C HAPTER 6 ______________________________________________ CALIBRATION AND ADJUSTMENT

RH : 11.5 Ref1 ? 11.3 <cr>

Press any key when ready...

Place the sensor head in the calibration hole of the NaCl bottle

(wet end reference) in the humidity calibrator.

Wait for 30 minutes.

Press any key and enter the second point value and press <cr>.

RH : xx.x Ref2 ? yy.y <cr>

The stabilization of the sensor can be monitored here as well by entering c <cr> instead of the reference value.

Using display/keypad commands

Leave the calibrator and the transmitter for at least 4 hours in the same space so that their temperatures have time to equalize.

Remove the filter cap on the transmitter.

Disconnect the security lock jumper before turning the transmitter on!

Place the sensor head in the calibration hole of the LiCl bottle (dry end reference) in the humidity calibrator.

Select

Cali

in the main menu and then

RH

; select

Not changed and then two-point calibration

RH 2 point cal

. Change the first point reading with the arrow keys to correspond the reference humidity and press ENT; pressing an arrow once changes the reading by 0.05 %RH.

Place the sensor head in the calibration hole of the NaCl bottle

(wet end reference) in the humidity calibrator.

Wait for 30 minutes.

V AISALA _________________________________________________________________________ 57