Operation: Respiration Profiles
Entering Pressure Settings (Inspira ASVv)
Inspiration Pressure
The Inspiration pressure is a pressure threshold, which is used to detect airway overpressure conditions. If the average airway pressure rises above this pressure threshold, an alarm is sounded (if enabled) and the display shows the fault.This indicates that the residual volume may be increasing in the animal’s lungs and user intervention is required immediately. Ventilation will still occur after this threshold has been broken.
The default Inspiration Pressure is 35 cmH2O.To set the inspiration pressure press the
[Set] key followed by the [Pressure] key. The display should now display a list of options “Press 1=Insp, 2=Asst”. Press the [1] on the numeric keypad to select the inspiration pressure. The display should now read “Enter Insp Pressure” and display the current pressure setting. Using the numeric keypad, enter the inspiration pressure setting in cmH2O. Inspira will only accept whole number values for this setting. Press [Enter] to enter the selected pressure setting.
Step Key Display
Publication 5404-001-REV-C
advertisement
Related manuals
advertisement
Table of contents
- 5 Warranty and Repair Information
- 6 General Safety Summary
- 7 Theory of Operation
- 7 Features
- 9 Initial Setup and Location Requirements
- 9 Typical Operation
- 10 Airway Tubing Connections: Front Panel
- 11 Electrical Connections: Front Panel
- 12 Electrical Connections: Rear Panel
- 13 Air Cylinder Installation
- 14 Fuse Replacement
- 15 Turning Ventilator On
- 16 Main Operating Screens
- 17 Key Functions
- 18 Respiratory Profiles
- 20 Entering Animal Weight
- 21 Starting and Stopping Ventilation
- 21 Displaying Ventilation Parameters
- 22 Warnings and Faults
- 24 Entering Respiration Rate
- 24 Entering Tidal Volume
- 25 Entering Sigh Breath
- 25 Disabling Sigh Breath
- 27 Entering I:E Ratio
- 28 Inspiration Pressure
- 29 Assist Pressure
- 30 Inspiration Pressure
- 31 Assist Pressure
- 31 PEEP Pressure
- 32 Sigh Pressure
- 33 Volume Mode
- 33 Pressure Mode
- 33 Assist Mode
- 34 User Override
- 34 Using Digital and Analog I/O Ports
- 35 Remote Control/Monitoring (Serial Communications)
- 38 Changing Air Cylinders
- 39 Hardware Setup
- 41 Diagnostics
- 46 Power Failure
- 46 General Maintenance and Cleaning
- 47 General Specifications
- 48 I/O Specifications
- 49 Serial Commands, Queries and Responses
- 51 Glossary of Terms
- 52 Sample Waveforms
- 53 Frequently Asked Questions
- 55 Recommended Cylinder Size
- 56 Recommended Port Size
- 57 Using Inspira with Pressurized and Anesthetic Gasses
- 58 Generating PEEP with Inspira ASVv