Configuring Temperature/ Resistance test. SKF Baker AWA-IV 4 kV, Baker AWA-IV 12 kV, Baker AWA-IV 12 HO, Baker AWA-IV Series, Baker AWA-IV 2 kV
Add to My manuals214 Pages
advertisement
![Configuring Temperature/ Resistance test. SKF Baker AWA-IV 4 kV, Baker AWA-IV 12 kV, Baker AWA-IV 12 HO, Baker AWA-IV Series, Baker AWA-IV 2 kV | Manualzz Configuring Temperature/ Resistance test. SKF Baker AWA-IV 4 kV, Baker AWA-IV 12 kV, Baker AWA-IV 12 HO, Baker AWA-IV Series, Baker AWA-IV 2 kV | Manualzz](http://s2.manualzz.com/store/data/061559025_1-1bf1c4587ff90e1b99f9e6a936424ecf-360x466.png)
Test procedures
9) The new Test ID will be displayed and all tests will be turned off.
10) Turn on all required tests by clicking the ON/OFF buttons in the left-most column.
The buttons turn green when they are on.
Figure 87. New Test tab view with new Test ID created and selected. All tests turned on.
Configuring Temperature/ Resistance test
1) The temperature and resistance tests share the same setup window. Click on either the Temperature or Resistance test configuration buttons; by default, they are labeled Manual or 3 Lead/High V . The Temperature/ Resistance Test setup window will appear next.
2) Because the Temperature entry step has been selected, the Temperature Enable radio button is green.
3) Click on the Manual Temperature Entry radio button.
NOTE
The Baker AWA-IV will accept both ° C (Celsius) or ° F (Fahrenheit) temperatures and a temperature range of -32° C to 250° C. If you choose not to use
Temperature entry, click on the radio button and it will be disabled (grayed out).
PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual 85
Test procedures
4) Because the resistance test was chosen, the Resistance Enable radio button is also green. The motor in this example is wye-wound, which is indicated on the window and can be changed via the motor’s Nameplate tab.
Figure 88. Temperature/Resistance test parameters.
5) The 3 Leads and Automatic radio buttons will be selected by the software for the
3-phase motor. The analyzer will automatically acquire the resistance readings.
6) Depending on the model, some analyzers are equipped with a separate set of resistance test leads to perform a lead-to-lead low-voltage resistance test.
Resistance values must be greater than 0.500 ohms for high-voltage leads. If the high-voltage leads are used on a motor with resistances less than 0.500 ohms, the analyzer will prompt the operator to switch to low-voltage leads. For this example, the AWA model being used does not have the low-voltage leads, so there is no Res
Leads box to check.
7) Check the Delta R (%) box and ensure it is set to 10%. If the spread of resistance readings are more than 10%, the motor will fail.
8) Because temperature is enabled, the Correct to box is checked and defaults to 25° C and the value for copper is set to the IEEE 118 constant.
9) Target Corrected Resistance is a feature that refines the pass/fail criteria. If checked, the analyzer will fail a motor if the readings are not within tolerances. For example, the resistance reading for a motor can be taken using a DVM. Let’s say that the reading is 3.1 ohms. In this case, we can check the Target Corrected Resistance box and enter the value of 3.1 +/- 10%. One issue with using this feature is that enabling target corrected resistance makes this Test ID specific to a single motor. If resistance data is not available, or you want to be able to use the Test ID for other motors, do not enable this feature.
10) Click on the Close button.
86 PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual
advertisement
Related manuals
advertisement
Table of contents
- 21 Formatting
- 21 Information devices
- 23 Symbols on equipment
- 23 Labels on equipment
- 24 Safety precautions
- 24 Test related
- 26 Emergency stop button
- 27 Baker ZTX E-stop and remote E-stop
- 28 General operation, maintenance, and service information
- 28 Cleaning and decontamination
- 28 Technical assistance / authorized service centers
- 28 Unpacking the unit
- 28 Pollution degree II
- 28 Power requirements
- 29 Environmental conditions
- 29 Power pack lifting and shipping
- 29 Lifting the instrument
- 30 Operating and shipping positions
- 31 Database management
- 31 Consequences of not organizing data
- 32 Starting the software
- 32 Creating a new database
- 33 Opening an existing database
- 34 Using multiple databases
- 35 Data Transfer feature
- 35 Transferring motor and test results data
- 38 Transferring Test IDs
- 39 Archiving a database
- 40 Restoring a database
- 43 Baker AWA-IV 2 kV/4 kV model front panel
- 43 USB ports
- 43 Ethernet connector
- 44 Emergency power shut-off
- 44 Resistance leads
- 44 High-voltage test leads
- 44 Voltage output control knob (6 kV and 12 kV models only)
- 44 Baker AWA-IV 6 kV/12 kV model front panel
- 45 Baker AWA-IV 6 kV model distinctions
- 45 Setting up the Baker AWA-IV tester
- 45 Selecting an optimal environment
- 45 Making basic connections and starting the analyzer
- 46 Connecting test leads to motor under test
- 46 Configuring a printer
- 46 Using the footswitch
- 47 Starting the software
- 48 Creating a new database
- 49 Opening an existing database
- 50 Using version 4 software for the first time
- 51 Main window
- 52 Main menu
- 52 File menu
- 53 View menu
- 54 Database menu
- 54 Window menu
- 55 Tools menu
- 55 Help menu
- 56 Toolbar
- 57 Explore tab
- 58 Motor ID tab
- 59 Route tab
- 61 Modifying the display of lists in the Motor ID and Route tabs
- 62 Viewing test data
- 63 Using the Data tab
- 64 Motor location fields
- 64 Motor ID field
- 65 Data tab—Nameplate view
- 67 Data tab—Application view
- 69 Data tab—Results Summary view
- 70 Data tab—Surge view
- 72 Data tab—PI view
- 73 Data tab—Step/Ramp-Voltage view
- 75 Using the Tests tab
- 76 Test configuration
- 76 Temperature/Resistance test setup window
- 79 Manually entering resistance measurements
- 80 DC Tests setup window
- 83 Surge test setup window
- 85 E bar graph
- 86 Creating a Surge test reference
- 90 Viewing Surge test results
- 91 Using the Trending tab
- 92 Max Delta R%
- 92 Resistance Trending Graphs
- 93 Insulation Resistance/MegOhm
- 94 HiPot
- 95 Relative humidity
- 95 Special software trending features
- 97 Before testing begins
- 97 Recommended testing sequence
- 98 Balance resistance test or line-to-line resistance
- 98 MegOhm test
- 98 DA/PI test
- 99 HiPot test
- 99 Step Voltage test
- 99 Surge test
- 100 Recommended test voltages for insulation resistance testing
- 100 Recommended test voltages for HiPot and Surge tests
- 101 Performing an example test
- 101 Creating a Motor ID
- 103 Creating a Test ID
- 105 Configuring Temperature/ Resistance test
- 107 Configuring DC tests
- 109 Configure Surge test
- 111 Running an automatic test
- 115 Reviewing test results/data
- 118 Printing reports
- 123 Creating a new motor voltage class
- 124 Surge testing notes and recommendations
- 124 Surge testing with rotor removed (typically motor shop testing)
- 124 Surge testing with rotor installed (typically field testing)
- 124 Surge testing DC motors
- 124 False P-P EAR failures
- 124 Surge test underpowered
- 125 Surge testing through capacitors
- 127 Predictive maintenance
- 128 Quality control
- 128 Motor troubleshooting
- 128 Field coils
- 129 Hi L in Baker AWA-IV 2 kV and Baker AWA-IV 4 kV
- 130 Using the Hi L technique
- 131 Fine tuning the technique
- 138 Power pack setup
- 138 Operating position
- 139 Combining Baker AWA-IV host and power pack tests
- 139 Creating IDs and setting up the test
- 142 Running the combined Baker AWA-IV and power pack tests
- 144 Testing with the Baker PP30 three-phase test lead power pack
- 145 Conducting DC tests with the Baker PP30 three-phase test lead power pack
- 147 Conducting Surge tests with the Baker PP30 three-phase test lead power pack
- 150 Testing with the Baker PP24 single-phase test lead power pack
- 150 Conducting DC tests with the Baker PP24 single-phase test lead power pack
- 151 Conducting Surge tests with the Baker PP24 single-phase test lead power pack
- 153 Principles of armature insulation testing
- 154 Connecting Baker AWA-IV to the Baker ZTX accessory
- 155 Armature preparation
- 156 Configuring a Surge test for armature bar-to-bar testing
- 162 Reviewing test results/data
- 164 Printing reports
- 165 Generating CSV files
- 167 Self-help and diagnostics
- 167 Repair parts
- 167 Step #1: Basic information
- 168 Step #2: Applications or service problem?
- 168 Applications: What to do first
- 168 Common application problems
- 170 Service: What to do first
- 170 Open condition display
- 170 HiPot display checks
- 171 Open ground check
- 171 Answer these questions
- 171 Limited output surge waveform
- 172 Proper storage of leads/unit
- 172 Checking test leads for broken sections
- 172 Manual break check
- 172 Overcurrent trip test
- 172 Open circuit test to verify analyzer operation
- 173 Third-party software warning
- 174 Warranty return
- 174 Warranty return form
- 175 Calibration information
- 175 Baker AWA-IV 2 kV and 4 kV tester specifications
- 177 Baker AWA-IV 6 kV, 12 kV, and 12 kVHO tester specifications
- 179 Applicable standards
- 181 Version 4.0 database definition
- 181 Nameplate table—(MotorID)
- 182 Test results table—(TestResults)
- 186 Memo table—(Memo)
- 187 Polarization Index Test Results table—(TestResultsPI)
- 189 Step Voltage test results table—(TestResultsPrgHiPot)
- 190 Surge test results table—(SurgeWaveform)
- 191 Test results parameters table—(TestResultsParameters)
- 195 Test ID table—(TestId)
- 199 Step Voltage test ID table—(TestIdPrgHiPot)
- 200 Reference Surge waveform table— (RefSrgWaveForm)
- 201 Database Information table—(DatabaseInfo)
- 201 Work list table—(Route)
- 201 Motor voltage class table—(MotorVoltageClasses)
- 203 Software fault messages
- 203 Resistance failure types
- 204 DC test failure types
- 204 Surge test failure types
- 205 Fault analysis chart
- 205 AWA-IV static testing parameters, indicators, and common causes