C HAPTER 6 ______________________________________________ CALIBRATION AND ADJUSTMENT
•
Check that the reading corresponds with the reading of the reference instrument. If not, adjust the second point.
•
If necessary, adjust with the arrow switches to the correct value and press ENT. At the second calibration point the second LED from the left flashes.
Calibration of the analogue outputs
The analogue outputs have been calibrated at the factory and since they are very stable, calibration of the outputs should be performed only when there is reason to believe that their adjustments have changed.
Using serial commands
Disconnect the security lock jumper!
ACAL <cr>
The outputs on channels 1 and 2 are measured and the measured values (mA or V) entered as calibration coefficients.
Example: both channels have 0...10 V outputs (set with AMODE command); enter the voltages measured at the analogue outputs:
>ACAL <cr>
Ch1 U1 ( V ) ? 0.123 <cr>
Ch1 U2 ( V ) ? 9.98 <cr>
Ch2 U1 ( V ) ? 0.120 <cr>
Ch2 U2 ( V ) ? 9.98 <cr>
Using display/keypad commands
Disconnect the security lock jumper!
•
Connect an ammeter/voltmeter to the output of channel 1, select
Cali
in the main menu and
Analog outputs
in the Cali menu. The following is displayed (the quantity can be either mA or V):
V AISALA _________________________________________________________________________ 63
advertisement
Related manuals
advertisement
Table of contents
- 5 CHAPTER 1 GENERAL INFORMATION
- 7 CHAPTER 2 PRODUCT DESCRIPTION
- 10 Use in high pressure
- 11 CHAPTER 3 INSTALLATION
- 13 Mounting; overview
- 14 Mounting the probe directly to the process
- 15 Installing the probe through the ball valve assembly
- 21 Mounting the probe with a quick connect
- 22 Sample cell
- 24 Grounding
- 26 Connections
- 27 Connection to an AC supply
- 29 CHAPTER 4 COMMISSIONING
- 37 CHAPTER 5 COMMANDS
- 59 CHAPTER 6 CALIBRATION AND ADJUSTMENT
- 60 Two-point calibration and adjustment adjustment procedure
- 60 Using serial commands
- 61 Using display/keypad commands
- 62 Using LED commands
- 63 Humidity calibration table
- 63 One point offset correction
- 63 Using serial commands
- 64 Using display/keypad commands
- 64 Using LED commands
- 65 Two-point temperature calibration and adjustement
- 65 Using serial commands
- 66 Using display/keypad commands
- 66 Using LED commands
- 67 Using serial commands
- 67 Using display/keypad commands
- 68 Using LED commands
- 71 CHAPTER 7 MAINTENANCE
- 76 CHAPTER 8 TECHNICAL DATA
- 80 CHAPTER 9 OPTIONS
- 81 CHAPTER 10 SPARE PARTS AND ACCESSORIES
- 83 APPENDIX 1 SERIAL COMMANDS
- 105 APPENDIX 2 POWER SUPPLY MODULE
- 109 APPENDIX 3 INSTALLING AND USING THE RS 485/422 SERIAL PORT MODULE
- 119 APPENDIX 4 INSTALLING AND USING THE CURRENT LOOP MODULE
- 129 APPENDIX 5 ERROR MESSAGES
- 133 APPENDIX 6 WIRING DIAGRAM
- 135 APPENDIX 7 ALARM OUTPUT UNIT
- 141 APPENDIX 8 PRESSURE CONVERSION CHART
- 141 Multiplication factors