Deleting a Saver Period
To delete an existing saver period
1 Open the Dome CXtra Services menu and select Backlight Saver.
2 Click the associated data line and then click Delete Period.
Viewing Schedule for Week
To view the saver periods for a week
1 Open the Dome CXtra Services menu and select Backlight Saver.
2 Click View Week to see the schedule.
3 Click OK.
DIN Test Service
The DIN Test service tests your Dome CX, Dome EX, and Dome QX displays for conformance with German standard DIN 6868-57. A successful acceptance test such as the one performed by DIN Test service is mandatory before DIN 6868-57 application category A and
B devices
*
can be accepted in a hospital or other diagnostic setting.
For more information about DIN acceptance criteria, obtain a copy of the DIN 6868-57 standard through the following web site: http://www2.din.de.
Besides acceptance testing, the DIN Test service provides constancy testing. Constancy tests can be performed periodically after successful acceptance testing to detect any degradation in display performance. The tests detect degradation by measuring deviation from performance values recorded during the acceptance test.
* Application category A includes image display devices for the diagnosis of digital images taken by means of projection radiography, including storage phosphors or solid-state detectors, for example, of the skeleton, thorax or mamma. However, image display devices for the diagnosis of digital images from dental radiography, fluoroscopy, or subtraction angiography are not included in application category A.
Application category B includes all image display devices suitable for the diagnosis of digital images from dental radiography, fluoroscopy, and subtraction angiography as well as of images generated by means of sectional imaging procedures (such as computed tomography, magnetic resonance tomography, sonography) and nuclear-medical procedures.
DIN Test Service | 69
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 7 Display Compatibility
- 8 Dome CXtra Services
- 10 System Requirements
- 10 Installing the Dome CXtra Software
- 11 Installing or Removing Services
- 12 Installing the Photometer
- 14 Starting Dome CXtra
- 15 Opening the Dome CXtra Services Menu
- 16 CXtra Utilities
- 16 Exiting Dome CXtra
- 16 Uninstalling Dome CXtra
- 18 Configuring the RightLight Service
- 27 Checking Conformance Manually
- 30 Checking Current Conformance Status
- 31 Reviewing Conformance Events
- 32 Reviewing Manual Conformance Data
- 37 Disabling RightLight Popup Warnings
- 44 Configuring the DICal Service
- 45 Using DICOM Response Function as Default Calibration
- 46 Choosing Display Calibration
- 48 Performing a Custom DICOM Calibration
- 49 Reviewing Calibration Events
- 50 Performing a Conformance Test
- 50 Reviewing Conformance Test Data
- 54 Reviewing Conformance Test Events
- 56 Advanced Options
- 58 AAPM TG18 Test Patterns
- 58 Miscellaneous Test Patterns
- 61 Displaying Test Patterns
- 62 Adding a Custom Test Pattern
- 64 Modifying or Deleting Custom Test Patterns
- 64 Dome Visual and Interactive Evaluations
- 70 Creating a Saver Period
- 72 Deleting a Saver Period
- 72 Viewing Schedule for Week
- 74 Setting Up for Testing
- 84 Running a DIN Acceptance Test
- 85 Running a DIN Constancy Test
- 87 Viewing DIN Test Results
- 90 AAPM TG18 Test Setup
- 96 Completing AAPM TG18 Tests
- 98 Viewing AAPM TG18 Test Results
- 100 Logging On as Privileged User/Administrator
- 101 Logging Out of Privileged Mode
- 101 Changing Your Password
- 102 Viewing Reports
- 106 Using the Reporting Control
- 109 SNMP Management Console
- 109 The SNMP Model
- 110 Configuring the SNMP Service