capturing Standards. CMI/MPH INTOXILYZER 5000
Advertisement
Advertisement
When the instrument requests you to detach the collector device, complete the following steps:
1.
Remove the TOXTRAP collector tube from the TOXTRAP Holder.
2.
3.
4.
Recap the TOXTRAP tube.
Disconnect the TOXTRAP Holder from the breath tube; the
TOXTRAP Holder may remain attached to the "PUMP
EXHAUST/SAMPLE CAPTURE" port on a standard instrument or to the "EXHAUST/SAMPLE CAPTURE" port on an instrument equipped with the Vapor Recirculation option.
Depress the "START TEST" switch.
NOTICE
~*~~**********~********.***~**********.********~*****************
TO PREVENT UNNECESSARY WEAR OF THE TOXTRAP HOLDER, DO NOT ALLOW A
COLLECTOR TUBE TO REMAIN IN THE TOXTRAP HOLDER FOR EXTENDED
PERIODS OF TIME.
CONNECT AND DISCONNECT THE COLLECTOR TUBE ONLY
WHEN THE DISPLAY REQUESTS YOU TO ATTACH AND DETACH THE COLLECTOR
DEVICE.
'
Captur~ng Standards
Activating S6 (Sample capture) does 'not program the instrument to request the attachment of a collector device following a calibration check.
Therefore, to capture a standard alcohol vapor from a wet bath simulator, you must fill the sample chamber with the standard alcohol vapor during the breath test mode .
The' following steps describe the procedure.
1.
Activate the sample capture option (S6 "on") and the ABA mode sequence ,(51 "off," S2 and S3 "on").
Complete the
"SETUP" procedures on page 10 and begin the mode sequence by pushing "START TEST."
2.
When the display reads "PLEASE BLOW/R INTO MOUTHPIECE UNTIL
TONE STOPS"; "PLEASE BLOW/R" (flashing), attach the outlet port of a wet bath simulator containing a standard alcohol solution to the breath tube.
(For additional information about the wet bath simulator, see "CONTROL CHECK," page 36.)
3.
Fill the sample chamber with the standard vapor by blowing into the simulator's inlet port.
4.
Detach the simulator from the breath tube and capture the' standard alcohol vapor as you would a breath sample.