And The TOXTRAP Holder. CMI/MPH INTOXILYZER 5000
Advertisement
Advertisement
Regardless of which mode sequence the Intoxilyzer 5000R breath analysis instrument is set·. to perform, activating 56 Sample capture Option) programs the instrument to do three things
1.
2.
Request the attachment of .a
collector device following completion of each breath sample analysis.
Request the detachment completion analysis.
of the of purge the collector device .following
operation followinq each sample
3.
Print "SAMPLE CAPTURE RE!JUESTED" on the evidence card.
TOXTRAP collector tubes' and a TOXTRAP Holder are accessory items available for purchase.
TOXTRAP tubes can be purchased in lots
'of 100 by ordering Part # 015014.
purchasedby ordering Part # 011109.
The TOXTRAP Holder can be
(See "ACCESSORIES," page
61.) The following diagrams and text show the TOXTRAP tube as well as the TOXTRAP Holder and give instruction on their use.
Instructions For Using The Collector Tube
And The Toxtrap Holder
TOXTRAP collector tubes are individually wrapped in waterproof plastic bags.
The labeling on each plastic bag is important because it gives the name of the breath analysis instrument the tub~was designed to be used with.
Therefore, prior to using a
TOXTRAP collector tube, make sure the label on the bag reads "USE
WITH THE INTOXILYZER 5000."
Two brass exhaust ports extend from the back of the instrument.
On a standard instrument, the ports are' labeled "PUMP
.EXHAUST /sAMPLE CAPTURE" and "BREATH EXHAUST.
n
On. an ~ instrument equipped with the Vapor Recirculation option, the ports are labeled "SIMULATOR RETURN"and "EXHAUST/SAMPLE CAPTURE.
I I
48
j
For clarity, the ·following instructions will refer to the right and left side of. the TOXTRAF Holder as indicated below.
Right Sid.
,
Lett Side
Preliminary Hookup:
1.
2.
Make sure all connectors are properly matched.
On a standard instrument, before beginning a: breath sample analysis, . insert a new mouthpiece in the end of the breath tube and connect the tubing on the right side of the Toxtrap
Holder to the instrument's "PUMP EXHAUST/SAMPLE CAPTURE" port.
I
5000 Series Instrument Without Recir¢ulation
PUMP EXHAUSTISAMPLE CAPTURE port
.49
On an instrument equipped with the Vapor Recirculation option, connect the tubing on the right side of the TOXTRAP
Holder to the "EXHAUST/SAMPLE CAPTURE" port.
SQQQseries Instrument With Recirculation
EXHAUST/SAMPLE· CAPTURE port
SIMULATOR RETURN port
When the instrument displays a command requesting you to attach the collector device, complete the following steps:
1.
Make sure the label on the bag enclosing the TOXTRAP collector tube reads "USE WITH THE INTOXILYZER SQQQ SERIES"; then remove the tube from the bag and the end caps from the tube.
~----------
...... --+1__
50
-\
I
I
I
I
\
"
.~~
., ...
/
I
I
I
I
2.
.Insert the ends of the collector tube into theTOXTRAP
Holder as shown.
I
I
I i
II
.
,
:1
>!
r
}.:I
.
1
) i
'-I
"
"
~.I
,
-I,.
.I"'.
I?
I
(
.'
.: r }
3.
4.
\1
I,·
Remove the mouthpiece a.nd connect the unattached tube of the
TOXTRAP Holder to the breath tube.
With Recirculation without Recirculation
Depress the "START TEST" ~witch .
51