Performing a Custom DICOM Calibration
A custom DICOM calibration uses the actual luminance values of your display to adjust the response curve. After you complete the calibration, the new values are applied in place of the default
DICOM calibration whenever DICOM Calibration is selected on the services menu.
NOTE: This procedure requires installation of the optional LCD or
LCD II photometer as described on
To perform a custom calibration
1 If you have one of the display models listed below, follow instructions provided with the display to verify that the contrast is set to the value shown.
Display Model
GX2MP
GX2MP (JUN 2007 or later)
PX191
PX1910M (2007 or later)
PX1910M
PX212M
PX212M (RoHS)
Contrast
70%
70%
50%
60%
50%
80%
50%
2 If the Privilege service has been installed, log on as a privileged
3 Open the Dome CXtra Services menu. Select DICal Calibration >
Calibration Wizard.
4 Follow the instructions for using the photometer in each
Calibration Wizard window and click Next.
NOTE: Abort the calibration at any time by pressing Ctrl+C.
5 If necessary, verify that your display calibration conforms to
the DICOM standard. (See page 46 for instructions.)
NOTE: If you change panel brightness using the brightness control on the display, you must repeat the preceding calibration steps.
advertisement
Related manuals
advertisement
Table of contents
- 7 Display Compatibility
- 8 Dome CXtra Services
- 10 System Requirements
- 10 Installing the Dome CXtra Software
- 11 Installing or Removing Services
- 12 Installing the Photometer
- 14 Starting Dome CXtra
- 15 Opening the Dome CXtra Services Menu
- 16 CXtra Utilities
- 16 Exiting Dome CXtra
- 16 Uninstalling Dome CXtra
- 18 Configuring the RightLight Service
- 27 Checking Conformance Manually
- 30 Checking Current Conformance Status
- 31 Reviewing Conformance Events
- 32 Reviewing Manual Conformance Data
- 37 Disabling RightLight Popup Warnings
- 44 Configuring the DICal Service
- 45 Using DICOM Response Function as Default Calibration
- 46 Choosing Display Calibration
- 48 Performing a Custom DICOM Calibration
- 49 Reviewing Calibration Events
- 50 Performing a Conformance Test
- 50 Reviewing Conformance Test Data
- 54 Reviewing Conformance Test Events
- 56 Advanced Options
- 58 AAPM TG18 Test Patterns
- 58 Miscellaneous Test Patterns
- 61 Displaying Test Patterns
- 62 Adding a Custom Test Pattern
- 64 Modifying or Deleting Custom Test Patterns
- 64 Dome Visual and Interactive Evaluations
- 70 Creating a Saver Period
- 72 Deleting a Saver Period
- 72 Viewing Schedule for Week
- 74 Setting Up for Testing
- 84 Running a DIN Acceptance Test
- 85 Running a DIN Constancy Test
- 87 Viewing DIN Test Results
- 90 AAPM TG18 Test Setup
- 96 Completing AAPM TG18 Tests
- 98 Viewing AAPM TG18 Test Results
- 100 Logging On as Privileged User/Administrator
- 101 Logging Out of Privileged Mode
- 101 Changing Your Password
- 102 Viewing Reports
- 106 Using the Reporting Control
- 109 SNMP Management Console
- 109 The SNMP Model
- 110 Configuring the SNMP Service