HP 8590D Spectrum Analyzer Calibration Guide

Add to My manuals
135 Pages

advertisement

HP 8590D Spectrum Analyzer Calibration Guide | Manualzz

Frequency Characteristics

These are not specifications. Characteristics provide useful but nonwarranted information about instrument performance.

Stability

Drift

Drift* (Option 013)

175 kHz/5 minutes after 2 hour warmup and 5 minutes after setting center frequency

Frequency Span 5 10 MHz, Free Run ~2 kHz/minute of sweep time

’ (Option 013)

Because the analyzer is locked at the center frequency before each sweep, drift occurs only during the

:ime of one sweep. For Line, Video or External trigger, additional drift occurs while waiting for the appropriate rigger signal.

Resolution Handwidth (-3 dB)

Range

Shape

1 kHz to 3 MHz, selectable in 1, 3 and 10 increments, accuracy f20% and 5 MHz. Resolution bandwidths may be selected manually, or coupled to frequency span.

Synchronously tuned four poles. Approximately Gaussian shape.

60 dBl3 dB Bandwidth Ratio

Resolution Bandwidth

100 kHz to 3 MHz

30 kHz

3 kHz to 10 kHz

1 kHz

15:l

16:l

15:l

16:l

Video Bandwidth (-3 dB)

Range 30 Hz to 1 MHz, selectable in 1, 3, 10 increments, accuracy f30% and 3 MHz. Video bandwidths may be selected manually, or coupled to resolution bandwidth and frequency span.

Shape Post detection, single pole low-pass filter used to average displayed noise.

FFT Bandwidth Factors

Noise Equivalent Bandwidtht

3 dB Bandwidtht

Sidelobe Height

Amplitude Uncertainty

Shape Factor (60 dB BW/3 dB BW) t Multiply entry by one-divided-by-sweep time.

FLATTOP HANNING

3.63x 1.5x

3.60x 1.48x

<-90 dB -32 dB

0.10 dB 1.42 dB

2.6 9.1

UNIFORM

l x l x

-13 dB

3.92 dB

>300

2-10 Specifications end Characteristics HP 8590D Spectrum Analyzer