C HAPTER 6 ______________________________________________ CALIBRATION AND ADJUSTMENT

Connect an ammeter/voltmeter to the analogue outputs (connector

X2). Give command ¡¡l¡ . At the first calibration point the

LED on the left flashes; adjust the first point (offset) with the arrow switches to the same reading with the reference and press

ENT switch.

After adjusting the offset point and pressing ENT the second LED from left flashes. Press ENT without changing the output value.

Two-point temperature calibration and adjustement

Using serial commands

Leave the reference instrument and the transmitter for at least 4 hours in the same space so that their temperatures have time to equalize. Remove the filter cap prior to calibration.

Disconnect the security lock jumper before turning the transmitter on!

Check the transmitter against the reference.

Give command CT <cr>, enter the first point value and press <cr>:

>CT <cr>

T : xx.x Ref1 ? yy.y <cr>

Press any key when ready

If you want to see how the sensor stabilizes to the reference temperature, enter c <cr> instead of the first reference:

T : 0.90 Ref1 ? c <cr>

T : 0.55 Ref1 ? c <cr>

T : 0.55 Ref1 ? 0.0 <cr>

Press any key when ready...

Change the temperature and again check the transmitter against the reference.

Check that the reading corresponds with the reading of the reference instrument. If not, adjust the second point.

Press any key, enter the second point value and press <cr>.

T : xx.x Ref2 ? yy.y <cr>

V AISALA _________________________________________________________________________ 61