6. Measurement. KYORITSU 2033
KYORITSU 2033 is a digital AC/DC clamp meter designed for safe and accurate current measurement. With its auto-ranging feature and a wide measuring range of up to 300A, it simplifies the process of measuring AC and DC currents without the need for manual range selection. Its compact size and tear-drop-shaped jaws make it ideal for use in tight spaces and crowded cable areas.
إعلان
إعلان
6. Measurement
6-1 AC Current Measurement
# DANGER
● Do not make measurement on a circuit above 300V AC. This may cause shock hazard.
● Do not make measurement with the battery compartment cover removed from the instrument.
● Keep your fingers and hands behind the barrier during measurement.
(1) Set the Function Selector switch to the " A" position. "AC" should be shown on the lower left corner of the display.
(2) Press the jaw trigger to open the transformer jaws and clamp them onto the conductor under test, then take the reading on the display.
Position the conductor at the center of the transformer jaws for accurate measurement.
Note :
◇ During current measurement, keep the transformer jaws fully closed.
Otherwise,accurate measurement cannot be made. The maximum measurable conductor size is approx. 24mm in diameter.
◇ Unlike in DC current measurement, zero adjustment is not necessary in AC current measurement. There is no polarity in the reading either.
Correct
— 10 —
Wrong
6-2 DC Current Measurement
# DANGER
● Do not make measurement on a circuit above 300V AC. This may cause shock hazard.
● Do not make measurement with the battery compartment cover removed from the instrument.
● Keep your fingers and hands behind the barrier during measurement.
(1) Set the Function Selector switch to the " A" position DC should be shown on the upper left corner of the display.
(2) With the transformer jaws closed and without clamping them onto the conductor, presses the Zero ADJ. Button for about one second to zero adjust the display.
(3) Press the jaw trigger to open the transformer jaws and clamp them onto the conductor under test and take the reading on the display.
Position the conductor at the center of the transformer jaws for accurate measurement.
Note
◇ During current measurement, keep the transformer jaws fully closed.
Otherwise, accurate measurement cannot be made. The maximum measurable conductor size is approx. 24mm in diameter.
◇ When the current flows from the upside (the display side)to the underside of the instrument, the polarity of the reading is positive and vice versa. (See the figure at the left.)
Correct
“+” Current
Wrong
— 11 —
“-” Current
تنزيل