4

Viewing the stored values

Press the M-button 8 briefly, when the device is switched off. The display first shows « M » AO and an average value. The device then switches to the last stored value.

Pressing the M-button again displays the previous value. Pressing the M-button repeatedly enables you to move from one stored value to another.

Memory full

 Pay attention that the maximum memory capacity of 30 memories is not exceeded. When the 30 memory is full, the oldest value is automatically overwritten with the 31st value.

Values should be evaluated by a doctor before the memory capacity is reached – otherwise data will be lost.

Clearing all values

If you are sure that you want to permanently remove all stored values, hold down the M-button (the device must have been switched off beforehand) until « CL » appears and then release the button. To permanently clear the memory, press the M-button while « CL » is flashing. Individual values cannot be cleared.

How not to store a reading

As soon as the reading is displayed press and hold the ON/OFF button 1 until « M » AO is flashing. Confirm to delete the reading by pressing the M-button

8

.

6. Battery Indicator and Battery change

Low battery

When the batteries are approximately ¾ empty the battery symbol AN will flash as soon as the device is switched on (partly filled battery displayed). Although the device will continue to measure reliably, you should obtain replacement batteries.

Flat battery – replacement

When the batteries are flat, the battery symbol AN will flash as soon as the device is switched on (flat battery displayed). You cannot take any further measurements and must replace the batteries.

1. Open the battery compartment 5 on the bottom of the device.

2. Replace the batteries – ensure correct polarity as shown by the symbols in the compartment.

Which batteries and which procedure?

 Use 4 new, long-life 1.5 V, size AA alkaline batteries.

Do not use batteries beyond their date of expiry.

Remove batteries if the device is not going to be used for a prolonged period.

Using rechargeable batteries

You can also operate this device using rechargeable batteries.

 Only use «NiMH» type reusable batteries.

Batteries must be removed and recharged when the flat battery symbol appears. They should not remain inside the device as they may become damaged (total discharge as a result of low use of the device, even when switched off).

 Always remove the rechargeable batteries if you do not intend to use the device for a week or more.

 Batteries cannot be charged in the blood pressure monitor.

Recharge batteries in an external charger and observe the information regarding charging, care and durability.

7. Using a Mains Adapter

You can operate this device using the Microlife mains adapter

(DC 6V, 600 mA).

 Only use the Microlife mains adapter available as an original accessory appropriate for your supply voltage.

 Ensure that neither the mains adapter nor the cable are damaged.

1. Plug the adapter cable into the mains adapter socket 4 in the blood pressure monitor.

2. Plug the adapter plug into the wall socket.

When the mains adapter is connected, no battery current is consumed.

8. Error Messages

If an error occurs during the measurement, the measurement is interrupted and an error message, e.g. « ERR 3 », is displayed.

Error Description Potential cause and remedy

« ERR 1 » Signal too weak

The pulse signals on the cuff are too weak.

Re-position the cuff and repeat the measurement.*

« ERR 2 » Error signal During the measurement, error signals were

« ERR 3 » No pressure in the cuff detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.

An adequate pressure cannot be generated in the cuff. A leak may have occurred.

Check that the cuff is correctly connected and is not too loose. Replace the batteries if necessary. Repeat the measurement.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • Systolic and diastolic blood pressure measurement
  • Pulse rate measurement
  • Heart arrhythmia indicator
  • Memory for 30 readings
  • Traffic light indicator for easy interpretation of results
  • Mains adapter option

Related manuals

Frequently Answers and Questions

Who should use this device?
It is suitable for home use by adults who want to monitor their blood pressure.
How often should I measure my blood pressure?
It is recommended to take at least two measurements per day, one in the morning and one in the evening.
What does the heart arrhythmia indicator mean?
It indicates that an irregular heartbeat was detected during the measurement. If you see this indicator, you should consult with your doctor.
How do I store my readings?
The device has a memory that can store up to 30 readings.
Can I use this device with a mains adapter?
Yes, the device can be used with a mains adapter (optional).

advertisement

Table of contents