 software can be downloaded from "www.mygss.eu".

If you do not have Internet access, we will send you a DVD on request.

• Unzip the "PSW1000_Vxx.zip" file and start the

"setup_PSW1000_Vxx.exe" programme by a double click.

• Select the desired language and click the OK button to confirm.

• Select the desired language and click the OK button to confirm.

- 10 - RLC 8

• Start the setup wizard using button "Next >".

• Read the license agreement.

• If you accept the license agreement select "I accept the agreement" and click the "Next >" button.

—> Keep the software version of the PSW 1000 always up-to-date in order to be able to remote control also the newest products.

—> After installing the PSW 1000 soft ware, update the software for the cassettes if necessary.

• Click the "Next >" button.

- 11 - RLC 8

• Specify the directory in which the PSW 1000 should be installed

(e.g. C:\Programme\GSS\PSW1000).

• Click the "Next >" button.

• Enter a name for the shortcut to the programme which will be created in the start menu.

• Click the "Next >" button.

- 12 - RLC 8

• Click on "Install" in order to proceed with the installation of the programme, or on "Back" to make corrections or changes.

—> The installation progress is shown.

• Click on "Finish", to complete the installation.

- 13 - RLC 8

5 C onneCting to a network

5.1 lan

Configur ation

The "static" (fixed) IP address 192.168.0.120 and Port 60002 is set to the

RLC 8.

If another IP address range is used in your network or this IP address is already in use, first the IP address of the RLC 8 must be changed. If several RLC 8 are used in the network, different IP addresses must be assigned for the RLCs:

• Set a static IP address to your PC which is in the address range of the RLC

8 (e.g. 192.168.0.2).

• Connect your PC directly to the RLC 8 via a LAN cable.

• Start any internet browser.

• Enter the IP address of the RLC 8 (http://192.168.0.120) at the address line of the browser and press the "Enter" button:

• Enter the user name "admin" and the password ("geheim" is the standard value) and click the "Login" button.

- 14 - RLC 8

• Enter the settings necessary for your network.

Use only ports in the range of 35000 – 60100 or 61000 – 65000!

On request change the password.

Note the settings!

• Click the "Apply" button in order to save the changes.

The changed settings are transmitted to the RLC 8.

The RLC 8 will be rebooted.

Now the RLC 8 is configured for your LAN network.

• Disconnect the LAN connection to your PC.

• Connect the RLC 8 to your network.

- 15 - RLC 8

5.2 C heCk the Configur ation at the Control unit of the head

end station

• Activate the "System information" software by simultaneously holding down the buttons and for more than 2 seconds in the "Selecting the cassette" mode.

—> The display darkens.

Afterwards the "System informatio n – control unit" –

"SYSTEM INFO: BE_Remote" menu appears.

—> The "System information" software also can be called up from the standby mode of the display.

SYSTEM

BE-Remote

INFO:

V 45

• Press the button so many times until the network configuration is shown in the display of the RLC 8.

REMOTE IP-ADDR

1 9 2 . 1 6 8 .

0 0 0 . 1 2 0

• Press the button.

REMOTE IP-MASK

2 5 5 . 2 5 5 . 2 5 5 .

0 0 0

• Press the

• Press the

button.

REMOTE GATEWAY

1 9 2 . 1 6 8 .

0 0 0 .

0 0 1

button.

REMOTE

60002

IP-PORT

- 16 - RLC 8

6 f irmware uPdate

• Enter the IP address of the RLC 8 (default http://192.168.0.120) at the address line of the browser and press the "Enter" button:

• Enter the user name "admin" and the password ("geheim" is the standard value) and click the "Login" button.

In section Firmware the current firmware version is indicated.

• Check whether a newer version is available on www.mygss.eu.

• Download and unpack it to your PC.

• Click on button , select the stored file and click on button "Open".

- 17 - RLC 8

7 r emote aCCess via

i nternet

To get remote access via internet the router connected to the RLC 8 must be connected to the internet. In addition the "public" IP address with which the router is connected to the internet must be known.

—> Therefore you either need a "fixed", "static" public IP address, or a DNS account which you can get from respective providers of dynamic DNS services (e.g. DynDNS.com).

For remote maintenance via internet "Port Forwarding" must be configured at the router.

—> Pay attention for the operating instruction of the router.

—> If several RLC 8 are used, for each port used "Port Forwarding" must be configured.

—> To get remote access via internet you need a PC with internet access

• Start the PSW 1000 software.

—> The installation of the PSW 1000 software is described in chapter

4 (page 9).

• Click the button.

—> The "Connection settings" window is activated.

• Select tab "Ethernet" and enter

• Select tab "Ethernet" and enter at a connection at a connection via the Internet the "external"

(public) IP address of the router or its "dynamic

DNS account" and the port of the router, for which port forwarding to the management system is configured e.g. IP 212.20.172.0 and port

59999

.

• If a password was set before enter the password

(case-sensitive).

• Click on button "OK".

- 18 - RLC 8

If no password is set, the following message appears:

• Click on button "OK".

—> The connecting button changes from to .

If you have entered a wrong password, the following message appears:

• Enter the password (case-sensitive).

—> The connecting button changes from to .

—> The password setting is to be done in menu Plant settings > Security

(PSW 1000).

- 19 - RLC 8

Declaration of CE conformity

GSS Grundig Systems GmbH • Beuthener Straße 43 • D-90471 Nuremberg

Phone: +49 (0) 911 / 703 8877 • Fax: +49 (0) 911 / 703 9210 www.gss.de/en • [email protected]

Service: Phone: +49 (0) 911/703 2221; Fax: +49 (0) 911/703 2326; [email protected]

Alterations reserved. Technical data E. & O.E. © by GSS Grundig Systems GmbH 001/2017