C HAPTER 6 ______________________________________________ CALIBRATION AND ADJUSTMENT
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
•
Place the sensor head in the calibration hole of the NaCl bottle
(wet end reference) in the humidity calibrator.
•
Wait for 30 minutes.
•
Press any key and enter the second point value and press <cr>.
RH : xx.x Ref2 ? yy.y <cr>
•
The stabilization of the sensor can be monitored here as well by entering c <cr> instead of the reference value.
Using display/keypad commands
•
Leave the calibrator and the transmitter for at least 4 hours in the same space so that their temperatures have time to equalize.
Remove the filter cap on the transmitter.
Disconnect the security lock jumper before turning the transmitter on!
•
Place the sensor head in the calibration hole of the LiCl bottle (dry end reference) in the humidity calibrator.
•
Select
Cali
in the main menu and then
RH
; select
Not changed and then two-point calibration
RH 2 point cal
. Change the first point reading with the arrow keys to correspond the reference humidity and press ENT; pressing an arrow once changes the reading by 0.05 %RH.
•
Place the sensor head in the calibration hole of the NaCl bottle
(wet end reference) in the humidity calibrator.
•
Wait for 30 minutes.
V AISALA _________________________________________________________________________ 57
advertisement
Related manuals
advertisement
Table of contents
- 5 CHAPTER 1 GENERAL INFORMATION
- 7 CHAPTER 2 PRODUCT DESCRIPTION
- 10 Use in high pressure
- 11 CHAPTER 3 INSTALLATION
- 13 Mounting; overview
- 14 Mounting the probe directly to the process
- 15 Installing the probe through the ball valve assembly
- 21 Mounting the probe with a quick connect
- 22 Sample cell
- 24 Grounding
- 26 Connections
- 27 Connection to an AC supply
- 29 CHAPTER 4 COMMISSIONING
- 37 CHAPTER 5 COMMANDS
- 59 CHAPTER 6 CALIBRATION AND ADJUSTMENT
- 60 Two-point calibration and adjustment adjustment procedure
- 60 Using serial commands
- 61 Using display/keypad commands
- 62 Using LED commands
- 63 Humidity calibration table
- 63 One point offset correction
- 63 Using serial commands
- 64 Using display/keypad commands
- 64 Using LED commands
- 65 Two-point temperature calibration and adjustement
- 65 Using serial commands
- 66 Using display/keypad commands
- 66 Using LED commands
- 67 Using serial commands
- 67 Using display/keypad commands
- 68 Using LED commands
- 71 CHAPTER 7 MAINTENANCE
- 76 CHAPTER 8 TECHNICAL DATA
- 80 CHAPTER 9 OPTIONS
- 81 CHAPTER 10 SPARE PARTS AND ACCESSORIES
- 83 APPENDIX 1 SERIAL COMMANDS
- 105 APPENDIX 2 POWER SUPPLY MODULE
- 109 APPENDIX 3 INSTALLING AND USING THE RS 485/422 SERIAL PORT MODULE
- 119 APPENDIX 4 INSTALLING AND USING THE CURRENT LOOP MODULE
- 129 APPENDIX 5 ERROR MESSAGES
- 133 APPENDIX 6 WIRING DIAGRAM
- 135 APPENDIX 7 ALARM OUTPUT UNIT
- 141 APPENDIX 8 PRESSURE CONVERSION CHART
- 141 Multiplication factors