Checking Conformance Manually. Dome CXtra

Add to My manuals
126 Pages

advertisement

Checking Conformance Manually. Dome CXtra | Manualzz

To set manual alert parameters

1 Select either the DICOM LUM test or the White Level test or both.

• For the DICOM LUM test, check the selection box. Then enter a value for Maximum allowed. The DICOM LUM test is classified a failure if it produces a value greater than the maximum allowed.

Enter a value ranging from 0 (zero) to 2 (default=1.00). A value of

0 (zero) indicates perfect conformance with the DICOM Grayscale

Standard Display Function.

• For the CX/EX White Level test, check the selection box. Then enter a value for Tolerance ±. The CX/EX White Level test is classified a failure if the measured white level is outside the range defined by this percent, applied to the Target Luminance defined on the Backlight tab. Enter a value ranging from 0 (zero) to 25

(default =10.00).

2 Select the number of luminance measurements you want to make during each manual test. The default number of measurements is 15.

NOTE: When you select All, 256 measurements are taken on 8-bit display systems; 1024 measurements are taken on 10-bit systems.

Checking Conformance Manually

If you have configured RightLight properly and have an external photometer installed, you can run a manual conformance test.

If you have the Privilege service installed, you must log on as a privileged user before you run a manual conformance test. (See

page 96.)

As with automatic conformance testing, the manual conformance test shows how well your display conforms to the DICOM standard.

You are able to specify, however, the number of measurements to be taken. See “Manual Alerts tab” on page 22 for configuration

information. See “Installing the Photometer” on page 8 for how to

install the photometer.

RightLight Service | 23

24 | Dome CXtra

Before starting a manual conformance test, select Dome CX/EX

DICOM Calibration on the services menu.

NOTE: If you run a conformance test with another response function

(such as Dome CX/EX Exponential Calibration or Dome CX/EX Native

Calibration), the system presents a warning message. Continuing to run the test causes the system to show the Dome CXtra icon in the warning state.

To run a manual conformance test

1 Open the Dome CXtra Services menu. Select RightLight

Conformance > Manual Conformance Wizard.

2 Follow the Conformance Wizard instructions.

After you finish with the Conformance Wizard, RightLight enters

the event in both the Reporting History log and the Conformance

History log. All RightLight history events from manual conformance tests are recorded in the Reporting History log. RightLight provides additional information on conformance events in graphical and tabular formats in the Conformance History log.

If the manual conformance test produces a warning condition because the white level is beyond the tolerance set on the

Manual Alerts tab (page 22), then you can correct the warning by calibrating the white level via the conformance test. Select the check box to correct the warning, and click Next. The wizard executes automatically.

Alternatively, you can adjust the tolerance for the failing test to

bring the display into conformance. (See also “White Level

Calibration Service” on page 35.)

NOTE: White Level calibration during conformance testing is available only for Dome E2, Dome E3, Dome E3c, Dome E3cHB, and

Dome E5 displays. You must be an administrator or a privileged user to perform a manual conformance test and to calibrate the white level during the manual conformance test.

RightLight Service | 25

advertisement

Related manuals

advertisement

Table of contents