Reviewing test results/data. SKF Baker AWA-IV 4 kV, Baker AWA-IV 12 kV, Baker AWA-IV 12 HO, Baker AWA-IV Series, Baker AWA-IV 2 kV

Add to My manuals
214 Pages

advertisement

Reviewing test results/data. SKF Baker AWA-IV 4 kV, Baker AWA-IV 12 kV, Baker AWA-IV 12 HO, Baker AWA-IV Series, Baker AWA-IV 2 kV | Manualzz

Test procedures

Reviewing test results/data

After the test results have been saved to the database, they can be reviewed using the Data tab in the right pane of the software’s Main window. The Results Summary tab has a Date/

Time area on the top part of the window and a spreadsheet style view of the data on the bottom.

Figure 98. Main window—Data tab; Results Summary view.

Figure 99. Date/time results summary section close-up.

The Date/Time area shows a quick summary of the time and date of tests, and whether the tests passed or failed. Double click on a test date and time to move to a new record.

You can also right click on a record to open a Delete dialog box if you choose to remove a specific test record.

PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual 95

Test procedures

If you want to expand a column to view more information, hover the mouse cursor over a column’s edge until the cursor changes to a two-headed arrow, then click and hold the left button down while dragging the edge to create the desired column width.

Figure 100. Date/time results summary section close-up.

The Results Summary view presents the test data acquired in a spreadsheet style. The test date and time are shown across the top of the window with specific measurement results shown in each column.

Figure 101. Test results spreadsheet; lower section.

96 PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual

Test procedures

The PI test can be reviewed by pressing the PI tab. PI view will display the PI/DA graph along with a table of the current and MegOhm readings gathered at specific times. The PI voltage,

DA ratio, and PI ratio are displayed on the right side. Because this test was a DA, only the PI ratio has I=0 No PI indicating there is no PI value because the current was zero.

Figure 102. Reviewing PI test data.

Click on the Surge tab to view surge test data.

Figure 103. Surge waveform view.

PUB CM/I4 71-015 EN V13.2 Static Motor Analyzer—Baker AWA-IV User Manual 97