"CMI INC INTOXILYZER-ALCOHOL ANALYZER FL MODEL 5000R--PUSH BUTTON
TO START ·TEST"; "PUSH BUTTON" (flashing); "TIME ##HR ##MIN" "DATE
MM/DD/YY"
Clock Errors
Consult a repair technician if:
1.
The time or· date or' both are erratic and "CLOCK ERROR" appears periodically on the display accompanied by a lowhigh tone sounding intermittently.
2.
The instrument fails.to update the time while turned "off."
Other Error Messages
Fcllowing are other error messages that may appear on the display.
The error messages are accompanied by a low-high tone sounding intermittently for approximately five seconds.
Displayed Error Message
Description and
Corrective operator Action
"UNSTABLE REF" The microprocessor was unable to obtain a stable reference signal from the processor.
The instrument halts the test, prints "Il'!VALID TEST"; "uNABLE
TO OBTAIN STABLE REFERENCE," and prepares itself to start another test.' If the instrument is' set in "DVM
TEST" when "UNSTABLE REF" appears on the display, the instrument returns
READY" followed to by riNdT the diagnostic checks.
When the display reads "CMI
INC INTOXILYZER-ALCOHOL
A..~ALYZER FL MODEL.5000R--PUSH
BUTTON TO etc., "begin another test by pushing the "START
TEST" button.
If "UNSTABLE
REF" again appears on the display, trip the computer
"RESET" switch located an the back of the instrument.
The
55
-"INVALID TEST" instrument will return to "NOT
READY" .and
subsequently perform the diagnostic checks.
See "Diagnostic Checkstratthe beginning of this section:
"MALFUNCTIONS AND DISPLAYED
ERROR . MESSAGES. " If the instrument completes the diagnostic findin~ a checks without malfunction, running another test.
try
If
"UNSTABLE REF" againappears on the display, turn the instrument "off" an.d consult a repair technician.
The "START TEST" button was pushed at the wrong time, the evidence card was pulled from the printer, or the instrument's pump inadequately purged the sample chamber.
The instrument cancels the test, prints "INVALID TEST"
(if the evidence card was not pulled from the printer) and prepares itself to begin another test.
When' the display reads "eMI
INC INTOXILYZER~ALCOHOL
ANALYZER FL HODEL 5000R--PUSH
BUTTON TO etc.," begin another test by pushing the "START
TEST" button.
,Hak.e
sure the
"START TEST" button is pushed only at the proper time and the evidence card remains in the card slot until the instrument releases it.
If the instrument again displays
"INVALID TEST" following the purge ("AIR BLANK") operation, turn.
the instrument OFF and consult a repair technician.
I
!
. ,
.•......
/
~I
I
I
'.
I
I
I
I
I
I
,
I
I
I
I
I
I
I
~ ...
'
:~ "INVALID SAMPLE"
"INHIBITED - RFI"
57
The instrument detected residual mouth . alcohol in the subject's breath sample.
The instrument completes the mode sequence, prints "INVALID
SAMPLE . XXX" in place of
"SUBJECT TEST
.iii"
and prints
"INVALID SAMPLE MOUTH
ALCOHOL" at end of the test sequence and prepares itself to begin another test.
Since normal body processes eliminate residual mouth alcohol within approximately
15 minutes, observe the subject for at least 15 minutes before beginning another breath-analysis.
During the observation time, the subject may not smoke, eat, drink, or introduce any substance into his mouth.
Furthermore, if the Subject regurgitates, note the time and delay beginning a breath analysis for at least 15 minutes.
High level radio. frequency interference is present.
The instrument halts the test, prints "INVALID TEST";
"INHIBITED RFI" and prepares to start another test.
If the instrument -isset in "DVM
TEST" when -radio frequency interference activates the RFI detector, returns to the' instrument
"NOT READY" followed by the diagnostic checks.
Locate the RFI source and either remove the source from the instrument's operational environment or instrument to move a environment free fromRFI.
the new
"LOW SAMPLE VOL"
"INVALID MODE"
"INTERFERENT"
The subject did not provide an adequate breath sample within approximately three minutes.
The inst~ument
"SUBJECT TEST.
highest BAC/BrAC displays
iii"
(the value obtainable trom the given breath samples) and completes the mode sequence.
On the evidence card, the instrument indicates the highe~t obtainable BAC/BrAC value by printing an asterisk before "SUBJECT TEST
(-)
.lli."
The asterisk (-) is a cross reference to the message printed at the bottom of the evidence card: ,,LOW SAMPLE
VOL VALUE
HIGHEST OBTAINED.
It
PRINTED WAS
When the display reads "eMI
INC INTOXILYZER-ALCOHOL
ANALYZER FL MODEL SOOOR--PUSH
BUTTON TO etc.," you may begin another breath analysis.
The Mode selection switches located on the side of the instrument are set improperly.
Check and reset the switches before beginning another test.
(See "switch ,priori ties And
Invalid Mode," page 27.)
The subject's breath sample or the standard alcohol vapor from an attached wet bath simulator contains a substance, such as acetone, that absorbs infrared energy in the same frequency range as alcohol.
The instrument compensates for the amount of infrared energy absorbed by acetone, completes the mode sequence, subject's followed
SUBTRACTED. " and by prints the
BAC/BrACvalue