TH2515 Operation Manual

5.3.4

Temperature correction factor

When the instrument is used under the following environmental conditions:

0 ℃ ~18 ℃ , ≤80%RH; 28 ℃ ~40 ℃ , ≤80%RH

Measurement Accuracy: Multiply the basic measurement accuracy value listed in chapter 5.3.1

-5.3.4 by the temperature correction factor K in the following table.

Temperature( ℃ ) 0 - 5 5 - 18 18 - 28

Temperature correction factor K

4 2 1

28 - 35

2

35 - 40

4

-------------------------------------------------------------------------------------------------------------------------------