Operation: Respiration Profiles
Entering I:E Ratio
• The I:E ratio is the ratio of the time needed for inhalation compared to time needed for exhalation.The volume of the inhalation breath is always the same as the volume of the exhalation breath.
• There are no predetermined safe range values for I:E ratios.
• The default I:E ratio is 50%
To enter a different I:E ratio press the [SET] key followed by the [I:E RATIO] key. The display should now read “Enter % Inspiration, (20%-80%)”. Use the numeric keypad to enter the whole number percentage of inspiration (which must be between 20 and 80). The percent of expiration is determined from the value for inspiration. (Note: Due to electromechanical system limitations, Inspira might not accept some I:E Ratio values for certain respiration profiles. If a desired I:E ratio is not accepted when using the small cylinder, try switching to the larger cylinder. This may allow acceptance of the desired ratio.)
Step Key Display
Publication 5404-001-REV-C
advertisement
Related manuals
advertisement
Table of contents
- 5 Warranty and Repair Information
- 6 General Safety Summary
- 7 Theory of Operation
- 7 Features
- 9 Initial Setup and Location Requirements
- 9 Typical Operation
- 10 Airway Tubing Connections: Front Panel
- 11 Electrical Connections: Front Panel
- 12 Electrical Connections: Rear Panel
- 13 Air Cylinder Installation
- 14 Fuse Replacement
- 15 Turning Ventilator On
- 16 Main Operating Screens
- 17 Key Functions
- 18 Respiratory Profiles
- 20 Entering Animal Weight
- 21 Starting and Stopping Ventilation
- 21 Displaying Ventilation Parameters
- 22 Warnings and Faults
- 24 Entering Respiration Rate
- 24 Entering Tidal Volume
- 25 Entering Sigh Breath
- 25 Disabling Sigh Breath
- 27 Entering I:E Ratio
- 28 Inspiration Pressure
- 29 Assist Pressure
- 30 Inspiration Pressure
- 31 Assist Pressure
- 31 PEEP Pressure
- 32 Sigh Pressure
- 33 Volume Mode
- 33 Pressure Mode
- 33 Assist Mode
- 34 User Override
- 34 Using Digital and Analog I/O Ports
- 35 Remote Control/Monitoring (Serial Communications)
- 38 Changing Air Cylinders
- 39 Hardware Setup
- 41 Diagnostics
- 46 Power Failure
- 46 General Maintenance and Cleaning
- 47 General Specifications
- 48 I/O Specifications
- 49 Serial Commands, Queries and Responses
- 51 Glossary of Terms
- 52 Sample Waveforms
- 53 Frequently Asked Questions
- 55 Recommended Cylinder Size
- 56 Recommended Port Size
- 57 Using Inspira with Pressurized and Anesthetic Gasses
- 58 Generating PEEP with Inspira ASVv