![](http://s1.manualzz.com/store/data/006657317_1-1969b957a8715efeabd5951efa20f988-128x128.png)
advertisement
![Xcalibur II manual | Manualzz Xcalibur II manual | Manualzz](http://s1.manualzz.com/store/data/006657317_1-1969b957a8715efeabd5951efa20f988-360x466.png)
MAINTENANCE INSTRUCTIONS
5. Adjust the x and y tilt screws and type card raw on 0.5. Repeat point 5 until an image results where the maximum intensity resides on the right hand spot. This is typically 5.5 turns of each of the x and y tilt screws from point 1 above (screws are 1/4 mm pitch, 20 turns range).
6. Position the mouse cursor over the right hand spot. Right mouse click and select integration
at pixel position. Record the intensity figure from the history window.
7. Fine adjust x tilt screw [i.e. screw 5(1) in Figure 9.10.1]. Type card raw on 0.5. Repeat points
6 – 7 until a maximum intensity figure results.
8. Fine adjust y tilt screw [i.e. screw 5(2) in Figure 9.10.1]. Type card raw on 0.5. Record intensity of the right hand spot as in point 6 above. Repeat point 8 until a maximum intensity figure results.
9. The optics are now aligned.
9.10.3 Collimator Alignment of Enhance Ultra
1. Remount the collimator by pushing on and tighten the grub screws 2 (see Figure 9.10.1). Note the collimator has a specific orientation.
2. We now need to align the collimator. Dismount the 0.6 mm collimator cap by sliding it off.
4. Adjust the collimator x and y adjustment screws 7 and 8 (see Figure 9.10.1). Repeat 3 -4 until the X-ray beam is visible as a single intense circular spot.
5. Mount the 1mm collimator cap.
7. Fine adjust the collimator x and y adjustment screws 7 and 8 (see Figure 9.10.1). Repeat 6 -7 until the X-ray beam is visible as a single intense circular spot.
8. Finally, dismount the 1mm collimator cap and re-mount the 0.6 mm cap.
10. Adjust the collimator x and y adjustment screws 7 and 8. Repeat 9 - 10 until the X-ray beam is visible as a single intense circular spot.
11. The X-ray beam and focusing optics are now fully aligned. We now need to point the focused
X-ray beam through the centre of the diffractometer’s sphere of confusion. The sample position.
9.10.4 Aligning the beam to the centre of the goniometer –
Enhance Ultra
1. Set X-ray generator settings to 12 kV and 1mA
2. Flip beamstop to the “up” position (see Figure 9.10.2)
3. Mount the alignment disc with a 0.3 mm pinhole on a goniometer head and place on the goniometer.
4. Optically align the alignment disc at the centre of the goniometer, as with a crystal sample, using F12 etc.
5. Move the detector to 60 mm using km4 detector 60.
Version 1.4 Xcalibur_Manual_v1.4
Page 127
Xcalibur
USER MANUAL
advertisement
Related manuals
advertisement
Table of contents
- 10 1. Health and Safety Information
- 10 1.1 General
- 11 1.2 Electrical Safety
- 11 1.2.1 Potential Electrical Hazards
- 11 1.2.2 Recommended Precautions
- 11 1.2.3 First Aid
- 12 1.3 Mechanical Handling Safety
- 12 1.4 Safe Mechanical Practice
- 12 1.5 Moving Parts
- 12 1.6 X-ray Radiation
- 13 1.7 Extreme Temperatures
- 14 1.8 Vacuum
- 14 1.9 High Pressures
- 14 1.10 Hazardous or Toxic Materials
- 14 1.11 Modifications and Service
- 15 2. Introduction
- 15 2.1 Scope
- 15 2.2 How To Use This Manual
- 15 2.3 System Overview
- 16 3. Specifications
- 16 3.1 Environmental Requirements
- 16 3.2 Services
- 16 3.2.1 Electrical Supply
- 16 3.2.2 Water Cooling
- 17 3.2.3 Helium Gas Supply (where applicable)
- 17 3.3 Performance Data
- 17 3.3.1 X-ray Tube (Typical Operating Conditions)
- 18 3.3.2 Sapphire 2 CCD Detector
- 19 3.3.3 Sapphire 3 CCD Detector
- 20 3.3.4 Onyx CCD Detector
- 20 Onyx CCD Detector Theta and Resolution Ranges
- 21 3.3.5 PC CCD Interface
- 21 3.3.6 Four-circle Kappa Geometry X-ray Goniometer
- 22 4.4 Electrical Data
- 23 4. Technical Description
- 23 4.1 Overview
- 24 4.2 CCD Detector Technology
- 25 4.2.1 Beryllium Window
- 26 4.2.2 Phosphor
- 26 4.2.3 Taper
- 26 4.2.4 CCD
- 26 4.2.5 Readout Speed
- 26 4.2.6 Binning
- 27 4.2.7 Dark Current and MPP Mode
- 27 4.2.8 Radiation Damage
- 27 4.2.9 Full Well Depth and 18-bit Digitisation
- 27 4.2.10 Anti-blooming
- 27 4.2.11 Vacuum
- 27 4.2.12 Fast Shutter
- 28 4.2.13 Zingers and Cosmic Ray Events
- 28 4.3 Four-Circle Kappa Geometry Goniometer
- 29 4.4 X-ray Generator
- 30 4.5 Software
- 30 4.5.1 Directory Structure
- 31 4.5.2 Basic Menu Philosophy
- 31 4.6 KMW200CCD Chiller
- 31 4.7 KMW3000C Chiller
- 31 4.8 Low Temperature Option
- 31 4.9 Safety Features
- 32 5. Handling, Installation, Storage and Transit Information
- 32 5.1 Reception and Handling
- 32 5.1.1 Delivery
- 32 5.1.2 Unpacking
- 33 5.1.3 Mechanical Handling
- 33 5.1.3.1 Weights, Dimensions and Lifting Points
- 34 Delivery
- 34 5.2 Installation and Setting to Work
- 34 5.2.1 Preparation of Site and Services
- 34 5.2.1.1 Environmental Requirements
- 34 5.2.1.2 System Layout
- 35 5.2.1.3 Electrical Services
- 36 5.2.1.4 Water Supply
- 36 5.2.1.5 Low Temperature Option
- 36 5.2.1.6 CCD Camera Pumping
- 36 5.2.1.7 Helijet Option
- 37 5.2.2 Setting to Work
- 37 5.2.2.1 Equipment Required
- 37 5.2.2.2 Personnel Required for Installation
- 37 5.2.2.3 Setting up Procedures
- 39 Storage
- 40 6. Operation
- 40 6.1 Controls and Indicators
- 41 6.2 Initial Switch on Procedure
- 42 6.3 X-ray Tube Warm-up Procedure
- 43 6.4 Software
- 43 6.4.1 Software Updates
- 44 6.4.2 Software Installation
- 44 6.4.2.1 MGC interface software
- 45 6.4.2.2 CrysAlis Software
- 45 6.4.3 Changing Machine Correction and Set-up Files
- 46 6.5 Normal Operation
- 46 6.5.1 General Commands
- 48 6.5.2 Changing Xcalibur Settings
- 52 6.5.3 Standard Diffraction Experiment
- 52 6.5.3.1 Crystal Mounting and Alignment
- 54 6.5.3.2 Diffraction Photographs to Determine Crystal Quality
- 55 6.5.3.3 Unit Cell Determination
- 57 6.5.3.4 Data Collection
- 58 6.5.3.5 Data Processing and Reduction
- 59 6.5.3.5.1 Orientation Matrix
- 60 6.5.3.5.2 Run List
- 61 6.5.3.5.3 Scan Width
- 61 6.5.3.5.4 Background Evaluation
- 62 6.5.3.5.5 Special Corrections
- 63 6.5.3.5.6 Outlier Rejection
- 64 6.5.3.5.7 Output Format
- 65 6.5.3.6 Changing the Output Format from Data Reduction
- 66 6.5.3.7 Absorption Correction
- 74 6.5.3.8 GRAL - Space Group Determination
- 80 6.5.3.9 Structure Solution and Refinement
- 80 6.5.4 Ewald explorer
- 86 6.5.5 Dc Movie - Replay of Data Collection Movie
- 87 6.5.6 Reconstruction of Precession Photographs
- 91 6.5.7 Dc opti - Optimisation of Data Collection Strategy
- 96 6.5.8 Indexing and Data Reduction of Incommensurate Samples
- 97 6.5.9 Indexing and Data Reduction of Twinned Samples
- 99 6.5.10 Extracting Data from Powder Samples
- 100 6.5.11 Refining of Machine Parameter File
- 103 6.5.12 Glossary of CrysAlis Commands
- 110 6.6 Normal Shutdown
- 110 6.7 Emergency Shutdown
- 111 6.7.1 Emergency Shutdown Procedure
- 112 7. Mechanical Changeover of Detectors and X-ray Sources
- 112 7.1 Interchange of CCD Detectors
- 112 7.1.1 Installation of a Sapphire 2 and Sapphire 3 CCD detectors
- 113 7.1.2 Removal of a Sapphire 2 and Sapphire 3 CCD detectors
- 113 7.1.3 Installation of the Onyx CCD camera
- 116 7.1.4 Removal of the Onyx CCD camera
- 117 7.2 Procedure for Interchange of the Molybednum and Copper Enhance X-ray Source
- 121 8. Maintenance Schedules
- 121 8.1 Introduction
- 121 8.2 Weekly Maintenance Schedule
- 121 8.3 Monthly Maintenance Schedule
- 122 8.4 Six Monthly Maintenance Schedule
- 122 8.5 Yearly Maintenance Schedule
- 123 8.6 10,000 Hours Maintenance Schedule
- 124 9. Maintenance Instructions
- 124 9.1 Special Tools
- 124 9.2 Refining the Machine Parameter File
- 125 9.3 Changing the X-ray Tube of Enhance
- 126 9.4 Changing the X-ray Tube of Enhance ULTRA
- 128 9.5 X-ray Beam Stop Alignment
- 129 9.6 Changing the Collimator of Enhance
- 130 9.7 Changing the Collimator of Enhance Ultra
- 130 9.8 Aligning the X-ray Collimator of Enhance
- 130 9.9 Aligning the Enhance X-ray Source
- 133 9.10 Aligning the Enhance Ultra X-ray Source
- 133 9.10.1 X-ray Beam Alignment of Enhance Ultra
- 135 9.10.2 Optic Alignment of Enhance Ultra
- 136 9.10.3 Collimator Alignment of Enhance Ultra
- 136 9.10.4 Aligning the beam to the centre of the goniometer – Enhance Ultra
- 137 9.11 Checking the Door Safety Interlocks
- 137 9.12 Checking the Emergency stop
- 138 9.13 Checking the X-ray Radiation Levels
- 138 9.14 CCD Detector – Pumping Out Vacuum
- 140 9.15 Dismantling Xcalibur
- 143 10. Trouble Shooting
- 146 11. Spares
- 148 12. Disposal Instructions
- 148 12.1 X-ray Tube and CCD Detector
- 148 12.2 Third Party Equipment
- 149 13. Additional Information
- 149 13.1 Third Party Information
- 150 13.2 Drawings
- 150 13.2.1 Mechanical Drawings
- 150 Xcalibur Suggested Layout
- 150 System and Component Dimensions
- 152 13.2.2 Electrical Drawings
- 154 14. CE Conformity notice
- 155 Appendices
- 155 Appendix 1 X-ray Tubes Wave Lengths
- 155 Appendix 2 Standard Crystal Parameters
- 155 Appendix 3 Temperature Scales Conversion
- 155 Appendix 4 Maintenance Records
- 160 Appendix 5 Example of Local Rules for the Xcalibur System Set Up at Oxford Diffraction