Performing a Custom DICOM Calibration. Dome CXtra

Add to My manuals
126 Pages

advertisement

Performing a Custom DICOM Calibration. Dome CXtra | Manualzz

44 | Dome CXtra

Performing a Custom DICOM Calibration

A custom DICOM calibration uses the actual luminance values of your display to adjust the response curve. After you complete the calibration, the new values are applied in place of the default

DICOM calibration whenever DICOM Calibration is selected on the services menu.

NOTE: This procedure requires installation of the optional LCD or

LCD II photometer as described on

page 8.

To perform a custom calibration

1 If you have one of the display models listed below, follow instructions provided with the display to verify that the contrast is set to the value shown.

Display Model

GX2MP

GX2MP (JUN 2007 or later)

PX191

PX1910M (2007 or later)

PX1910M

PX212M

PX212M (RoHS)

Contrast

70%

70%

50%

60%

50%

80%

50%

2 If the Privilege service has been installed, log on as a privileged

user. (See page 96.)

3 Open the Dome CXtra Services menu. Select DICal Calibration >

Calibration Wizard.

4 Follow the instructions for using the photometer in each

Calibration Wizard window and click Next.

NOTE: Abort the calibration at any time by pressing Ctrl+C.

5 If necessary, verify that your display calibration conforms to

the DICOM standard. (See page 46 for instructions.)

NOTE: If you change panel brightness using the brightness control on the display, you must repeat the preceding calibration steps.