HP 8590D Spectrum Analyzer Calibration Guide

Add to My manuals
135 Pages

advertisement

HP 8590D Spectrum Analyzer Calibration Guide | Manualzz

Option Specifications

Tracking Generator Specifications (Option 010 or 011)

All specifications apply over 0 ‘C to + 55 “C. The spectrum-analyzer/tracking-generator combination will meet its specifications after 2 hours of storage at a constant temperature within the operating temperature range, 30 minutes after the spectrum-analyzer/trackinggenerator is turned on and after CAL FREQ, CAL AMPTD, CAL TRK GEN, and

TRACKING PEAK have been run.

Warm-Up

30 minutes

Output Power Level

Range

50 0 (Option 010)

75 Q (Option 011)

Resolution

Absolute Accuracy

Vernier

Range

Accuracy

Oto-15dBm

+ 42.8 to + 27.8 dBmV

0.1 dB f 1.5 dB (at 300 MHz and - 10 dBm source power)

(Option 011: use i38.8 dBmV instead

of

-10 dBm)

15 dB f 1 .O dB (referenced to - 10 dBm source power)

(Option 011: referenced to

t38.8

dBmV instead of -10 dBm)

Output Power Sweep

Range

50 Q (0ption 010)

75 Q (Option

Resolution

011)

Accuracy (zero span)

- 15 dBm to 0 dBm

+27.8 to t 42.8 dBmV

0.1 dB

~2 dB peak-to-peak

Output Flatness

(referenced to 300 MHz) ±1.75 dB

Spurious Outputs

50 Q (Option 010)

75 Q (Option 011)

Harmonic Spurs

Nonharmonic Spurs

(0 dBm output, 100 kHz to 1.8 GHz)

(t

42.8 dBmV output, 1 MHz to 1.8 GHz)

<-25 dBc

<-30 dBc

2-8 Specifications and Characteristics HP 8590D Spectrum Analyzer

Dynamic Range

Tracking Generator Feedthrough

50 n (Option

010)

75

Q (Option

011)

<- 106 dBm

<-57.24 dBmV

Option Specifications

HP 8580D Spectrum Analyzer

Specifications and Characteristics 2-8